Remote Commands
R&S
®
EVSG1000
246
User Manual 1178.6227.02 ─ 06
SETATTMODE
This command selects the attenuation of the active mode ("ILS LOC","ILS GP", "ILS
MB", "VOR" or "COM" mode).
Parameters:
<Attenuation>
Select the attenuation type from the listed ones.
AUTO
The RF attenuation mode is selected automatically.
LN
(Low Noise) 15 dB pre-amplification
Provides a high sensitivity. Suitable when scanning the area for
distant signals.
NORM
0 dB
Provides a normal sensitivity.
LD
(Low Dist) 15 dB attenuation
Provides a low sensitivity. Suitable when analyzing a nearby sig-
nal, to avoid overload due to high-level signals.
Return values:
<State>
READY.
Command was executed successfully.
Manual operation:
See
See
See
See
SETUP:DEMODAMFM
<Mode>
Gets or sets the audio demodulation mode.
Parameters:
<Mode>
AM | FM
Return values:
<State>
READY.
The command was executed successfully.
SETUP:DEMOD_OUT_AGC
<Mode>
Determines whether the level for demodulation output is automatically corrected using
the automatic gain control (AGC).
Parameters:
<Mode>
OFF | ON
Common Instrument and Measurement Setup