Remote control commands
R&S
®
NRPM
164
Mannual 1425.8663.02 ─ 08
If disabled, each measurement from the sensor module is averaged. If enabled,
only one sampling sequence per measurement is recorded, thus increasing the
measurement speed. With a higher measurement speed, the measured values of
an individual measurement are immediately delivered.
The averaging filter is not used, i.e. the following settings are ignored:
●
[SENSe<Sensor>:]TRACe:AVERage[:STATe]
●
[SENSe<Sensor>:]TRACe:AVERage:COUNt
Parameters:
<state>
*RST:
OFF
[SENSe<Sensor>:]TRACe:TIME
<time>
Sets the trace length, i.e. the time to be covered by the trace sequence. This time
period is divided into several equal intervals, in which the average power is deter-
mined. The number of intervals equals the number of trace points, which is set
with the command
Parameters:
<time>
Range:
10.0e-6 to 3.0
*RST:
0.01
Default unit: Seconds
Manual operation:
[SENSe<Sensor>:]TRACe:UPSample[:TYPE]
<type>
Selects an output mode for the acquired trace data.
Parameters:
<type>
SINC | HOLD
HOLD
Returns the trace data unchanged. The course of power
over time is represented as sampled by the sensor mod-
ule's data acquisition and processing logic.
Configuring the measurement modes