Data Management
R&S
®
EVSG1000
221
User Manual 1178.6227.02 ─ 06
"None"
Message data is not included in the slot data record.
Remote command:
Graph Log ← Data selection (Select, GBAS/SCAT-I mode only)
Defines whether graphical information such as the time domain and constellation val-
ues are stored.
Remote command:
Slot <x> Log ← Data selection (Select, GBAS/SCAT-I mode only)
Activates or deactivates data recording for the corresponding slot.
To toggle between the slots A to D and slots E to H in the menu, press the "More soft-
keys" key.
Export List to USB
Exports the data list to a connected USB storage device.
Remote command:
Clear All Lists
Deletes all entries in all data lists.
Remote command:
Rec IQ
If enabled, the raw, unprocessed I/Q data is also recorded. This feature requires the
R&S
EVSG1-K25 option.
I/Q data is stored as a separate file with the same file name as the data list, but the
extension
.iq
. Furthermore, the data list contains two additional entries:
"IQSamples": indicates how many samples the data set includes; depending on the
measurement time this number can vary.
"IQPosition": indicates the starting position of the data set within the I/Q data file
Remote command:
8.2.4
Activating Data Recording
By default, data is recorded in a specific time interval (see
as recording is active. Alternatively, you can use an external trigger, if available, to
determine when data is stored. For example, you can store data at regular distances
during a moving measurement, based on a rotary position transducer.
Recording Measurement Data