Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
120
User Manual 1173.9286.02 ─ 15
Parameters:
<HalfDecade>
Start offset frequency of the half decade you want to display.
Note that the half decade you want to display has to be part of
the current measurement range.
Range:
100 mHz...300 mHz to 3 GHz...10GHz
*RST:
Half decade display is off.
Example:
DISP:TRAC:X:HDEC 1KHZ
Displays the half decade beginning with 1 kHz.
Manual operation:
See
DISPlay[:WINDow]:TRACe:X[:SCALe]:SCOPe
<Scope>
This command selects the way the application scales the horizontal axis.
Parameters:
<Scope>
HDECade
Shows a particular half decade only.
You can select a particular half decade with .
MANual
Shows a custom part of the measurement range.
You can select the start and stop offsets with .
MRANGe
Shows the complete measurement range.
*RST:
MRANGe
Example:
DISP:TRAC:X:SCOP MRAN
Shows the complete measurement range on the x-axis.
Manual operation:
See
DISPlay[:WINDow]:TRACe:X[:SCALe]:STARt
<StartFrequency>
This command selects the start frequency of the display range.
Before you can use the command you have to select a manual display range for the x-
axis with
DISPlay[:WINDow]:TRACe:X[:SCALe]:SCOPe
Parameters:
<StartFrequency>
Start offset frequency of a half decade.
Note that the start offset you want to display has to be part of the
current measurement range.
Range:
100 mHz to 3 GHz
*RST:
Manual display range is OFF
Example:
DISP:TRAC:X:STAR 100HZ
Defines 100 Hz as the start of the display range.
Manual operation:
See
Graphical Display of Phase Noise Results