Graphic Measurement Modes
R&S
®
EVSG1000
143
User Manual 1178.6227.02 ─ 06
Remote command:
Transducer Correction
Configures the level correction. The specified value is added to the measured power
levels to compensate for an inherent offset by the measurement setup, for example the
antenna.
Remote command:
5.1.4
Configuring the Trace
Access
: [Meas]
Access VNC
: [a]
Trace settings determine how the display values are calculated from the measured val-
ues.
Trace mode
Defines the update mode for subsequent traces.
"Clear/Write"
Overwrite mode (default): the trace is overwritten by each sweep
"Average"
The average is determined over several sweeps
The
determines the number of averaging procedures.
"Max Hold"
The maximum value is determined over several sweeps and dis-
played. The R&S
EVSG1000 saves each trace point in the trace
memory only if the new value is greater than the previous one.
The
determines the number of sweeps to evaluate.
"RMS"
The RMS value for each trace point over several sweeps is deter-
mined and displayed.
The
determines the number of sweeps to evaluate.
Remote command:
Average Count
Determines the number of sweeps over which trace evaluation is performed (e.g. aver-
aging or maxhold).
RF Spectrum Mode (Option R&S