Analysis
R&S
®
FSW-K40
62
User Manual 1173.9286.02 ─ 15
Comment ← Select Limit Line
Shows the comment of the selected limit line. If the limit line has no comment, this field
stays empty.
View Filter ← Select Limit Line
Turns filter for the list of limit lines on and off.
By default, the list includes all limit lines that are stored on the R&S
FSW.
"Show Com-
patible"
Filters the list of limit lines by compatibility.
If on, the list includes only those limit lines that are compatible to the
current measurement setup.
"Show Lines
For PNoise"
Filters the list of limit lines by compatibility to phase noise measure-
ments.
If on, the list includes only those limit lines that are compatible to
phase noise measurements.
New / Edit / Copy To ← Select Limit Line
All three buttons open the "Edit Limit Line" dialog box to create or edit limit lines.
When you use the "New" button, the dialog box contains no data.
When you use the "Edit" button, the dialog box contains the data of the previously
selected limit line.
When you use the "Copy To" button, the dialog box also contains a copy the data of
the previously selected limit line.
Remote command:
New:
Copy:
Delete ← Select Limit Line
Deletes the selected limit line.
Remote command:
Disable All Lines ← Select Limit Line
Turns all active limit lines off.
6.3.3 Creating and Editing Standard Limit Lines
Access
: "Overview" > "Limit Analysis" > "Line Config" > "New" / "Edit" > "Copy To"
Using Limit Lines