Remote Control
R&S
®
ESRP
739
User Manual 1175.7445.02 ─ 08
Example:
ESP:RANG1:TRAN 'test'
Sets the transducer called test for range 1.
Manual operation:
See
Configuring the Reference Range
The following commands define the reference range for the SEM sweep list.
............................................................................................. 739
[SENSe:]ESPectrum:FILTer[:RRC]:ALPHa
........................................................................739
[SENSe:]ESPectrum:FILTer[:RRC][:STATe]
....................................................................... 740
.........................................................................................740
........................................................................................... 740
[SENSe:]ESPectrum:BWID
<Bandwidth>
This command defines the bandwidth used for measuring the channel power (refer-
ence range). This setting takes only effect if channel power is selected as power refer-
ence type (see
Parameters:
<Bandwidth>
minimum span ≤ value ≤ span of reference range
*RST:
3.84 MHz
Example:
ESP:RTYP CPOW
Sets the power reference type to channel power.
ESP:BWID 1 MHZ
Sets the Tx bandwidth to 1 MHz.
Manual operation:
See
[SENSe:]ESPectrum:FILTer[:RRC]:ALPHa
<Value>
This command sets the alpha value of the RRC filter. This setting takes only effect if
channel power is selected as power reference type (
command) and if the RRC filter is activated (
command).
Parameters:
<Value>
0 to 1
*RST:
0.22
Example:
ESP:RTYP CPOW
Sets the power reference type to channel power.
ESP:FILT ON
Activates the use of an RRC filter.
ESP:FILT:ALPH 0.5
Sets the alpha value of the RRC filter to 0.5.
Manual operation:
See