Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
144
User Manual 1173.9286.02 ─ 15
Example:
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
CALC:DELT2 ON
Switches on delta marker 2.
CALC:DELT2:Y?
Outputs measurement value of delta marker 2.
Usage:
Query only
8.12.3 Configuring Markers
DISPlay:MINFo[:STAT]
<DisplayMode>
This command turns the marker information in all diagrams on and off.
Parameters:
<DisplayMode>
ON | 1
Displays the marker information in the diagrams.
OFF | 0
Hides the marker information in the diagrams.
*RST:
1
Example:
DISP:MINF OFF
Hides the marker information.
Manual operation:
See
DISPlay:MTABle
<DisplayMode>
This command turns the marker table on and off.
Parameters:
<DisplayMode>
ON | 1
Turns on the marker table.
OFF | 0
Turns off the marker table.
AUTO
Turns on the marker table if 3 or more markers are active.
*RST:
AUTO
Example:
DISP:MTAB ON
Activates the marker table.
Manual operation:
See
Using Markers