FSMR Measurement
Examples
1166.3311.12-01 1.15
E-1
DEMOD
RELATIVE
+- PEAK/2
2. Relative Audio measurements – audio frequency response
Press the
DEMOD
hotkey.
The analog demodulation menu is displayed.
In the DEMOD menu press the
RELATIVE
softkey.
The RELATIVE menu is displayed.In the relative menu, only the
detectors which have been switched on in the DETECTOR menu
are available for relative measurements.
In the RELATIVE menu press the
RELATIVE DB %
softkey.
The actual measurement value of the detector will be saved as a
reference value and the result display will change to a relative
reading (indicated with
K
0.00 %).
Change the modulation frequency to 50 kHz
The following settings are used on the generator:
1. Select the AM Modulation, 50 kHz modulation frequency.
RELATIVE
RELATIVE
DB %
Relative display – Change the Unit to dB
Press the
DEMOD
hotkey.
The analog demodulation menu is displayed.
In the DEMOD menu press the
RELATIVE
softkey.
The RELATIVE menu is displayed.
In the RELATIVE menu press the
+- PEAK/2
softkey.
The actual result display will change from a relative reading
indicated in % to a dB reading.
Fig.1-11
Relative measurement in the demod mode