Measurements & Settings
Frequency
Settings
Software Manual 1300.6637.42 - 03
56
2.4.1.5 Fixed LO
The
Fixed LO
is the fixed local oscillator frequency for measurement
Modes.
I
Fixed LO, IF = RF + LO
I
Fixed LO, IF = abs(RF – LO)
Changing the
Fixed LO
will replace all LO values in the Frequencies List (the list of
measurement steps).
2.4.1.6 Fixed IF
The
Fixed IF
is the fixed intermediate frequency for measurement
Modes
.
I
Fixed IF, LO = RF + IF
I
Fixed IF, LO = abs(RF – IF)
Changing the
Fixed IF
will replace all IF values in the Frequencies List (the list of
measurement steps).
2.4.1.7 Image Rejection
The
Image Rejection
is the suppression applied to the second sideband during
calculations for measurement
Modes
:
I
Fixed IF, LO = RF + IF
I
Fixed IF, LO = abs(RF – IF)
I
Fixed LO, IF = RF + LO
I
Fixed LO, IF = abs(RF – LO)