Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1078
User Manual 1178.8520.02 ─ 08
Offset
Analyzer
Generator
F
ator
Deno
Numerator
F
F
min
Suffix:
<si>
.
irrelevant
<gen>
Parameters:
<Value>
<numeric value>
*RST:
1
Example:
//Define multiplication factor of 4/3; the transmit frequency of the
generator is 4/3 times the analyzer frequency
SOUR:EXT:FREQ:NUM 4
SOUR:EXT:FREQ:DEN 3
Manual operation:
See
"(Automatic) Source Frequency (Numerator/Denominator/
SOURce<si>:EXTernal<gen>:FREQuency:OFFSet
<Offset>
This command defines the frequency offset of the generator with reference to the ana-
lyzer frequency.
Select the offset such that the frequency range of the generator is not exceeded if the
following formula is applied to the start and stop frequency of the analyzer:
Offset
Analyzer
Generator
F
ator
Deno
Numerator
F
F
min
Suffix:
<si>
.
irrelevant
<gen>
Parameters:
<Offset>
<numeric value>, specified in Hz, kHz, MHz or GHz, rounded to
the nearest Hz
*RST:
0 Hz
Default unit: HZ
Example:
//Define an offset between generator output frequency and ana-
lyzer frequency
SOUR:EXT:FREQ:OFFS 10HZ
Manual operation:
See
"(Automatic) Source Frequency (Numerator/Denominator/
SOURce<si>:EXTernal<gen>:POWer[:LEVel]
<Level>
This command sets the output power of the selected generator.
Setting basic measurement parameters
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...