Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
76
User Manual 1173.9286.02 ─ 15
8.3 Controlling the Phase Noise Measurement Channel
The following commands are necessary to control the measurement channel.
........................................................................................ 76
.............................................................................................. 76
.......................................................................................... 77
...................................................................................79
INSTrument:CREate:DUPLicate
This command duplicates the currently selected measurement channel, i.e creates a
new measurement channel of the same type and with the identical measurement set-
tings. The name of the new channel is the same as the copied channel, extended by a
consecutive number (e.g. "IQAnalyzer" -> "IQAnalyzer2").
The channel to be duplicated must be selected first using the
INST:SEL
command.
Example:
INST:SEL 'IQAnalyzer'
INST:CRE:DUPL
Duplicates the channel named 'IQAnalyzer' and creates a new
measurement channel named 'IQAnalyzer2'.
Usage:
Event
INSTrument:CREate[:NEW]
<ChannelType>, <ChannelName>
This command adds an additional measurement channel.
The number of measurement channels you can configure at the same time depends on
available memory.
Parameters:
<ChannelType>
Channel type of the new channel.
For a list of available channel types see
<ChannelName>
String containing the name of the channel. The channel name is
displayed as the tab label for the measurement channel.
Note: If the specified name for a new channel already exists, the
default name, extended by a sequential number, is used for the
new channel (see
Example:
INST:CRE IQ, 'IQAnalyzer2'
Adds an additional I/Q Analyzer channel named "IQAnalyzer2".
Controlling the Phase Noise Measurement Channel