FSMR Measurement
Examples
1166.3311.12-01 1.5
E-1
1. Zero the Power Sensor
PRESET
Reset the instrument.
Press the PRESET key.
The main receiver menu is opened. The receiver mode is set.
PWR METER
Switching to Power Meter Mode
Press the PWR METER hotkey.
The power meter menu is displayed.
FREQ
RECEIVER
FREQUENCY
Set the measuring frequency
Press the
FREQ
key on the front panel.
The measuring receiver frequency menu is displayed.
In the frequency menu press the
RECEIVER FREQUENCY
softkey.
The frequency entry is displayed. Enter the measuring frequency
with the number keys and the appropiate unit key. The frequency
will be used by the instrument to calculate the calibration factors for
the power sensor head.
In the RECEIVER FREQUENCY entry window key in “
1000 MHZ”
.
PWR METER
ZERO
Zero the power sensor
Press the
PWR METER
hotkey.
The power meter menu is displayed.
In the POWER METER menu press the
ZERO
softkey.
In the message box, confirm CONTINUE and press ENTER.
The zeroing of the sensor will be performed.
After the zeroing is completed, a message will appear on the
screen.