Remote Commands
R&S
®
EVSG1000
343
User Manual 1178.6227.02 ─ 06
Usage:
Query only
Manual operation:
See
"Time reference status (Time Ref.)"
GBAS:RESETBER
Resets the values for the following measurement results:
●
"Bit error rate (BER) before FEC"
●
●
Parameters:
<State>
READY.
The command was executed successfully.
Usage:
Event
Manual operation:
See
GBAS:STOPSTREAM
This command terminates the data stream for both RX Boards in GBAS mode.
Return values:
<State>
READY.
The command was executed successfully.
GBAS:STREAM
<ParamSel>, [+MSG,] <RX Board>, <Slots>
Starts data streaming. In streaming mode, a measured data set is automatically sent to
the remote interface. One data set is provided per slot, that is: every 62.5
ms. Stream-
ing can produce large quantities of output. It is mandatory to have a fast data connec-
tion and an application that is able to handle the data.
Chapter 10.12.3.6, "GBAS Mode"
Parameters:
<ParamSel>
Determines which parameter selection is output.
FULL
EVS300 compatibility mode
ALL
All values of the current measurement mode are sent as comma
separated text, using actual units.
MEDIUM
A fixed selection of parameters is exported.
SHORT
A short fixed selection of parameters is exported.
RAW
Only the non-interpreted measurement values from the bit-
stream in binary format are exported
GBAS/ SCAT-I Mode