Basic Safety Instructions 

1171.0000.42-05.00 

Page 3 

Electrical safety 

If the information on electrical safety is not observed either at all to the extent necessary, electric shock, 
fire and/or serious personal injury or death may occur. 

 

1.  Prior to switching on the product, always ensure that the nominal voltage setting on the product 

matches the nominal voltage of the AC supply network. If a different voltage is to be set, the power 
fuse of the product may have to be changed accordingly.  

2.  In the case of products of safety class I with movable power cord and connector, operation is 

permitted only on sockets with an earthing contact and protective earth connection. 

3.  Intentionally breaking the protective earth connection either in the feed line or in the product itself is 

not permitted. Doing so can result in the danger of an electric shock from the product. If extension 
cords or connector strips are implemented, they must be checked on a regular basis to ensure that 
they are safe to use. 

4.  If the product does not have a power switch for disconnection from the AC supply network, the plug of 

the connecting cable is regarded as the disconnecting device. In such cases, always ensure that the 
power plug is easily reachable and accessible at all times (corresponding to the length of connecting 
cable, approx. 2 m). Functional or electronic switches are not suitable for providing disconnection from 
the AC supply network. If products without power switches are integrated into racks or systems, a 
disconnecting device must be provided at the system level. 

5.  Never use the product if the power cable is damaged. Check the power cable on a regular basis to 

ensure that it is in proper operating condition. By taking appropriate safety measures and carefully 
laying the power cable, you can ensure that the cable will not be damaged and that no one can be 
hurt by, for example, tripping over the cable or suffering an electric shock.  

6.  The product may be operated only from TN/TT supply networks fused with max. 16 A (higher fuse 

only after consulting with the Rohde & Schwarz group of companies). 

7.  Do not insert the plug into sockets that are dusty or dirty. Insert the plug firmly and all the way into the 

socket. Otherwise, sparks that result in fire and/or injuries may occur.  

8.  Do not overload any sockets, extension cords or connector strips; doing so can cause fire or electric 

shocks. 

9.  For measurements in circuits with voltages V

rms

 > 30 V, suitable measures (e.g. appropriate 

measuring equipment, fusing, current limiting, electrical separation, insulation) should be taken to 
avoid any hazards. 

10.  Ensure that the connections with information technology equipment, e.g. PCs or other industrial 

computers, comply with the IEC60950-1/EN60950-1 or IEC61010-1/EN 61010-1 standards that apply 
in each case. 

11.  Unless expressly permitted, never remove the cover or any part of the housing while the product is in 

operation. Doing so will expose circuits and components and can lead to injuries, fire or damage to the 
product. 

12.  If a product is to be permanently installed, the connection between the PE terminal on site and the 

product's PE conductor must be made first before any other connection is made. The product may be 
installed and connected only by a licensed electrician. 

13.  For permanently installed equipment without built-in fuses, circuit breakers or similar protective 

devices, the supply circuit must be fused in such a way that anyone who has access to the product, as 
well as the product itself, is adequately protected from injury or damage. 

Содержание HZ-14

Страница 1: ...R S HZ 14 Probe Set for E and H Near Field Measurements Operating Manual 1026 7767 12 01 Test and Measurement Operating Manual ...

Страница 2: ...f the product The product is used for its designated purpose if it is used in accordance with its product documentation and within its performance limits see data sheet documentation the following safety instructions Using the product requires technical skills and a basic knowledge of English It is therefore essential that only skilled and specialized staff or thoroughly trained personnel with the...

Страница 3: ...he positions specified by the manufacturer without the product s ventilation being obstructed If the manufacturer s specifications are not observed this can result in electric shock fire and or serious personal injury or death Applicable local or national safety regulations and rules for the prevention of accidents must be observed in all work performed 1 Unless otherwise specified the following r...

Страница 4: ...at it is in proper operating condition By taking appropriate safety measures and carefully laying the power cable you can ensure that the cable will not be damaged and that no one can be hurt by for example tripping over the cable or suffering an electric shock 6 The product may be operated only from TN TT supply networks fused with max 16 A higher fuse only after consulting with the Rohde Schwarz...

Страница 5: ... damage may occur It is the responsibility of the employer operator to select suitable personnel for operating the products 2 Before you move or transport the product read and observe the section titled Transport 3 As with all industrially manufactured goods the use of substances that induce an allergic reaction allergens such as nickel cannot be generally excluded If you develop an allergic react...

Страница 6: ... sunlight must be avoided Keep cells and batteries clean and dry Clean soiled connectors using a dry clean cloth 3 Cells or batteries must not be short circuited Cells or batteries must not be stored in a box or in a drawer where they can short circuit each other or where they can be short circuited by other conductive materials Cells and batteries must not be removed from their original packaging...

Страница 7: ...le Adequately secure the product in the vehicle to prevent injuries or other damage in the event of an accident Waste disposal 1 If products or their components are mechanically and or thermally processed in a manner that goes beyond their intended use hazardous substances heavy metal dust such as lead beryllium nickel may be released For this reason the product may only be disassembled by special...

Страница 8: ... only be disposed of at a Rohde Schwarz service center or at a suitable depot Normas de Seguridad para Baterías Según BattV Este equipo lleva una batería que contiene sustancias per judiciales que no se debe desechar en los contenedores de basura domésticos Después de la vida útil la batería sólo se podrá eliminar en un centro de servicio de Rohde Schwarz o en un depósito apropiado Consignes de sé...

Страница 9: ...ance with EN 50419 Once the lifetime of a product has ended this product must not be disposed of in the standard domestic refuse Even disposal via the municipal collection points for waste electrical and electronic equipment is not permitted Rohde Schwarz GmbH Co KG has developed a disposal concept for the environmental friendly disposal or recycling of waste material and fully assumes its obligat...

Страница 10: ...ment system stan dards The Rohde Schwarz quality management system is certified according to standards such as ISO9001 and ISO14001 Environmental commitment Energy efficient products J J Continuous improvement in J J environmental sustainability ISO14001 certified environmental J J management system Cher client Vous avez choisi d acheter un pro duit Rohde Schwarz Vous disposez donc d un produit fa...

Страница 11: ...information and upgrades To keep your instrument up to date and to be informed about new application notes related to your instrument please send an e mail to the Customer Support Center stating your instrument and your wish We will take care that you will get the right information Europe Africa Middle East Phone 49 89 4129 12345 customersupport rohde schwarz com North America Phone 1 888 TEST RSA...

Страница 12: ...7767 12 01 1 Contents 1 Introduction 3 2 Operation 5 2 1 Magnetic Field Probes 5 2 2 Preamplifier for Magnetic Probes 7 2 3 Electric Probe 8 2 4 Probe Test jig 9 2 5 Probe Power Adapter for Electric Probe 10 3 Maintenance 11 3 1 Spare Parts 11 ...

Страница 13: ...r spectrum analyzer with the active electric probe obtaining power via the RF cable using the probe power adaptor The low noise preamplifier provided is used with the magnetic probes to increase their sensitivity and because of their low power requirements this and the electric probe can be conveniently powered from the probe power outlet of most R S Test Receivers and Spectrum Analyzers The probe...

Страница 14: ...ses through the end face of the probe tip generates a voltage at the probe output Near field Electric Probe The near field electric probe is an active probe with power supplied via the RF cable The probe power adapter includes an acoustic overload warning The pointed tip of this probe makes it especially useful for exactly locating sources of interference Probe Test jig The probe test jig is used ...

Страница 15: ...he probes in the set for comparative measurements care should be taken to obtain repeatable results i e the probe orientation in space should be the same for each measurement 2 To correctly measure fields the magnetic field probes must be orientated for maximum coupling with the magnetic field i e the end face of the probe top is perpendicular to the magnetic field When using the 30 MHz to 1 GHz p...

Страница 16: ...end face of the probe Table 1 Antenna factor for low frequency magnetic probe typical 9 kHz 89 dB FA m FV 100 kHz 70 dB FA m FV 1 MHz 52 dB FA m FV 10 MHz 43 dB FA m FV 30 MHz 42 dB FA m FV Fig 2 Antenna factor for magnetic probe 9 kHz to 30 MHz typical 0 10 20 30 40 50 60 70 80 90 100 0 001 0 01 0 1 1 10 100 Frequency MHz Antenna Factor dBuA m uV Table 2 Antenna factor for high frequency magnetic...

Страница 17: ... low noise amplification over the range 9 kHz to 1 GHz The preamplifier for magnetic probes has a wide dynamic range and with its 1 dB compression point of 8 dBm it can be used as a general purpose amplifier but at all times the maximum signal input level of 10 dBm must be strictly observed limit of safe operation The probes must be used without the amplifier if higher signal levels are to be meas...

Страница 18: ...e measuring equipment including the preamplifier and cause erroneous readings of any low level signals which exist at the same time Therefore care must be taken when making impulsive measurements as very high signal levels may be present 2 3 Electric Probe Unlike the magnetic probes the electric probe sensing element is a point sensor and is not affected by probe orientation The pointed end of thi...

Страница 19: ...obe the one with the maximum output should be used With the tracking generator output set to 10 dBm and the spectrum analyzer or measuring receiver set as appropriate a plot similar to Fig 6 should be displayed Typically the plot should be within 3 dB of those shown large changes e g 10 dB or more indicate a damaged probe Fig 6 Typical frequency response measured with the probe test jig Probe Test...

Страница 20: ... 2 5 Probe Power Adapter for Electric Probe The probe power adapter is used to power the electric probe It can be conveniently powered from the probe power outlet of most R S Test Receivers and Spectrum Analyzers Fig 7 Probe power adapter for electric probe ...

Страница 21: ...robes or the adapters No chemicals may be used for cleaning purposes 3 1 Spare Parts For the probe set for E and H near field measurements R S HZ 14 the following spare parts are available 0633 0740 00 Magnetic Near Field Probe 9 kHz 30 MHz 0633 0757 00 Magnetic Near Field Probe 30 MHz 1 GHz 0633 0763 00 Electric Near Field Probe 9 kHz 1 GHz 1313 1968 00 Pre Amplifier for Magnetic Near Field Probe...

Отзывы: