Measurement configuration
The following commands are required to activate external generator control and to con-
figure a calibration measurement with an external tracking generator.
SOURce<si>:EXTernal<gen>:FREQuency
......................................................................1076
SOURce<si>:EXTernal<gen>:FREQuency:COUPling[:STATe]
.......................................... 1076
SOURce<si>:EXTernal<gen>:FREQuency[:FACTor]:DENominator
....................................1077
SOURce<si>:EXTernal<gen>:FREQuency[:FACTor]:NUMerator
....................................... 1077
SOURce<si>:EXTernal<gen>:FREQuency:OFFSet
......................................................... 1078
SOURce<si>:EXTernal<gen>:POWer[:LEVel]
................................................................. 1078
SOURce<si>:EXTernal<gen>[:STATe]
............................................................................ 1079
SOURce<si>:POWer[:LEVel][:IMMediate]:OFFSet
...........................................................1079
SOURce<si>:EXTernal<gen>:FREQuency
<Frequency>
This command defines a fixed source frequency for the external generator.
Suffix:
<si>
.
irrelevant
<gen>
Parameters:
<Frequency>
Source frequency of the external generator.
*RST:
1100050000
Default unit: HZ
Example:
//Define frequency of the generator
SOUR:EXT:FREQ 10MHz
Manual operation:
See
SOURce<si>:EXTernal<gen>:FREQuency:COUPling[:STATe]
<State>
This command couples the frequency of the external generator output to the
R&S
FSV/A.
Suffix:
<si>
.
irrelevant
<gen>
Parameters:
<State>
ON | OFF | 0 | 1
Setting basic measurement parameters
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...