Remote Control
FPC1000
225
User Manual 1178.4130.02 ─ 02
Example:
//More complex limit line
CALC:LIM:DEF 'Limit','Empty
comment',HZ,ABS,DBM,
0,-70,100000000,-70,100000000,-80,1000000000,
-80,1000000000,-70,2000000000,-70
CALC:LIM:UPP:SEL 'limit.abslim'
Usage:
Setting only
CALCulate<n>:LIMit<k>:DELete
This command removes a limit line from the evaluation.
Removing a limit line also ends the limit check.
Prerequisites for this command
●
Select a limit line or define a threshold limit with one of the following commands:
–
CALCulate<n>:LIMit<k>:LOWer:SELect
–
CALCulate<n>:LIMit<k>:LOWer:THReshold
–
CALCulate<n>:LIMit<k>:UPPer:SELect
–
CALCulate<n>:LIMit<k>:UPPer:THReshold
Suffix:
<n>
.
irrelevant
<k>
Example:
//Remove limit line
CALC:LIM2:DEL
CALCulate<n>:LIMit<k>:FAIL?
This command queries the result of a limit check.
To get a valid result, you have to perform a complete sweep with synchronization to
the sweep end before reading out the result. This is only possible in single sweep
mode.
Prerequisites for this command
●
Select a limit line or define a threshold limit with one of the following commands:
–
CALCulate<n>:LIMit<k>:LOWer:SELect
–
CALCulate<n>:LIMit<k>:LOWer:THReshold
–
CALCulate<n>:LIMit<k>:UPPer:SELect
–
CALCulate<n>:LIMit<k>:UPPer:THReshold
Suffix:
<n>
.
irrelevant
<k>
Remote Commands of the Spectrum Application
北京海洋兴业科技股份有限公司(证券代码:839145)
www.hyxyyq.com