[SENSe:]POWer:ACHannel:BANDwidth|BWIDth[:CHANnel<channel>]
●
●
[SENSe:]POWer:ACHannel:PRESet:RLEVel
●
[SENSe:]POWer:BANDwidth|BWIDth
...............................................................................717
[SENSe:]POWer:BANDwidth|BWIDth
<Percentage>
This command defines the percentage of the power with respect to the total power.
This value is the basis for the occupied bandwidth measurement (see
Parameters:
<Percentage>
10 to 99.9PCT
*RST:
99PCT
Example:
POW:BWID 95PCT
Manual operation:
See
"% Power Bandwidth (span > 0)"
11.5.1.4
Measuring Spurious Emissions
The following commands configure and perform spurious emission measurements.
................................................................................717
.........................................................................................718
................................................................................... 726
.................................................................................................727
Controlling the Measurement
Note that the spurious measurement must be initialized before you can start configur-
ing the sweep list or list evaluation.
Commands useful to retrieve results described elsewhere:
●
.................................................................................... 718
INITiate<n>:SPURious
This command initiates a Spurious Emissions measurement.
Suffix:
<n>
.
irrelevant
Example:
INIT:SPUR
Starts a Spurious Emissions measurement.
Usage:
Event