Operating the instrument
R&S
®
FE50DTR
27
Manual 1179.3180.02 ─ 02
6
Operating the instrument
You can operate the R&S
FE50DTR in single receive mode (Rx), single transmit
mode (Tx) or simultaneous mode. In simultaneous mode, the R&S
FE50DTR
operates in Rx and Tx mode simultaneously also referring to dual transmit receive
6.1
Operating the R&S
FE50DTR in Rx mode
In Rx mode, the R&S
FE50DTR receives a high-frequency RF signal from the
device under test (DUT). It downconverts this signal using its own internal local
oscillator to an intermediate frequency (IF) signal that is a low-frequency signal.
The R&S
FE50DTR outputs and transmits the IF signal to the vector signal ana-
lyzer.
Test setup
Vector signal
analyzer
DUT
RF shield box
Ref Out
RF
LAN
IF Out A
LAN
RF A
Ref In
R&S
FE50DTR
LAN
IF Out
Ref
RF In
The following step-by-step descriptions use an R&S
FSV/A as vector signal ana-
lyzer covering the following topics:
●
●
FE50DTR with a vector signal analyzer"
●
"To configure frequency settings at the vector signal analyzer"
If supported, the descriptions for other Rohde
&
Schwarz vector signal analyzers
are analogous.
To connect the R&S
FE50DTR for Rx mode
The R&S
FE50DTR is
.
► Connect all connectors for connections between the R&S
FE50DTR and
R&S
FSV/A, and R&S
FE50DTR (RF in) and DUT as in
Operating the R&S
FE50DTR in Rx mode