Analysis
R&S
®
FSW-K40
52
User Manual 1173.9286.02 ─ 15
Remote command:
DISPlay[:WINDow]:TRACe:X[:SCALe]:SCOPe
X-Axis Start / Stop
Defines the start and stop frequency of the horizontal axis.
Note that the displayed frequency range is a detail of the measurement range. Regard-
less of the displayed frequency range, the application still performs all measurement
over the measurement range you have defined.
The range depends on the measurement range. and possible increments correspond
to the half decades.
Available for a manual "X Axis Scope".
Remote command:
X-axis start:
DISPlay[:WINDow]:TRACe:X[:SCALe]:STARt
X-axis stop:
DISPlay[:WINDow]:TRACe:X[:SCALe]:STOP
Half Decade
Selects the half decade that is displayed.
Available if you have selected the half decade "X Axis Scope".
Remote command:
DISPlay[:WINDow]:TRACe:X[:SCALe]:HDECade
Y Axis Scaling
Selects the type of scaling for the vertical axis.
"Auto"
Automatically scales the vertical axis.
"Top & Bot-
tom"
Allows you to set the values at the top and bottom of the vertical axis.
"Top & Range" Allows to set the value at the top of the vertical axis and its range.
"Bottom &
Range"
Allows you to set the value at the bottom of the vertical axis and its
range.
Remote command:
Automatic scaling:
DISPlay[:WINDow]:TRACe:Y[:SCALe]:AUTO
Manual scaling:
DISPlay[:WINDow]:TRACe:Y[:SCALe]:MANual
Top / Range / Bottom
Define the top and bottom values or the range of the vertical axis.
Top defines the top values of the vertical axis. The unit is dBm/Hz.
Bottom defines the bottom value of the vertical axis. The unit is dBm/Hz.
Range defines the range of the vertical axis. The unit is dB.
The availability of the three fields depends on the type of manual "Y Axis Scaling" you
have selected.
Configuring Graphical Result Displays