Operating Concepts
R&S
®
NRP-Zxx
33
Getting Started 1176.8822.02 ─ 04
Setup
Figure 6-2: Setup with an R&S
NRP2 base unit
1 = Signal source
2 = R&S NRP
‑
Z power sensor
3 = SENSOR connector of the R&S
NRP2
4 = R&S
NRP2 base unit
Incorrectly connecting/disconnecting the R&S NRP
‑
Z power sensors may
damage the power sensors or lead to erroneous results.
Ensure that you connect/disconnect your power sensors as described in
Chapter 4.4, "Connecting the R&S NRP
Chapter 4.5, "Disconnecting the R&S NRP
Starting a measurement
1. Connect the power sensor to the [SENSOR] connector of the R&S
NRP2 as
shown in
2. Preset the R&S
NRP2.
3. Execute zeroing.
Note:
Turn off all test signals before zeroing. An active test signal during zero-
ing causes an error.
4. Connect the power sensor to the signal source.
The result window indicates the result (in dBm) obtained with sensor A.
R&S
NRP2