Remote Control
R&S
®
ESRP
740
User Manual 1175.7445.02 ─ 08
[SENSe:]ESPectrum:FILTer[:RRC][:STATe]
<State>
This command activates or deactivates the use of an RRC filter. This setting only takes
effect if channel power is selected as power reference type (see
Parameters:
<State>
ON | OFF
*RST:
ON
Example:
ESP:RTYP CPOW
Sets the power reference type to channel power.
ESP:FILT OFF
Deactivates the use of an RRC filter.
Manual operation:
See
[SENSe:]ESPectrum:RRANge
This command returns the current position (number) of the reference range.
Example:
ESP:RRAN?
Returns the current position (number) of the reference range.
[SENSe:]ESPectrum:RTYPe
<Type>
This command sets the power reference type.
Parameters:
<Type>
PEAK | CPOWer
PEAK
Measures the highest peak within the reference range.
CPOWer
Measures the channel power within the reference range (integral
bandwidth method).
*RST:
CPOWer
Example:
ESP:RTYP PEAK
Sets the peak power reference type.
Manual operation:
See
Configuring Power Classes
The following commands define the power classes for SEM measurements.
CALCulate<n>:LIMit<k>:ESPectrum:LIMits
.......................................................................741
CALCulate<n>:LIMit<k>:ESPectrum:MODE
......................................................................741
CALCulate<n>:LIMit<k>:ESPectrum:PCLass<Class>:COUNt
............................................. 742
CALCulate<n>:LIMit<k>:ESPectrum:PCLass<Class>[:EXCLusive]
......................................742
CALCulate<n>:LIMit<k>:ESPectrum:PCLass<Class>:LIMit[:STATe]
.................................... 742
Remote Commands in Spectrum Analyzer Mode