The command is only available with the preselector option B2 or electronic
attenuator option B25.
3.1.11 INSTrument Subsystem
INSTrument:SELect
This remote control command selects active operation of the R&S FS-K30 option
by specifying its name.
Example
“
INST:SEL NOIS
'The R&S FS-K30 option will be selected as the active option.
Characteristics
*RST value: SANalyzer
SCPI: conforming
INSTrument:NSELect
This remote control command selects active operation of the R&S FS-K30 option
by specifying its associated option number.
Example
INST:NSEL 19
'The R&S FS-K30 option will be selected as the active option.
Characteristics
*RST value: 1
SCPI: conforming
3.1.12 SENSe Subsystem
The SENSe command is used to set and get the values of parameters in the remote
instrument. The get variant of the SENSe command differs from set in that it takes no
parameter values (unless otherwise stated) but is followed by the character ‘?’ and will
return the parameter’s value in the same format as it is set.
E.g.
SENS:SWE:TIME 10MS
– sets the sweep time to 10 milliseconds
SENS:SWE:TIME?
- response 0.01
- returns the current
sweep time
The SENSe subsystem is divided into a number of subsystems. The main areas being,
the commands used to control the equipment and measurement settings. The SENSe
command will be divided into equipment settings, general measurement settings and
specific measurement settings (one subsystem for each type of measurement).