Remote Control
R&S
®
ESRP
957
User Manual 1175.7445.02 ─ 08
Manual operation:
See
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
<Address>
This command changes the GPIB address of the unit.
Parameters:
<Address>
0 to 30
*RST:
(no influence on this parameter, factory default 20)
Example:
SYST:COMM:GPIB:ADDR 18
Usage:
SCPI confirmed
Manual operation:
See
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator
<Terminator>
This command changes the GPIB receive terminator.
According to the standard the terminator in ASCII is <LF> and/or <EOI>. For binary
data transfers (e.g. trace data) from the control computer to the instrument, the binary
code (0AH) used for <LF> might be included in the binary data block, and therefore
should not be interpreted as a terminator in this particular case. This can be avoided by
changing the receive terminator to EOI.
Output of binary data from the instrument to the control computer does not require
such a terminator change.
Parameters:
<Terminator>
LFEOI | EOI
*RST:
(no influence on this parameter, factory default
LFEOI)
Example:
SYST:COMM:GPIB:RTER EOI
Manual operation:
See
SYSTem:LXI:INFo?
This query returns the current parameters of the LXI class C.
Only user accounts with administrator rights are able to use LXI functionality.
Return values:
Return values
(example)
<current version> | <LXI class> | <Computername> |
<MAC adress> | <IP adress> | <Auto MDIX>
Usage:
Query only
Manual operation:
See
Network Connection