7
Subject to change without notice
Specifications
S p e c i f i c a t i o n s
Set of HZO40 probe consisting of:
1
HZO40 200 MHz differential probe 1 piece
2
Hook Clip, red
1 piece
Hook Clip, black
1 piece
3
Alligator Clip, red
1 piece
Alligator Clip, black
1 piece
4
USB Power Cord (2 m)
1 piece
5
9 V Battery
1 piece
Manual
1 piece
Electrical specifications
Bandwidth (-3 dB):
200 MHz
Attenuation Ratio:
10 : 1
Probe Risetime (10%-90%):
1.75 ns
Gain Accuracy:
±1%
Absolute Maximum Rated Input Voltage
(each side to ground):
±60 V
Maximum Differential
Input Voltage (DC+AC Peak):
±20 V
Maximum Common
Mode Input Voltage
±60 V
Input Resistance / Input capacity
each side to ground:
500 kΩ, 7 pF
between inputs:
1 MΩ, 3.5 pF
Output Voltage Swing:
±2 V (driving 50 Ω scope input)
Offset (typical):
±2 mV
CMRR (typical):
-80 dB at 60 Hz,
-50 dB at 10 MHz
Power Requirements:
one 9 V battery or USB power
adapter (5 V to 9 V, 200 mA)
Approx. Battery Life:
7.5 hours (alkaline battery)
Battery/Power Cord:
The supplied voltage must be
less than 12 V and greater than
4.5 V or else the probe could be
damaged
Mechanical specifications
Approximate Weight:
170 g
(not including battery and
accessories)
BNC Cable Length:
120 cm
Length of Input Leads:
15 cm
Dimensiones (LxWxH):
111 x 22 x 14 mm
Environmental specifications
Temperature range
operating:
5 °C to +40 °C
non-operating:
-
20 °C to +70 °C
Altitude operating:
3,000 m
non-operating:
15,300 m
Humidity:
25 - 85% room humidity
SPEcificaTiOnS
all data valid at 23 °C after 30 minutes warm up
4
1
2
3
5
Содержание Hameg HZ040