Data Management
R&S
®
EVSG1000
223
User Manual 1178.6227.02 ─ 06
8.2.5
Viewing Recorded Data
The "Recording" view displays the recorded data for the current measurement mode
from the selected data file (see
The following functions are available in the "Recording" view for all modes.
DL Line
Selects a particular line (record) in the data recording file.
8.2.6
Analyzing Recorded I/Q Data
Optionally, raw, unprocessed I/Q data is recorded to a separate file during recording
(see
Additional information on the I/Q data is stored in the CSV data list, indicating where in
the
.iq
file the I/Q data for a specific data set starts and how many samples it con-
tains. You can use this information to synchronize the raw I/Q data with the evaluated
measurement data for each data set.
You can also use the stored I/Q data to reproduce the measured signal on a signal
generator. Using the R&S
ARB
Toolbox, for example, you can convert the
.iq
data file
to a
.wv
file.
To convert the .iq data file to a .wv file
1. In the R&S
ARB
Toolbox, right-click the
.iq
data file.
2. Select "Convert" > "Convert from Data".
3. Select the following settings for the data format:
● "File Type": "Binary"
● "Data format": "Integer"
● "Endianness": "Little-Endian (x86)"
● "Precision": "32 Bit"
● "Signedness": "Signed (Two's Complement)"
4. Define the clock rate as
125 kHz
.
8.3
Creating and Storing Screenshots
Access
: [Setup] > "Screenshots"
You can create a screenshot of the current display on the R&S
EVSG1000 at any time
during operation. Creating screenshots is useful to document measurement results, for
example.
Creating and Storing Screenshots