Parameters:
<Slot>
0 | 1 | 2 | 3 | 4 | 5 | 6 | 7
The slot for which results are determined.
Manual operation:
See
SETUP:GBAS_PPS_SOURCE?
<Source>
SETUP:GBAS_PPS_SOURCE?
<Source>
Gets or sets the source of the PPS signal required as a time reference.
Return values:
<Source>
TRIG
Signal provided at the Trigger In connector
PPS SMA
Signal provided at the PPS In connector
PPS GPS
Signal provided at the RS232-GPS connector
10.6.2
Retrieving GBAS/ SCAT-I Results
The following commands are required to retrieve results from GBAS/ SCAT-I measure-
ments.
............................................................................................. 341
............................................................................................... 341
GBAS:EVM?
<Slot>
Queries the error vector magnitude measured over the period of the burst.
If no burst was detected, the arithmetic average of the noise power is returned.
GBAS/ SCAT-I Mode