background image

Remote Control

R&S

®

ESRP

1035

User Manual 1175.7445.02 ─ 08

Command

Known Differences

RLPOS

Adapts the position of the reference level even if the tracking generator normalization is
not active.

RQS

Supported bits:

1 (Units key pressed)

2 (End of Sweep)

3 (Device error)

4 (Command complete)

5 (Illegal command)

11.18.5

 

Model-Dependent Default Settings

If the GPIB language is switched over to an 85xx model, the GPIB address is automati-
cally switched over to 18 provided that the default address of the R&S

 

ESRP (20) is

still set. If a different value is set, this value is maintained. Upon return to SCPI, this
address remains unchanged.

The following table shows the default settings obtained after a change of the GPIB lan-
guage and for the commands IP, KST and RESET:

Model

# of Trace
Points

Start Freq.

Stop Freq.

Ref Level

Input Coupling

8566A/B

1001

2 GHz

22 GHz

0 dBm

AC

8568A/B

1001

0 Hz

1.5 GHz

0 dBm

AC

8560E

601

0 Hz

2.9 GHz

0 dBm

AC

8561E

601

0 Hz

6.5 GHz

0 dBm

AC

8562E

601

0 Hz

13.2 GHz

0 dBm

AC

8563E

601

0 Hz

26.5 GHz

0 dBm

AC

8564E

601

0 Hz

40 GHz

0 dBm

AC

8565E

601

0 Hz

50 GHz

0 dBm

AC

8594E

401

0 Hz

3 GHz

0 dBm

AC

Stop frequency

The stop frequency given in the table may be limited to the corresponding frequency
range of the R&S

 

ESRP.

Command LF sets the stop frequency for 8566A/B to a maximum value of 2 GHz.

Test points (trace points)

The number of trace points is switched over only upon transition to the REMOTE state.

GPIB Commands of HP Models 856xE, 8566A/B, 8568A/B and 8594E

Содержание ESRP Series

Страница 1: ...R S ESRP EMI Test Receiver User Manual User Manual Version 08 1175744502 2...

Страница 2: ...Co KG M hldorfstr 15 81671 M nchen Germany Phone 49 89 41 29 0 Fax 49 89 41 29 12 164 Email info rohde schwarz com Internet www rohde schwarz com Subject to change Data without tolerance limits is not...

Страница 3: ...Instructions and the product documentation in a safe place and pass them on to the subsequent users Riesgo de lesiones y da os en el instrumento El instrumento se debe usar de manera adecuada para pre...

Страница 4: ...die Produktdokumentation gut auf und geben Sie diese an weitere Benutzer des Produkts weiter Risque de blessures et d endommagement de l appareil L appareil doit tre utilis conform ment aux prescript...

Страница 5: ...ument up to date and to be informed about new application notes related to your instrument please send an e mail to the Customer Support Center stating your instrument and your wish We will take care...

Страница 6: ...Data Sheets and Brochures 24 2 7 Release Notes and Open Source Acknowledgment OSA 24 2 8 Application Notes Application Cards White Papers etc 24 3 Quick Start 25 3 1 Front and Rear Panel View 25 3 1...

Страница 7: ...2 4 4 Setting the Screen Colors 48 3 2 4 5 Setting the Display Power Save Function 50 3 2 4 6 Selecting and Configuring Printers 51 3 2 5 Windows Operating System 53 3 2 5 1 Virus Protection 54 3 2 5...

Страница 8: ...zed Display 81 3 4 4 3 Soft and Mini Soft Front Panel Display 81 3 4 4 4 Enlarging the Display Area 84 3 4 4 5 Displaying the Toolbar 84 3 4 4 6 Zooming into the Diagram 84 3 4 4 7 Adding a Title to t...

Страница 9: ...n with Traces 115 3 5 5 4 Configuring Automatic Loading 115 4 Measurement Modes 117 4 1 Receiver Mode 117 4 2 Spectrum Mode 117 4 3 I Q Analyzer Mode 118 4 4 Measurement Mode Root Menus HOME Key 118 5...

Страница 10: ...rned Values ASCII Format and Binary Format 153 5 3 Common Measurement Settings 154 5 3 1 Defining the Frequency and Span 155 5 3 2 Configuring the Level Display and the RF Input 157 5 3 3 Selecting th...

Страница 11: ...suring Channel Power and Adjacent Channel Power 201 6 1 1 3 Measuring the Carrier to Noise Ratio 225 6 1 1 4 Measuring the Occupied Bandwidth 228 6 1 1 5 Measuring with Spectrum Emission Masks 232 6 1...

Страница 12: ...327 6 2 5 Defining Automatic Settings AUTO SET Key 327 6 2 5 1 Softkeys of the Auto Set Menu 327 6 2 6 Setting the Bandwidths and Sweep Time BW Key 331 6 2 6 1 Softkeys of the Bandwidth Menu 331 6 2...

Страница 13: ...tra of Complex Signals 404 6 4 3 1 Separating Signals by Selecting an Appropriate Resolution Bandwidth 404 6 4 3 2 Intermodulation Measurements 405 6 4 4 Measuring Signals in the Vicinity of Noise 407...

Страница 14: ...asurement 465 8 4 4 Reflection measurement 465 8 4 5 Normalization 465 8 4 6 Modulation internal Tracking Generator only 469 8 5 Displayed Information and Errors 471 9 System Configuration 473 9 1 Man...

Страница 15: ...sponses to Queries 542 11 1 5 Command Sequence and Synchronization 543 11 1 5 1 Preventing Overlapping Execution 544 11 1 6 Status Reporting System 545 11 1 6 1 Hierarchy of Status Registers 545 11 1...

Страница 16: ...the Instrument Without a Network 601 11 2 7 2 Creating Users 601 11 2 7 3 Changing the User Password 602 11 2 7 4 Mapping Network Drives 602 11 2 7 5 Sharing Directories only with Microsoft Networks 6...

Страница 17: ...ower Measurements 688 11 5 1 2 Measuring Channel Power 695 11 5 1 3 Measuring the Occupied Bandwidth 716 11 5 1 4 Measuring Spurious Emissions 717 11 5 1 5 Testing Against a Spectrum Emission Mask 727...

Страница 18: ...Generator 868 11 7 2 Controlling the External Tracking Generator 875 11 8 Common Commands 883 11 9 System Configuration 887 11 9 1 General System Configuration 887 11 9 2 Checking the System Configur...

Страница 19: ...r Commands 962 11 16 2 Reading out the EVENt Part 963 11 16 3 Reading Out the CONDition Part 963 11 16 4 Controlling the ENABle Part 964 11 16 5 Controlling the Negative Transition Part 964 11 16 6 Co...

Страница 20: ...ing a File on the Instrument 992 11 17 12 Spectrum Emission Mask Measurement 992 11 17 12 1 Using Predefined Standard Wibro 993 11 17 12 2 Defining 5 Ranges with all Parameters 993 11 17 13 Spurious E...

Страница 21: ...4 14 1 4 Creating Users 1045 14 1 5 Changing the User Password 1045 14 1 6 Logging on to the Network 1045 14 1 7 Automatic Login Function 1046 14 1 8 Mapping Network Drives 1047 14 1 9 Sharing Directo...

Страница 22: ...ions safety issues are grouped according to subjects For example one subject is electrical safety The Basic Safety Instructions are delivered with the R S ESRP in different languages in print Througho...

Страница 23: ...hots When describing the functions of the product we use sample screenshots These screenshots are meant to illustrate as many as possible of the provided functions and possible interdependencies betwe...

Страница 24: ...ith the corre sponding help topic is displayed To change to the next tab press the tab on the touchscreen Navigating in the help topics To scroll through a page use the rotary knob or the UP ARROW and...

Страница 25: ...Preface R S ESRP 22 User Manual 1175 7445 02 08 Closing the help window Press the ESC key or a function key on the front panel How to Use the Help System...

Страница 26: ...ons It also provides an introduction to remote control a complete description of the remote control commands with programming examples and information on maintenance instrument interfaces and error me...

Страница 27: ...as optional accessories The brochure provides an overview of the instrument and deals with the specific char acteristics www rohde schwarz com brochure datasheet esrp 2 7 Release Notes and Open Source...

Страница 28: ...igure 3 1 Each element function keys and connectors is described in more detail in the subsequent sections Figure 3 1 Front panel of the R S ESRP No Description See 1 General device functions e g Chan...

Страница 29: ...pad on page 71 13 Rotary knob Chapter 3 4 2 5 Rotary Knob on page 72 14 Arrow keys Chapter 3 4 2 6 Arrow Keys UNDO REDO Keys on page 72 15 Undo redo functions Chapter 3 4 2 6 Arrow Keys UNDO REDO Keys...

Страница 30: ...te and time to establish a LAN connection Provides functionality to configure printers and hardcopys Displays the online help Selects the operating mode or firmware application Opens the Windows Start...

Страница 31: ...configure the sweep for example the measurement time the number of measurement points the measurement mode single or continuous measurements TRACE Provides functionality to configure data acquisition...

Страница 32: ...creen Display All measurement results are displayed in the screen on the front panel Additionally the screen display provides status and setting information and allows you to switch between various me...

Страница 33: ...6 Diagram footer with diagram specific information depending on measurement mode 7 Error indicator 8 Error message if available 9 Device status 10 Progress bar for measurement 11 Date and time displa...

Страница 34: ...ire a voltage of 28 V in order to be switched on and 0 V to be switched off The output supports a maximum load of 100 mA RF INPUT 50 The R S ESRP provides an RF input for connection of a device under...

Страница 35: ...e Before putting on the headphones make sure that the volume setting is not too high to protect your hearing 3 1 1 4 Optional Front Panel Connectors Tracking Generator Output GEN OUTPUT 50 Option R S...

Страница 36: ...Main Power Switch with fuse 3 1 2 1 Standard Rear Panel Connectors EMI impact on measurement results Electromagnetic interference EMI may affect the measurement results To suppress generated electrom...

Страница 37: ...ernal Monitor on page 45 EXT TRIGGER GATE IN The female connector for external trigger gate input is used to control the measure ment by means of an external signal The voltage levels can range from 0...

Страница 38: ...rs to connect devices like an exter nal keyboard or mouse You can also connect a memory stick to save and restore instrument settings and measurement data AUX PORT The 9 pole SUB D male connector prov...

Страница 39: ...the R S ESRP for the first time Risk of injury due to disregarding safety information Observe the information on appropriate operating conditions provided in the data sheet to prevent personal injury...

Страница 40: ...ESD can damage the electronic components of the instrument and the device under test DUT Electrostatic discharge is most likely to occur when you connect or disconnect a DUT or test fixture to the ins...

Страница 41: ...ing material Retain the original packing material If the instrument needs to be transported or ship ped later you can use the material to protect the control elements and connectors 3 2 1 2 Accessory...

Страница 42: ...Risk of injury when stacking instruments A stack of instruments can tilt over and cause injury if not stacked correctly Further more the instruments at the bottom of the stack can be damaged due to t...

Страница 43: ...oltage and frequency The AC power connector is located on the rear panel of the instrument Connect the R S ESRP to the AC power supply using the supplied power cable Since the instrument is assembled...

Страница 44: ...dicates that the DC power supply is ready for operation 3 In case of overvoltage or undervoltage the DC power supply switches off automati cally This state is indicated by means of a red LED 4 Set the...

Страница 45: ...o shut down the application properly 3 2 1 8 Maintenance The R S ESRP does not require regular maintenance Maintenance is essentially restricted to cleaning the R S ESRP It is however advisable to che...

Страница 46: ...w Align Results softkey Performing a self test The self test does not need to be repeated every time the instrument is switched on It is only necessary when instrument malfunction is suspected 1 Press...

Страница 47: ...em auto matically searches for a suitable device driver If Windows does not find a suitable driver it prompts you to specify a directory that contains the driver software If the driver software is on...

Страница 48: ...ect Start Settings Devices Device Manager Update Device drivers to install the driver 3 2 3 Connecting an External Monitor You can connect an external monitor to the MONITOR connector on the instrumen...

Страница 49: ...at the external reference signal is deactivated when switching from external to internal reference to avoid interactions with the internal reference signal Remote commands ROSC SOUR EXT ROSC EXT FREQ...

Страница 50: ...for daylight saving changes check box 5 Click OK 3 2 4 3 Aligning the Touchscreen When the device is delivered the touchscreen is initially aligned However it may become necessary to adjust the align...

Страница 51: ...brightness tint and color saturation of all screen objects 1 In the Screen Colors submenu see Displaying the Screen Colors Submenu on page 48 press the Set to Default softkey The Set User Colors to De...

Страница 52: ...ed Colors softkey For details refer to Defining and Using a User Defined Color Set on page 49 8 Click OK to accept the new settings and to close the dialog box Remote command DISP CMAP1 41 PDEF color...

Страница 53: ...at the steps for all objects for which you want to change the color 7 To change to predefined colors press the Predefined Colors softkey For details refer to Using the Predefined Color Set on page 48...

Страница 54: ...urns dark after the selected period of time Deactivating Display Power Save In the Display Setup submenu press the Display Pwr Save On Off softkey again Off is highlighted and the power save mode is s...

Страница 55: ...ork printer select the Printer option Note The Name Print to File and Orientation fields are only available if the Printer option is selected You can perform the following steps only with the Printer...

Страница 56: ...rintout For details how to proceed refer to Defining and Using a User Defined Color Set on page 49 Remote command HCOP CMAP DEF3 Screen Colors Hardcopy option The current screen colors without any cha...

Страница 57: ...tall anti virus software on the instru ment Rohde Schwarz does NOT recommend running anti virus software in the background on access mode on Windows based instruments due to potentially degrading inst...

Страница 58: ...described in Reactivating the automatic login function on page 56 Automatic Login Function When shipped the instrument automatically logs on the default Instrument user to Windows 7 using the default...

Страница 59: ...login 3 Save the changes to the file Reactivating the automatic login function 1 Press the Windows key or the CTRL ESC key combination on your keyboard to access the operating system of the R S ESRP s...

Страница 60: ...l chapter of the R S ESRP user manual To open the taskbar press the Windows key or the CTRL ESC key combination on your keyboard 3 2 6 Configuring the GPIB Interface A GPIB interface is integrated on...

Страница 61: ...ion using one of the following methods Copying the files to the instrument using USB devices e g a memory stick GPIB or LAN Using the Remote Installation feature in a LAN network Copying the files to...

Страница 62: ...ed and all found instruments are listed Note The setup procedure communicates with the instrument via LAN which means the file must pass the firewall Therefore add the Setup exe file to your firewall...

Страница 63: ...n by XML softkey An edit dialog box is displayed 5 Enter or browse for the name of an XML file on the instrument that contains the option key 6 Press Select If validation is successful the message opt...

Страница 64: ...Display Using the R S ESRP you can handle several different measurement tasks channels at the same time although they can only be performed asynchronously For each channel a separate tab is displayed...

Страница 65: ...indicates an invalid setting that does not provide useful results It is the user s responsibility to remedy such a situation Editing settings in the channel bar All settings that are displayed in the...

Страница 66: ...lt default not displayed Mode Indicates which sweep mode type is selected Auto FFT automatically selected FFT sweep mode Auto sweep automatically selected swept sweep mode FFT manually selected FFT sw...

Страница 67: ...t impedance of the instrument is set to 75 FRQ A frequency offset 0 Hz is set DC AC An external DC or AC calibration signal is in use Inp Input source digital I Q option R S FSV B17 only 3 4 1 4 Diagr...

Страница 68: ...x and y axis positions of the last 2 markers or delta markers that were set as well as their index are displayed within the diagram grid if available The value in the square brackets after the index...

Страница 69: ...rement function Func Result Result of the active marker or measurement function Mode dependant Information in Diagram Footer The diagram footer beneath the diagram contains the following information d...

Страница 70: ...following conditions applies Correction data has been switched off No correction values are available This occurs for example if a firmware update is performed followed by a cold start of the instrume...

Страница 71: ...page 75 All tasks necessary to operate the instrument can be performed using this user inter face Apart from instrument specific keys all other keys that correspond to an external keyboard e g arrow...

Страница 72: ...tion Repeats previously reverted operation Selection mode the cursor can be used to select and move markers in a zoomed display Zoom mode displays a dotted rectangle in the diagram that can be expande...

Страница 73: ...activated for the entire screen DIAGRAM TOUCH OFF touchscreen is deactivated for the diagram area of the screen but active for the surrounding softkeys toolbars and menus SCPI command DISP TOUCh STAT...

Страница 74: ...SC key Closes all kinds of dialog boxes if the edit mode is not active Quits the edit mode if the edit mode is active In dialog boxes that contain a Cancel button it activates that button For Edit dia...

Страница 75: ...the edit mode is activated For details on the edit mode refer to Chapter 3 4 3 3 Navigating in Dialog Boxes on page 78 3 4 2 6 Arrow Keys UNDO REDO Keys The arrow keys are used for navigation The UND...

Страница 76: ...the device Soft keys are dynamic i e depending on the selected function key a different list of soft keys is displayed on the right side of the screen A list of softkeys for a certain function key is...

Страница 77: ...ment results When you press a function key on the front panel the softkeys are displayed temporarily enabling you to perform a certain task with the softkeys They are then automatically hidden again w...

Страница 78: ...value In the documentation these dialog boxes are called edit dialog boxes Dialog boxes that are not only designed for parameter entry have a more complex structure and in the documentation are called...

Страница 79: ...c input the keypad provides only numbers 1 Enter the parameter value using the keypad or change the currently used parame ter value by using the rotary knob small steps or the UPARROW or DNARROW keys...

Страница 80: ...Table 3 3 5 To change from capital to small letters and vice versa press the sign key 6 When you have chosen the desired value wait for 2 seconds to use the same key again or start the next entry by...

Страница 81: ...Navigating in Dialog Boxes Some of the dialog boxes are not only for parameter entry and therefore have a more complex structure The following figure shows an example Changing the focus The focus on...

Страница 82: ...therwise you must switch to edit mode manually Switching to edit mode 1 Press the ENTER key 2 To quit the edit mode press the ESC key Working in dialog boxes To edit alphanumeric parameters use the ke...

Страница 83: ...mode is not active change into the edit mode Scroll through the list using the arrow keys or the rotary knob until the list entry you want to select is highlighted Confirm your choice by pressing the...

Страница 84: ...ween a split screen and a full screen maximized display In maximized display the table or diagram cur rently focused is maximized In split display both the diagram and any available tables are display...

Страница 85: ...ont Panel State On The extended display appears on the screen Using the F6 key you can toggle the Front Panel display on and off Remote SYST DISP FPAN STAT ON Working with the Soft Front Panel Basic o...

Страница 86: ...ress the DISPLAY key 2 In the Display Settings dialog box select Mini Front Panel State On The Mini Front Panel window appears on the screen It can be moved anywhere on the screen where it does not in...

Страница 87: ...ay on page 81 To do so drag the lower right corner of the win dow to the required size In standard display a small icon is displayed in the lower right corner of the screen 3 4 4 5 Displaying the Tool...

Страница 88: ...help To zoom into the diagram 1 Display the toolbar as described in Chapter 3 4 4 5 Displaying the Toolbar on page 84 2 Click on the Zoom On icon in the toolbar A dotted rectangle is displayed in the...

Страница 89: ...ction mode icon in the toolbar To return to original display Click on the Zoom Off icon in the toolbar A dotted rectangle is displayed in the diagram Remote commands 1 Activate the zoom mode DISP ZOOM...

Страница 90: ...beneath the diagram or removed from display 3 4 4 9 Selecting a Theme You can select a theme for the screen display The theme defines the colors used for keys and screen elements for example The defa...

Страница 91: ...date rate 1 Press the DISPLAY key The Display Settings dialog box is opened 2 Under Display Update Rate select Slow The display is updated less frequently and performance for measurements should impro...

Страница 92: ...ut of the signal generator to the RF input of R S ESRP Table 3 4 Signal generator settings e g R S SMU Frequency 128 MHz Level 30 dBm 3 5 1 1 Measuring the Level and Frequency Using Markers The level...

Страница 93: ...r the Peak softkey in the MKR menu in order to set the currently active marker to the maximum of the displayed signal Increasing the Frequency Resolution The frequency resolution of the marker is pred...

Страница 94: ...play dynamic range has increased Setting the marker level equal to the reference level The marker can also be used to shift the maximum value of the trace directly to the top edge of the diagram If t...

Страница 95: ...MHz 4 In the Setup menu select Reference Ext 5 Activate the marker by pressing the MKR key The marker is activated and set to the signal maximum The level and the fre quency of the marker are displaye...

Страница 96: ...ode 3 Set the start frequency to 100 MHz and the stop frequency to 400 MHz a Press the FREQ key b Press the Start softkey and enter 100 MHz c Press the Stop softkey and enter 400 MHz The R S ESRP disp...

Страница 97: ...D3 1 show the offset of the first and second harmonics from the fundamental Reducing Noise The signal analyzer offers three methods to differentiate the harmonics of a signal from the noise effectivel...

Страница 98: ...he measurement will take significantly more time The video band width that is displayed is marked with a bullet to indicate that it is no longer coupled to the resolution bandwidth see Figure 3 10 Fig...

Страница 99: ...ftkey in the BW menu 3 5 3 Measuring Signal Spectra with Multiple Signals 3 5 3 1 Separating Signals by Selecting the Resolution Bandwidth A basic feature of a Signal and Spectrum Analyzer is the abil...

Страница 100: ...y span to 300 kHz a Press the FREQ key and enter 128 015 MHz b Press the SPAN key and enter 300 kHz 4 Set the resolution bandwidth to 30 kHz and the video bandwidth to 1 kHz a Press the BW key b Press...

Страница 101: ...l drop is located exactly in the center of the screen only if the generator frequencies match the frequency display of the R S ESRP exactly To achieve exact matching the frequencies of the generators...

Страница 102: ...again by turning the rotary knob counterclockwise thus yielding a higher frequency resolution 6 Set the resolution bandwidth to 1 kHz To do so in the bandwidth menu press the Res BW Manual softkey an...

Страница 103: ...0 In the frequency range display the AM side bands can be resolved with a narrow bandwidth and measured separately The modulation depth of a carrier modulated with a sinusoidal signal can then be mea...

Страница 104: ...l in the center of the diagram and one delta marker each to the upper and lower AM side bands The R S ESRP calculates the AM modulation depth from the level dif ferences of the delta markers to the ma...

Страница 105: ...MPT key and enter 6 dBm b Press the Range softkey c Press the Range Linear softkey 6 Set triggering in response to the AF signal by using the video trigger to produce a static image a Press the TRIG k...

Страница 106: ...ffers easy to use functions that mea sure the power over a predefined time Measuring the Power of a GSM Burst During the Activation Phase Test setup Table 3 8 Signal generator settings e g R S SMU Fre...

Страница 107: ...trace The trigger level is displayed as a horizontal line labeled with the abso lute level for the trigger threshold in the measurement diagram 7 Configure power measurement in zero span a Press the M...

Страница 108: ...Hz display range the edges of TDMA bursts can be measured precisely The edges can be shifted to the screen area by using the trigger offset Test setup Table 3 9 Signal generator settings e g R S SMU F...

Страница 109: ...ey b Press the Trigger Offset softkey c By turning the rotary knob counterclockwise move the trigger offset until the burst edge can be seen in the center of the screen or enter 50 s The R S ESRP disp...

Страница 110: ...measured by comparing the power values during the activa tion phase and the deactivation phase of the transmission burst For this purpose the R S ESRP provides the function for measuring absolute and...

Страница 111: ...S ESRP shows the GSM burst continuously across the display 6 Use the trigger source Video and the trigger polarity Pos to trigger on the rising edge of the burst and shift the start of burst to the c...

Страница 112: ...vertical line to the end of the burst The R S ESRP displays the power during the activation phase of the burst Figure 3 18 Power measurement during the activation phase of the burst 8 Measure the powe...

Страница 113: ...ated signals The voltage at the output of the envelope detector remains constant for FM modulated signals as long as the fre quency deviation of the signal is located within the flat part of the passb...

Страница 114: ...127 50 MHz and the span to 300 kHz a Press the FREQ key and enter 127 50 MHz b Press the SPAN key and enter 300 kHz 4 Set the resolution bandwidth to 300 kHz a Press the BW key b Press the Res BW Manu...

Страница 115: ...be verified using the marker and delta marker Figure 3 20 Display of the filter edge of the 300 kHz filter as an FM discriminator character istic 6 Set the FM deviation to 50 kHz on the signal generat...

Страница 116: ...the peak of the curve b Press the Marker 2 softkey c Press the MKR key d Press the More softkey e Press the Min softkey Marker 2 delta marker is placed on the minimum of the curve The level differ enc...

Страница 117: ...strument configurations is displayed The File Name field is in edit mode and contains a suggestion for a new name 3 To change the suggested name enter a name for the settings file to be stored The nam...

Страница 118: ...via the keyboard or key pad Select the file from the selection list using the touchscreen or mouse Alternatively a Press the Select File softkey The focus is set on the files list b Using the rotary...

Страница 119: ...wing procedure Be aware that the chosen settings file is loaded both while booting and presetting 1 Press the SAVE RCL key 2 Press the Startup Recall softkey 3 Press the Startup Recall On Off softkey...

Страница 120: ...e R S ESRP The R S ESRP also provides function for IF analysis if you have equipped your R S ESRP with firmware application R S ESRP K56 IF analysis is not a separate measurement mode but is integrate...

Страница 121: ...zer Mode The I Q Analyzer mode provides measurement and display functions for digital I Q sig nals For more information on functionality available for the I Q Analyzer see Chapter 7 I Q Analyzer on pa...

Страница 122: ...igh interferer level The final measurement may be either automatic or in interactive mode Automatic control functions for line impedance stabilization networks LISN are available Time domain scans opt...

Страница 123: ...cordingly The display keeps the overall maximum level even after changing the frequency until a reset of the maxhold function 1 maximum levels note that the maximum and quasipeak peaks have been measu...

Страница 124: ...s unweighted It is independent of the selected detector for the bargraph measurement e g average or quasi peak A maximum of three traces can be displayed in parallel The display mode Clear Write Max H...

Страница 125: ...f the spectrum at the receiver input are measured in parallel by using fast fourier transform FFT of fre quency sections Time Domain scans require option R S ESRP K53 Stepped Scans in the Frequency Do...

Страница 126: ...with prescan the prescan is used to get a detailed overview of the emission spectrum The prescan result is then analyzed and the critical frequencies can be determined On the critical frequencies if f...

Страница 127: ...plays an exclamation mark if the selected measurement bandwidth is not compatible to the bandwidth required by CISPR for the corre sponding frequency range Meas Time Measurement time applied to the sc...

Страница 128: ...ritical frequencies Several data reduction methods are used Generating subrange maxima search method Subranges The whole frequency range is divided into equidistant frequency subranges A selectable nu...

Страница 129: ...asurement is performed after data reduction thus reducing the overall mea surement time The final measurement analyzes only the data that still remains after the preliminary measurement stages in othe...

Страница 130: ...n of anomalies and interfering signals The horizontal axis represents the frequency span The vertical axis represents time Time in the spectrogram runs chronologically from top to bottom Therefore the...

Страница 131: ...with IF analysis a single sweep may consist of several frames The exact number of frames depends on the measure ment time of the bargraph In the default state a frame is added to the spectrogram after...

Страница 132: ...e upper end of the color range it is always the light est color possible regardless of differences in amplitude e g black and blue in case of the Cold scheme Defining the Range of the Color Map The cu...

Страница 133: ...e range as far as possible is also a good way if you want to observe and put the focus on signals with a certain amplitude only Then only those signal amplitudes that you really want see are displayed...

Страница 134: ...nge from 105 5 dBm to 60 dBm blue green and yellow In the color map based on the linear color curve the same range is covered by blue and a few shades of green only The range from 60 dBm to 20 dBm on...

Страница 135: ...urements in the time domain a single measurement lasts until the defined measurement time has passed RUN CONT starts a continuous scan or measurement A continuous measurement lasts until it is interru...

Страница 136: ...is result display on and off Scan Turns the Scan result display on and off Spectrogram Turns the Spectrogram result display on and off 5 1 7 3 Bargraph Control The Measurement Configuration menu conta...

Страница 137: ...iately The scan stops at the frequency at which it was interrupted until it is con tinued While the scan is stopped the receiver settings can be changed e g for a detailed analysis of the recorded tra...

Страница 138: ...was interrupted the scan continues at the frequency at which it was interrupted The scan is continued with the settings in the scan table Remote command HOLD on page 609 SENSe FREQuency CENTer on page...

Страница 139: ...sure The final measurement starts or continues at the next entry of the peak list Abort the final measurement Stop Final Measurement The collected data is lost After all frequencies in the peak list h...

Страница 140: ...active mode Likewise it is possible to start measuring in interactive mode and later change into automatic mode Hold Final Measurement 137 Automatic Final 137 Interactive Final 137 Skip Frequency 137...

Страница 141: ...on page 609 5 1 7 6 Measurement Settings The Meas menu contains the basic measurement settings and the selection of result displays Chapter 5 1 7 2 Selecting the Result Display on page 132 Press the M...

Страница 142: ...automatically turned on Matches the scan trace number to the number of the bargraph Matches the color of the scan trace to a particular bargraph detector If you couple bargraph and scan trace the R S...

Страница 143: ...ly from one another The R S ESRP saves the settings accordingly Press the MEAS CONFIG key The R S ESRP opens the MeasConfig menu Clear Spectrogram 140 History Depth 140 Color Mapping 140 Trace to Spec...

Страница 144: ...ut to define the range of the color map 8 Color curve numerical input to define the shape of the color curve 9 Color scheme selection 10 Auto button automatically sets the value range of the color map...

Страница 145: ...nd automotive standards The resolution filters are implemented as digital Gaussian bandpass filters Concerning the attenuation characteristic the filters behave like analog filters but their measure m...

Страница 146: ...You can use several detectors at the same time The combined use of several detec tors multiple detection is important for EMI measurements This is due to specifica tions in the standards for example...

Страница 147: ...g a Time Domain Scan RMS detector The RMS detector evaluates the root mean square RMS value over the specified measurement time and displays the resulting value The integration time is the speci fied...

Страница 148: ...e the relatively long time constants used with quasipeak detectors entail long measurement times to obtain correct results With unknown signals the measurement time should be at least 1 s This ensures...

Страница 149: ...ger measurement times Measurement times shorter than 20 ms With measurement times shorter than 20 ms the detector weighting changes to plain average weighting When you change the receiver frequency or...

Страница 150: ...tive are not visible Each time the trace mode is changed the selected trace memory is cleared Note that in Spectrum mode the Max Hold and Min Hold modes are unavailable for statistics measurements Cle...

Страница 151: ...s indicated by the icon on the tab label If the level range or reference level is changed the R S ESRP automatically adapts the measured data to the changed display range This allows an amplitude zoom...

Страница 152: ...ENV4200 R S ENV432 Two line V networks R S ESH3 Z5 R S ENV216 For the R S ENV216 network a 150 kHz high pass filter is available for protection of the input After selecting the type of network you ca...

Страница 153: ...some control lines have to be routed through the wall of the shielded room You can also use a direct connection without a filter e g when you use the R S ESRP in a shielded room In that case you can...

Страница 154: ...unit or unit dB The frequency range covered by a set can be subdivi ded into max 10 subranges each with up to 4 transducer factors which follow each other without a gap i e the stop frequency of a sub...

Страница 155: ...general information about the measurement and instrument settings and characteristics It consists of three columns separated by a semicolon parameter name numeric value unit The data section contains...

Страница 156: ...1 6150000 000000 84 210000 Trace Frequency Level DeltaLi mit Phase Unused etc 5 2 9 Formats for Returned Values ASCII Format and Binary Format ASCII Format FORMat ASCII The command reads out a list of...

Страница 157: ...s The schema of the result string is as follows NoOfDataBytes value1 value2 value n with Header prefix 1 byte 1 byte NoOfDataBytes number of following data bytes 1024 in the example 10 bytes 1 byte Va...

Страница 158: ...equency When you change the frequency with the rotary knob the R S ESRP increases or decreases the 4th digit of the receiver frequency When you change the frequency with the cursor keys the R S ESRP i...

Страница 159: ...command SENSe FREQuency SPAN on page 793 Full Span Restores the span to the full available frequency range The full span is specified in the data sheet In receiver mode full span is available for IF A...

Страница 160: ...turn 10 dB Min off Note The maximum mixer level allowed is 0 dBm Mixer levels above this value may lead to incorrect measurement results which are indicated by the OVLD status dis play The increased...

Страница 161: ...nput does not exceed the allowed limits Exceeding the limit might damage the input mixer Do not use a 0 dB attenuation under any circumstances when you measure RFI volt age or unknown signals in combi...

Страница 162: ...ng another unit replace dB V with the corresponding unit The conversion is also possible when a transducer defines the used unit Remote command CALCulate n UNIT POWer on page 636 Grid Range Grid Min L...

Страница 163: ...field to define the measurement or resolution bandwidth The R S ESRP supports a selected set of resolution bandwidths If you enter a num ber that is not supported the R S ESRP rounds the value up to n...

Страница 164: ...e resolution bandwidth for IF spectrum analysis Available for IF Analysis firmware application R S ESRP K56 Remote command SENSe BANDwidth IF on page 639 5 3 4 Configuring the Scan The following setti...

Страница 165: ...started immediately Remote command TRIGger n SEQuence SOURce on page 807 Video Selects the video trigger The trigger event is a certain voltage level For the video triggering mode a level line showing...

Страница 166: ...the required bit pattern by changing the state of the individual ports Port x See AUX PORT on page 35 for information about the pin assignment of the con nector See OUTPut UPORt VALue on page 901 for...

Страница 167: ...nce scan peak search and final measurement The item opens the Scan Table tab The item includes blue state or excludes grey state the peak search and later stages from the test sequence The item opens...

Страница 168: ...s the configuration the scan is based on Scan Table The scan is performed based on the settings defineds of the scan table Current The scan is performed based on the current setting of the R S ESRP Re...

Страница 169: ...LIN Linear frequency steps The stepsize is fix and depends on the Stepsize LOG Logarithmic frequency steps The stepsize is a percentage of the current frequency Remote command SENSe SWEep SPACing on p...

Страница 170: ...s For each range you can customize the following parameters Range Start Range Stop Frequency Stepsize Resolution Bandwidth Meas Time Auto Ranging RF Attenuation Preamplifier For more information see C...

Страница 171: ...PSEarch METHod on page 649 No Of Peaks Defines the number of peaks the R S ESRP looks for during a peak search The range is from 1 to 500 peaks The number of peaks only takes effect if the peak searc...

Страница 172: ...margin relative to a limit line that is considered during a peak search For more information see Chapter 5 1 4 Peak List and Data Reduction on page 124 Remote command CALCulate n PEAKsearch PSEarch M...

Страница 173: ...ws the signal level of the peak The unit depends on the one you have selected Delta Limit Shows the distance of the peak to a limit line The delta limit is only calculated if you have activated a limi...

Страница 174: ...e unit The results are split into several data sections one for each active trace The data section begins with the entry Trace x Final followed by the trace char actersistics and the peak list data it...

Страница 175: ...u can select a detector for the scan and the final mea surement For more information see Chapter 5 1 3 Scans on page 121 Chapter 5 1 5 Final Measurement on page 126 Chapter 5 2 3 Trace Modes on page 1...

Страница 176: ...ge 654 Prescan Phase Selects the phase of the network you want to control during the scan During the scan you can control more than one phase at a time For more information see Chapter 5 2 5 V Network...

Страница 177: ...al axis frequency or time If more measured values than measurement points are available several measured values are combined in one displayed measurement point The trace functions include the followin...

Страница 178: ...821 Selecting the detector for the final measurement SENSe DETector t FMEasurement on page 611 More Traces Opens a submenu to select one of the traces not currently displayed in the main menu Copy Tra...

Страница 179: ...at frame was recorded For large history buffers the export operation may take some time Remote command FORMat DEXPort TRACes on page 615 FORMat DEXPort DSEParator on page 818 MMEMory STORe n TRACe on...

Страница 180: ...out Otherwise the edit dialog box for the last activated marker is opened and the current frequency time value is displayed Further information Displayed Marker Information on page 381 Chapter 5 4 2...

Страница 181: ...ch the marker is to be placed Remote command CALCulate n MARKer m TRACe on page 664 CALCulate n DELTamarker m TRACe on page 670 Marker Wizard Opens a configuration dialog for markers The marker wizard...

Страница 182: ...ll Marker Off Switches all markers off It also switches off all functions and displays that are associ ated with the markers delta markers Remote command CALCulate n MARKer m AOFF on page 660 Marker T...

Страница 183: ...scan range the marker frequency is currently in Remote command CALCulate n MARKer m SCOupled STATe on page 657 5 4 2 2 Positioning Markers The MKR key is used for search functions of measurement marke...

Страница 184: ...e selected trace Remote command CALCulate n MARKer m MAXimum NEXT on page 660 CALCulate n DELTamarker m MAXimum NEXT on page 667 Min Sets the active marker delta marker to the minimum of the selected...

Страница 185: ...nu to set the limits for maximum or minimum search in the x and y direction Remote command CALCulate n MARKer m X SLIMits STATe on page 659 Left Limit Search Limits Opens an edit dialog box to enter a...

Страница 186: ...B to 80 dB are allowed the resolution is 0 1 dB The default setting for the peak excursion is 6 dB Remote command CALCulate n MARKer m PEXCursion on page 647 5 4 2 3 Markers in Spectrograms In additio...

Страница 187: ...forms a search in the currently selected frame Y Search Performs a search on the marker frequency span 0 or time span 0 over all captured frames XY Search Performs a search in the frequency domain spa...

Страница 188: ...a New Limit Line Based upon an Existing Limit Line on page 195 Chapter 5 4 3 9 Activating Deactivating a Limit Line on page 196 5 4 3 1 Softkeys of the Lines Menu The following table shows all softke...

Страница 189: ...or all assigned traces For details see also Chap ter 5 4 3 9 Activating Deactivating a Limit Line on page 196 Remote command CALCulate n LIMit k STATe on page 687 New Opens the Edit Limit Line dialog...

Страница 190: ...order rule an error message is displayed and the values are discarded Delete Value New Deletes the selected data point x and y value All succeeding data points are shifted up accordingly This softkey...

Страница 191: ...ny effect on limit lines that represent absolute val ues for the y axis Remote command CALCulate n LIMit k LOWer OFFSet on page 681 CALCulate n LIMit k UPPer OFFSet on page 684 Display Lines Opens a s...

Страница 192: ...Lines Turns a frequency line coupled to the receiver frequency on and off The line that marks the receiver frequency is a green line labeled TF Available for the Scan result display Remote command CAL...

Страница 193: ...instrument can be activated simultaneously The number of limit lines stored in the instrument is only limited by the capacity of the flash disk used Which softkeys are available depends on the displa...

Страница 194: ...tched on softkey with highlighted background 4 When you press the Display Line 1 Display Line 2 softkey for the second time the edit dialog box for the line is opened again 5 When you press the Displa...

Страница 195: ...lines refer also to Chapter 5 4 3 3 Limit Lines Frequency Time Lines on page 190 The following set tings can be defined Setting Description Name The name under which the limit line is to be stored in...

Страница 196: ...zero span or if masks for modulated signals are required for span 0 Hz Scale Linear Logarithmic Y Axis Scale unit Unit of the y axis Scale mode Absolute The limit values refer to absolute levels or vo...

Страница 197: ...s a lower limit the margin is above the limit line Entering a new data point 1 Press the Insert value button in the dialog or select an existing data point in the table and press the Insert Value soft...

Страница 198: ...ftkey 5 4 3 8 Creating a New Limit Line Based upon an Existing Limit Line 1 In the Select Limit Line dialog box select the limit line you want to use as a basis for a new limit line For details see al...

Страница 199: ...has to consist of 2 or more data points 1 In the Select Limit Line dialog box select the limit line you want to activate deacti vate For details see also Chapter 5 4 3 5 Selecting a Limit Line on pag...

Страница 200: ...asurement parameters The general measure ment parameters apply to all measurements performed in Spectrum mode 6 1 Measurements In the Spectrum mode the R S ESRP provides a variety of different measure...

Страница 201: ...r in zero span Time Domain Power softkey for details see Chapter 6 1 1 7 Measuring the Power in Zero Span on page 270 EMI Measurement softkey for details see Chapter 6 1 1 8 Performing EMI Mea suremen...

Страница 202: ...oise ratio measurement Measurements without C N and measurements with reference to the bandwidth C No are possible For details see Chapter 6 1 1 3 Measuring the Carrier to Noise Ratio on page 225 OBW...

Страница 203: ...and opens a submenu For details see Chapter 6 1 1 9 Calculating Signal Amplitude Statistics on page 282 Remote command CALCulate n STATistics APD STATe on page 760 CCDF Activates the function to measu...

Страница 204: ...annel power and adjacent channel power is one of the most important tasks in the field of digital transmission for a signal analyzer with the necessary test routines While theoretically channel power...

Страница 205: ...s a multi carrier ACLR measurement There are two possible methods for measuring channel and adjacent channel power with a signal analyzer IBW method Integration Bandwidth Method Zero span method Fast...

Страница 206: ...el power but obviously increases measurement time The integrated bandwidth method IBW calculates channel power and ACLR from the trace data obtained during a continuous sweep over the selected span Mo...

Страница 207: ...la tion in one channel is the product of sweeptime in channel times the selected resolu tion bandwidth Assuming a sweeptime of 100 ms there are 30 kHz 4 19 MHz 100 ms 10 kHz 7 samples Whereas in Fast...

Страница 208: ...ults is calculated from 100 consecu tive measurements Take scaling into account if comparing power values Figure 6 4 Repeatability of adjacent channel power measurements on CDMA2000 signals in the fas...

Страница 209: ...powers are measured depends on the number of configured channels see of Adj Chan on page 213 For each channel the following values are displayed Label Description Channel Channel name as specified in...

Страница 210: ...p other instrument settings such as the used filter bandwidths frequency span and detector and trace settings can be optimized automat ically see Adjust Settings on page 219 For an overview of the sof...

Страница 211: ...vels Changes to an existing standard can be stored as a user defined standard see User Defined Configurations on page 210 In the Ch Power menu press Channel Setup then press the Channel Setup soft key...

Страница 212: ...nnel spacing for the adjacent or an alternate channel is changed all higher alternate channel spacings are multiplied by the same factor new spacing value old spacing value The lower adjacent channel...

Страница 213: ...by setting Limit Checking to On Performing a Channel Power Measurement A channel power measurement is started automatically according to the currently selected standard when you press the Ch Power AC...

Страница 214: ...nd adjacent channel spacing see Spacing on page 215 Channel bandwidth of transmission Tx adjacent Adj and alternate Alt channels see Bandwidth on page 214 Resolution bandwidth see Res BW Auto on page...

Страница 215: ...ased to ensure that adjacent channel powers are measured with adequate accu racy For general information on performing channel or adjacent channel power measure ments see Chapter 6 1 1 2 Measuring Cha...

Страница 216: ...annel power measurement Values from 0 to 12 are allowed The following measurements are performed depending on the number of the channels 0 Only the channel powers are measured 1 The channel powers and...

Страница 217: ...channel bandwidth is normally defined by the transmission standard The correct bandwidth is set automatically for the selected standard see Optimized Settings for CP ACLR Test Parameters on page 223 M...

Страница 218: ...nsmission channel with the lowest power is used as a refer ence channel Max Power TX Channel The transmission channel with the highest power is used as a refer ence channel Lowest Highest Channel The...

Страница 219: ...her adjacent channels is multiplied by factor 2 to result in 80 kHz ALT1 120 kHz ALT2 160 kHz ALT3 If starting from the default setting the spacing of the 5th adjacent channel ALT4 is set to 150 kHz t...

Страница 220: ...el to 0 35 POW ACH FILT ALT1 ON see SENSe POWer ACHannel FILTer STATe ALTernate channel on page 703 Activates the alpha value for the weighting filter for the alternate channel 1 POW ACH FILT ALPH ALT...

Страница 221: ...tivate or deactivate the limit to be considered during a limit check The check of both limit values can be activated independently Chan Pwr Hz CP ACLR Settings Ch Power ACLR If deactivated the channel...

Страница 222: ...tive power measurement in the adjacent channels Abs The absolute power in the adjacent channels is displayed in the unit of the y axis e g in dBm dB V Rel The level of the adjacent channels is display...

Страница 223: ...ated measured values are required for a reproducibility of 0 5 dB 99 of the measurements are within 0 5 dB of the true measured value This holds true for white noise The measured values are considered...

Страница 224: ...standard you can customize the following parameters number of adjacent channels channel bandwidth of transmission Tx adjacent Adj and alternate Alt channels channel spacings resolution and video band...

Страница 225: ...SRP or limiting the dynamic range by a too small S N ratio For details on manual settings see Optimized Settings for CP ACLR Test Parameters on page 223 The reference level is not influenced by the se...

Страница 226: ...BRO WIBRO WIBRO GSM GSM GSM RFID 14443 RFID 14443 RFID14443 TETRA TETRA TETRA PDC PDC PDC PHS PHS PHS CDPD CDPD CDPD APCO 25 Phase 2 APCO 25 P2 PAPCo25 For the R S ESRP the channel spacing is defined...

Страница 227: ...ceptable measurement speed and required selection to sup press spectral components outside the channel to be measured especially of the adjacent channels the resolution bandwidth must not be selected...

Страница 228: ...ches off this function Averaging which is often performed to stabilize the measurement results leads to a too low level indication and should therefore be avoided The reduction in the displayed power...

Страница 229: ...surement has been completed The frequency span should be set to approximately 4 times the channel bandwidth in order to measure the carrier to noise ratio correctly This setting is defined automati ca...

Страница 230: ...hes the measurement of the carrier noise ratio on or off If no marker is active marker 1 is activated The measurement is performed on the trace that marker 1 is assigned to To shift marker 1 and measu...

Страница 231: ...dulated signal is its occupied bandwidth In a radio communications system for instance the occupied bandwidth must be limited to enable distortion free transmission in adjacent channels The occupied b...

Страница 232: ...e new range The OBW measurement uses the same search limits as defined for marker search see Search Limits on page 182 However only the left and right limits are consid ered To ensure correct power me...

Страница 233: ...andwidth Occ BW is indicated in the marker results Furthermore the marker at the center frequency and the tempo rary markers are indicated The OBW calculation is repeated if the Search Limits are chan...

Страница 234: ...transmission standard the bandwidth speci fied by the standard for the transmission channel must be entered Remote command SENSe POWer ACHannel BANDwidth BWIDth CHANnel channel on page 699 Adjust Ref...

Страница 235: ...onding softkeys Spectrum Emission Mask on page 234 If you want a parameter set to be available permanently you can create an XML file for this configuration and if necessary export this file to anothe...

Страница 236: ...Level 236 RF Att Mode 236 RF Attenuator 236 Preamp 236 Transd Factor 236 Limit Check 1 4 237 Abs Limit Start 237 Abs Limit Stop 237 Rel Limit Start 237 Rel Limit Stop 237 Close Sweep List 238 Insert...

Страница 237: ...rum Emission Masks on page 232 Remote command SENS SWE MODE ESP see SENSe SWEep MODE on page 729 Sweep List Spectrum Emission Mask Opens a submenu to edit the sweep list and displays the Sweep List di...

Страница 238: ...the defined span and have a minimum span of 20 Hz The first and last ranges are adapted to the given span as long as the minimum span of 20 Hz is not violated Frequency values for each range have to...

Страница 239: ...e range RLEVel on page 737 RF Att Mode Sweep List dialog box Sweep List Spectrum Emission Mask Activates or deactivates the auto mode for RF attenuation Remote command SENSe ESPectrum RANGe range INPu...

Страница 240: ...ectrum Emission Mask Sets an absolute limit value at the start frequency of the range dBm This parameter is only available if the limit check is set accordingly see Limit Check 1 4 on page 237 Remote...

Страница 241: ...page 245 Remote command ESP RANG1 INS AFT see SENSe ESPectrum RANGe range INSert on page 735 Delete Range Sweep List Spectrum Emission Mask Deletes the currently focused range if possible The range nu...

Страница 242: ...the reference range integral bandwidth method If the Channel Power reference power type is activated the dialog box is exten ded to define additional settings Tx Bandwidth Defines the bandwidth used...

Страница 243: ...also to the ASCII File Export soft key ASCII File Export on page 240 Remote command MMEMory STORe n LIST on page 745 ASCII File Export Save Evaluation List List Evaluation Spectrum Emis sion Mask Ope...

Страница 244: ...ighest peak within the reference range Channel Power Measures the channel power within the reference range integral bandwidth method If the Channel Power reference power type is activated the dialog b...

Страница 245: ...power in the reference range is used The limits assigned to that power class are applied see PMin PMax on page 242 Remote command CALCulate n LIMit k ESPectrum PCLass Class EXCLusive on page 742 To de...

Страница 246: ...ement on page 248 Remote command SENSe ESPectrum PRESet STANdard on page 728 Save As Standard Spectrum Emission Mask Opens the Save As Standard dialog box in which the currently used SEM settings and...

Страница 247: ...le P 31 is indicated if the lowest power class is defined from infinity to 31 and the power is currently 17 dBm In addition to the graphical results of the SEM measurement displayed in the diagram a r...

Страница 248: ...he peaks in the list evaluation detected peaks can be indicated by blue squares in the diagram Show Peaks Furthermore you can save the evaluation list to a file Save Evaluation List Retrieving Results...

Страница 249: ...f the first range or the stop frequency of the last range select the appropriate span with the SPAN key If you set a span that is smaller than the overall span of the ranges the measurement includes o...

Страница 250: ...3 _SEM_LINE_REL 0 3 are automatically defined for each power class according to the current Sweep List settings every time the settings change The limit line defined for the current power class is ind...

Страница 251: ...xceed the limit are indicated in red and by an asterisk Although a margin functionality is not available for the limit check a margin threshold for the peak values to be displayed in the evaluation li...

Страница 252: ...dBm DL WIBRO UL PowerClass_23_INF xml WiBro TTA 23 INF dBm UL PowerClass_negINF_23 xml WiBro TTA 23 INF dBm UL WIMAX DL ETSI MHz 1 75 MHz 2 00 MHz 3 5 MHz 7 00 MHz 14 00 MHz 28 MHz System_Type_E xml...

Страница 253: ...goryA BW_15_0_MHz__CFlower1GHz xml LTE Cat A 1GHz DL EUTRA LTE DL CategoryA BW_20_0_MHz__CFhigher1GHz xm l LTE Cat A 1GHz DL EUTRA LTE DL CategoryA BW_20_0_MHz__CFlower1GHz xml LTE Cat A 1GHz DL EUTRA...

Страница 254: ...ccording to these conventions because the R S ESRP can only interpret XML files of a known structure For example files look in the C r_s instr sem_std directory Figure 6 5 Example Spectrum emission ma...

Страница 255: ...ESRP uses the same ranges in each power class Therefore the contents of the ranges of each defined power class have to be identical to the first power class An exception are the Start and Stop values...

Страница 256: ...Index n StartPower Unit dBm InclusiveFlag true Value StartPowerValue StopPower Unit dBm InclusiveFlag false Value StopPowerValue DefaultLimitFailMode Limit Fail Mode DefaultLimitFailMode Range Index n...

Страница 257: ...andard Yes Instrument Type FSL Name of the instrument No Application SA K72 K82 Name of the application No LinkDirection Name Downlink Uplink None Yes ShortName DL UL No Reference Power Yes Method TX...

Страница 258: ...ightingFilter Only if Referen cePower method is TX Channel Power and the range is the ref erence range Type RRC CFilter Type of the weighting filter Yes Roll Off Factor 0 1 Excess bandwidth of the fil...

Страница 259: ...dentical in all ranges No Sweep Mode Manual Auto Sweep Time Mode on page 236 Yes Time time in sec Sweep Time on page 236 No Amplitude No ReferenceLevel Value power in dBm Ref Level on page 236 Yes if...

Страница 260: ...ine in dB positive value means above the limit limit fail pass 0 fail 1 Fast Spectrum Emission Mask Measurements In order to improve the performance of the R S ESRP for spectrum emission mask measurem...

Страница 261: ...When the Fast SEM mode is activated the ranges for which these criteria apply are displayed as one single range The sweep time is defined as the sum of the individual sweep times initially but can be...

Страница 262: ...s The Spurious Emissions measurement defines a measurement that monitors unwan ted RF products outside the assigned frequency band generated by an amplifier The spurious emissions are usually measured...

Страница 263: ...to ranges The minimum span of a range is 20 Hz The individual ranges must not overlap but need not directly follow one another The maximum number of ranges is 20 The maximum number of sweep points in...

Страница 264: ...on page 262 Which data is dis played in the evaluation list can be defined in the List Evaluation menu see List Evaluation on page 268 The following information is provided in the evaluation list Colu...

Страница 265: ...possible to define limit lines for any measurement in Spectrum mode using the Lines key For Spurious measurements however a special limit line is available via the Sweep List and it is strongly recom...

Страница 266: ...e limit are indicated in red and by an asterisk Although a margin functionality is not available for the limit check a margin threshold for the peak values to be displayed in the evaluation list can b...

Страница 267: ...h 200 dB so that initially all peaks are detected In the Spurious menu press List Evaluation Margin and enter a value in dB Only delta values that exceed this margin are displayed in the result table...

Страница 268: ...ions regarding spurious emissions configuration see Chap ter 6 1 1 6 Measuring Spurious Emissions on page 259 Range Start Range Stop Sweep List dialog box Sweep List Spurious Emissions Sets the start...

Страница 269: ...the range Remote command SENSe LIST RANGe range SWEep TIME on page 725 Detector Sweep List dialog box Sweep List Spurious Emissions Sets the detector for the range For details refer to Chapter 6 3 1 5...

Страница 270: ...als the span of the range The x axis is linear The unit is dB Remote command SENSe LIST RANGe range TRANsducer on page 725 Limit Check 1 4 Sweep List dialog box Sweep List Spurious Emissions Sets the...

Страница 271: ...to Ranges and Range Settings on page 245 Remote command SENSe LIST RANGe range DELete on page 720 Adjust Ref Lvl span 0 Sweep List Spurious Emissions Adjusts the reference level to the measured total...

Страница 272: ...s Opens an edit dialog box to enter the number of peaks per range that are stored in the list Once the selected number of peaks has been reached the peak search is stopped in the current range and con...

Страница 273: ...level If the measured channel power is equal to the reference level the signal path cannot be overloaded Remote command SENSe ADJust LEVel on page 798 Adjust X Axis Spurious Emissions Adjusts the fre...

Страница 274: ...to be evaluated for the power measurement can be restricted using limit lines The left and right limit lines S1 S2 define the evaluation range and are indicated by vertical red lines in the diagram If...

Страница 275: ...e or a segment thereof The linear mean value of the equivalent voltages is calculated This can be used for instance to measure the mean power during a GSM burst Remote command CALCulate n MARKer m FUN...

Страница 276: ...t antenna or the rotation of the DUT Covering all test setups with one of the slow EMI weighting detectors over the required frequency range would lead to unacceptable measurement times Splitting the...

Страница 277: ...idth should cover at least one sweep point more is better If this condition is not met signals or interferences could be missed during final evaluation of narrowband interferers Example Linear axis In...

Страница 278: ...the number of markers you have turned on You can use up to 16 markers simultaneously The first marker is always on the most powerful peak while the last marker is always on the least powerful peak If...

Страница 279: ...d in a standard You can select and assign detec tors for EMI markers in the Marker Configuration on page 279 dialog box As EMC tests often require special measurement times you can also specify a dwel...

Страница 280: ...If you select the bandwidth with the Res BW CISPR softkey the R S ESRP automat ically changes the filter type to a 6 dB bandwidth Remote command Filter type SENSe BANDwidth BWIDth RESolution TYPE on...

Страница 281: ...Control Opens a dialog box to control a LISN The dialog box contains the following elements LISN Selects the V network that you have in use The R S FSV K54 supports the follow ing networks R S ESH3 Z5...

Страница 282: ...alog box that contains all marker characteristics necessary to perform successful EMI measurements The dialog box is made up out of two tabs The first tab controls markers 1 to 8 the second tab contro...

Страница 283: ...Auto Peak Turns automatic peak search for all markers on and off For more information see Performing a Peak Search on page 275 Dwell Time Sets the dwell time for all markers For more information see...

Страница 284: ...mark ers this is a absolute value for delta markers this is a relative value The corre sponding reference marker for delta markers is indicated in the Ref column Level Signal level at the marker posi...

Страница 285: ...t LIMit k CONDition on page 749 Result CALCulate n MARKer m FUNCtion FMEasurement RESult on page 751 CALCulate n DELTamarker m FUNCtion FMEasurement RESult on page 750 6 1 1 9 Calculating Signal Ampli...

Страница 286: ...gnal is therefore an important transmitter design criterion The crest factor is defined as the peak power mean power ratio or logarithmically as the peak level minus the average level of the signal To...

Страница 287: ...Statistic measurements on pulsed signals can be performed using a gated trigger An external frame trigger is required as a time frame reference For details see Defining Gated Triggering for APD and C...

Страница 288: ...splayed trace Mean amplitude Peak amplitude Crest factor Complementary Cumulative Distribution Function CCDF The Complementary Cumulative Distribution Function CCDF shows the probability that the mean...

Страница 289: ...The displayed amplitude range is indicated as Mean Pwr x dB In addition to the histogram a result table is displayed containing the following informa tion Number of samples used for calculation For ea...

Страница 290: ...Activates the function to measure the amplitude probability density APD and opens a submenu For general information on calculating signal statistics see Chapter 6 1 1 9 Calculat ing Signal Amplitude S...

Страница 291: ...Level on page 323 For the APD function this value is mapped to the right diagram border For the CCDF function there is no direct representation of this value on the diagram as the x axis is scaled rel...

Страница 292: ...ge 636 Display range DISP WIND TRAC Y 5DB see DISPlay WINDow n TRACe t Y SCALe on page 812 Range Log 1 dB x Axis Range Scaling APD Sets the level display range to 1 dB Remote command Logarithmic scali...

Страница 293: ...displayed probability range Values on the y axis are normalized which means that the maximum value is 1 0 The y axis scaling is defined via the y Unit Abs softkey The distance between max and min valu...

Страница 294: ...e AUTO ONCE on page 765 Gated Trigger On Off APD Activates and deactivates the gating for statistics functions for the ACP and the CCDF channel The trigger source is changed to EXTERN if this function...

Страница 295: ...min imum power for APD measurement or peak and mean power for CCDF measurement in order to obtain maximum power resolution Adjusts the reference level to the current input signal For details see also...

Страница 296: ...e switched on automatically As all markers the percent marker can be moved simply by touching it with a finger or mouse cursor and dragging it to the desired position Remote command CALCulate n MARKer...

Страница 297: ...istics SCALe X RLEVel on page 766 x Axis Range Scaling CCDF Opens the Range submenu to select a value for the level range to be covered by the statistics measurement selected Remote command CALCulate...

Страница 298: ...C LOG see DISPlay WINDow n TRACe t Y SPACing on page 636 Display range DISP WIND TRAC Y 1DB see DISPlay WINDow n TRACe t Y SCALe on page 812 Range Log Manual x Axis Range Scaling CCDF Opens an edit di...

Страница 299: ...in value must be at least one decade Remote command CALCulate n STATistics SCALe Y UPPer on page 767 y Axis Min Value Scaling CCDF Opens an edit dialog box to define the lower limit of the displayed p...

Страница 300: ...he gate ranges are defined using the Gate Ranges on page 291 softkey Remote command SENSe SWEep EGATe on page 809 SENSe SWEep EGATe SOURce on page 810 Gate Ranges CCDF Opens a dialog to configure up t...

Страница 301: ...also the Adjust Ref Lvl softkey Remote command CALCulate n STATistics SCALe AUTO ONCE on page 765 Defining Gated Triggering for APD and CCDF Statistic measurements on pulsed signals can be performed u...

Страница 302: ...of the period to be analyzed in the Period field Note The period is the same for all traces If you change the period for one trace it is automatically changed for all traces Make sure the defined per...

Страница 303: ...smaller than the length of the period 8 To define further ranges for the same period in the same trace repeat steps 5 7 for the same trace To define further ranges for the same period in a different t...

Страница 304: ...r intercept point is measured using the TOI softkey see TOI on page 200 About TOI Measurement If several signals are applied to a transmission two port device with nonlinear charac teristic intermodul...

Страница 305: ...g the 2nd harmonic of useful signal PU1 and signal PU2 fi1 2 fu1 fu2 6 fi2 2 fu2 fu1 7 Dependency on level of useful signals The level of the intermodulation products depends on the level of the usefu...

Страница 306: ...g The intermodulation products of the third order increase three times as much as the useful signals The intercept point is the fictitious level where the two lines intersect It cannot be measured dir...

Страница 307: ...n the range that is required Since phase noise decreases as the carrier offset increases its influence decreases with increasing frequency offset from the useful signals The following diagrams illustr...

Страница 308: ...f the intermodulation products sep arately using a small span The measurement time will be reduced in particular if the offset of the useful signals is large To find signals reliably when frequency sp...

Страница 309: ...16 Marker Norm Delta TOI The Marker X softkey activates the corresponding marker and opens an edit dialog box to enter a value for the marker to be set to Pressing the softkey again deactivates the s...

Страница 310: ...er values is used for AM modulation depth calculation A remote control programming example is described in Chapter 11 17 2 3 Measuring the AM Modulation Depth on page 976 and an example of how to perf...

Страница 311: ...ers 2 to 16 are delta markers that refer to marker 1 These markers can be converted into markers with absolute value display using the Marker Norm Delta softkey If marker 1 is the active marker press...

Страница 312: ...the fundamental wave must be selec tively attenuated with respect to the other harmonics with a high pass filter Obtainable dynamic range When harmonics are being measured the obtainable dynamic rang...

Страница 313: ...ng to 2 is At a resolution bandwidth of 10 Hz noise level 143 dBm S H I 40 dBm the opti mum mixer level is 51 5 dBm According to 1 a maximum measurable harmonic dis tortion of 91 5 dB minus a minimum...

Страница 314: ...rements Harmonic Distortion 311 No of Harmonics 312 Harmonic Sweep Time 312 Harmonic RBW Auto 312 Adjust Settings 312 Harmonic Distortion Opens a submenu to determine the settings for harmonics measur...

Страница 315: ...frequency search in the frequency range that was set before starting the harmonic measurement if harmonic measurement was with span 0 and adjusts the level Remote command CALCulate n MARKer m FUNCtio...

Страница 316: ...Further information Chapter 6 2 1 2 Initial Configuration on page 314 Task Chapter 6 2 1 1 Presetting the Instrument on page 313 6 2 1 1 Presetting the Instrument 1 Define the data set for the presett...

Страница 317: ...and Recalling Settings Files SAVE RCL Key on page 500 Table 6 5 Initial configuration Parameter Setting mode Spectrum sweep mode auto center frequency fmax 2 center frequency step size 0 1 span span...

Страница 318: ...and Stop Frequency on page 319 Chapter 6 2 2 3 Specifying the Frequency Axis by the Center Frequency and the Span on page 319 Chapter 6 2 2 4 Specifying the Step Size for the Arrow Keys and the Rotary...

Страница 319: ...rotary knob the center frequency changes in steps of 10 of the Center Frequency Stepsize The step size can be set to a fraction of the span span 0 or a fraction of the resolu tion bandwidth span 0 or...

Страница 320: ...EQuency CENTer STEP LINK on page 791 x RBW span 0 CF Stepsize Opens an edit dialog box to set the step size for the center frequency as a percentage of the resolution bandwidth Values between 1 and 10...

Страница 321: ...to enter a frequency offset that shifts the displayed frequency range by the specified offset The softkey indicates the current frequency offset The allowed values range from 100 GHz to 100 GHz The d...

Страница 322: ...Select Trace span 0 Signal Track span 0 Opens an edit dialog box to select the trace on which the signal is tracked Remote command CALCulate n MARKer m FUNCtion STRack TRACe on page 795 6 2 2 2 Speci...

Страница 323: ...d value The step size of the rotary knob is always 10 of it 6 2 2 5 Modifying the Frequency Axis by an Offset Press the Frequency Offset softkey and enter the offset to shift the displayed fre quency...

Страница 324: ...st Span on page 156 Full Span on page 156 Span Manual 321 Zero Span 321 Span Manual Opens an edit dialog box to enter the frequency span The center frequency remains the same when you change the span...

Страница 325: ...cifying the Amplitude on page 327 6 2 4 1 Softkeys of the Amplitude Menu The following table shows all softkeys available in the Amplitude menu It is possible that your instrument configuration does n...

Страница 326: ...ay range of the level axis Range Log 100 dB Range Log 50 dB Range Log 10 dB Range Log 5 dB Range Log 1 dB Range Log Manual Range Linear Range Lin Unit Range Log 100 dB Range Sets the level display ran...

Страница 327: ...C LOG see DISPlay WINDow n TRACe t Y SPACing on page 636 Display range DISP WIND TRAC Y 1DB see DISPlay WINDow n TRACe t Y SCALe on page 812 Range Log Manual Range Opens an edit dialog box to define t...

Страница 328: ...played at a power of 1 mW dBm Via the known input impe dance 50 or 75 conversion to other units is possible The following units are available and directly convertible dBm dBmV dB V dB A dBpW Volt Ampe...

Страница 329: ...y WINDow n TRACe t Y SCALe MODE on page 812 Noise Correction If activated the results are corrected by the instrument s inherent noise which increa ses the dynamic range ON A reference measurement of...

Страница 330: ...using the Ref Level Position and or Grid Abs Rel softkeys see Ref Level Position on page 326 and Grid Abs Rel on page 326 4 Set the attenuation using the RF Atten Manual or RF Atten Auto see RF Atten...

Страница 331: ...surement is com pleted the original trigger source is restored The trigger level is adjusted as fol lows For IF Power and RF Power triggers Trigger Level Reference Level 15 dB For Video trigger Trigge...

Страница 332: ...surement is used to determine the optimal reference level automatically see the Auto Level softkey Auto Level on page 329 The default value is 1 ms Remote command SENSe ADJust CONFigure LEVel DURation...

Страница 333: ...le filters Channel filters or RRC filters In this case sweep mode is used Remote command SWE TYPE FFT see SENSe SWEep TYPE on page 803 Auto Sweep Type Automatically sets the fastest available Sweep Ty...

Страница 334: ...age 338 Table 6 6 Tasks Chapter 6 2 6 2 Specifying the Bandwidth on page 337 Chapter 6 2 6 3 Selecting the Appropriate Filter Type on page 337 6 2 6 1 Softkeys of the Bandwidth Menu The following tabl...

Страница 335: ...channel bar When measuring Spurious Emissions using this softkey automatically opens the Sweep List dialog see Sweep List dialog box on page 265 Remote command SENSe BANDwidth BWIDth RESolution AUTO...

Страница 336: ...eo bandwidths result in longer sweep times due to the lon ger settling time Wide bandwidths reduce the signal noise ratio Remote command SENSe BANDwidth BWIDth VIDeo AUTO on page 800 Sweeptime Manual...

Страница 337: ...ssible methods of sweep ing Sweep on page 330 FFT on page 330 not available with 5 Pole filters channel filters or RRC filters see Chapter 6 2 6 3 Selecting the Appropriate Filter Type on page 337 Aut...

Страница 338: ...ne 1 1 Coupling Ratio Sets the following coupling ratio video bandwidth resolution bandwidth This is the default setting for the coupling ratio resolution bandwidth video bandwidth This is the couplin...

Страница 339: ...pling Ratio Sets the following coupling ratio resolution bandwidth span 100 This coupling ratio is the default setting of the R S ESRP This setting takes effect if you define the resolution bandwidth...

Страница 340: ...or Video BW Auto softkey 3 Set the sweep time using the Sweeptime Manual or Sweeptime Auto softkey 4 Press the Filter Type softkey and select the appropriate filters 6 2 6 3 Selecting the Appropriate...

Страница 341: ...RRC and Channel Filters For power measurement a number of especially steep edged channel filters are availa ble see the following table The indicated filter bandwidth is the 3 dB bandwidth For RRC fil...

Страница 342: ...DC 24 3 kHz a 0 35 RRC IS 136 25 kHz CFILter APCO 25 P2 30 kHz CFILter CDPD CDMAone 50 kHz CFILter 100 kHz CFILter 150 kHz CFILter FM Radio 192 kHz CFILter PHS 200 kHz CFILter GSM 300 kHz CFILter 500...

Страница 343: ...6 2 7 2 Specifying the Sweep Settings on page 345 6 2 7 1 Softkeys of the Sweep Menu The following table shows all softkeys available in the Sweep menu It is possible that your instrument configurati...

Страница 344: ...configuration refer to Chapter 6 3 1 Trace Configuration on page 358 Remote command INITiate n CONMeas on page 801 Sweeptime Manual Opens an edit dialog box to enter the sweep time Sweep time absolut...

Страница 345: ...ble for the I Q analyzer In frequency sweep mode the analyzer provides several possible methods of sweep ing Sweep on page 330 FFT on page 330 not available with 5 Pole filters channel filters or RRC...

Страница 346: ...32767 are allowed If the values 0 or 1 are set one sweep is performed The sweep count is applied to all the traces in a diagram If the trace configurations Average Max Hold or Min Hold are set the swe...

Страница 347: ...ram measurements only Determines whether the results of the last measurement are deleted before starting a new measurement On Repeats the single sweep measurement without deleting the spectrogram resu...

Страница 348: ...points 5 Press the Continuous Sweep or Single Sweep softkey to select the sweep mode 6 Press the Continue Single Sweep softkey to repeat the single sweep 6 2 8 Triggering the Sweep TRIG Key The TRIG...

Страница 349: ...x to select the trigger gate mode As gate modes all modes except Power Sensor are available For details see also Chapter 6 2 8 3 Using Gated Sweep Operation on page 352 The default setting is Free Run...

Страница 350: ...in the following range 24dBm RF Att Triggerlevel 5dBm RF Att max 30 dBm for Preamp OFF 40dBm RF Att Triggerlevel 11dBm RF Att max 30 dBm for Preamp ON with 500 MHz InputSignal 7 GHz Note If input valu...

Страница 351: ...e command TRIG SOUR TIMETRIGger n SEQuence SOURce on page 807 Trg Gate Level Opens an edit dialog box to enter the trigger gate level For details see also Chapter 6 2 8 3 Using Gated Sweep Operation o...

Страница 352: ...Opens an edit dialog box to define a repetition interval in which the measurement is triggered The shortest interval is 2 ms This softkey is only available if the trigger source Time is selected see T...

Страница 353: ...ey is not available For details also see Chapter 6 2 8 3 Using Gated Sweep Operation on page 352 Gate Mode Lvl Edge Gate Settings Sets the gate mode As settings level triggered or edge triggered gate...

Страница 354: ...to obtain a higher resolu tion for positioning gate delay and gate length When leaving the Gate Settings sub menu the original sweep time is retrieved For details also see Chapter 6 2 8 3 Using Gated...

Страница 355: ...yed versus time 1 Press the Gate Settings softkey to define the settings of the gate mode At the center frequency a transition to zero span is made and the time parameters gate delay and gate length a...

Страница 356: ...Spectrum Measurements R S ESRP 353 User Manual 1175 7445 02 08 Figure 6 14 TDMA signal with GATE OFF Figure 6 15 Timing diagram for GATE GATE DELAY and GATE LENGTH Configuration...

Страница 357: ...nput Output Menu The following table shows all softkeys available in the Input Output menu It is possi ble that your instrument configuration does not provide all softkeys If a softkey is only availab...

Страница 358: ...f output for the optional IF Video output When you turn on the output you can select to output either the intermediate fre quency or the video signal Note Video output does not return valid values in...

Страница 359: ...n be performed when the button is pressed By default when you press the probe s micro button the R S ESRP is set to single sweep mode and a single sweep is performed This allows you to start a measure...

Страница 360: ...with the probe Place the probe on the required position on the test equipment then press the micro button on the probe to perform a single sweep measurement Probe Configuration Principally the probe...

Страница 361: ...D PARTnumber on page 789 Micro Button Action Defines which action is taken when the probe s micro button is pressed RunSingle A single sweep is performed No Action No action is taken Remote command PR...

Страница 362: ...0 To open the Trace menu Press the TRACE key The Trace menu is displayed The Trace Configuration dialog box is displayed Menu and softkey description Chapter 6 3 1 1 Softkeys of the Trace Menu on page...

Страница 363: ...1 Trace 2 Trace 3 Trace 4 Trace 5 Trace 6 Selects the active trace 1 2 3 4 5 6 and opens the Trace Mode submenu for the selected trace In the default state trace 1 is in Clear Write mode The other tra...

Страница 364: ...page 652 Average Trace 1 Trace 2 Trace 3 Trace 4 Trace 5 Trace 6 The average is formed over several sweeps The Sweep Count determines the number of averaging procedures All available detectors can be...

Страница 365: ...and DISPlay WINDow n TRACe t MODE HCONtinuous on page 817 Detector Trace 1 Trace 2 Trace 3 Trace 4 Trace 5 Trace 6 Opens a submenu to select the detector manually or activate automatic selection Note...

Страница 366: ...ector Remote command DET NEG see SENSe WINDow DETector trace FUNCtion on page 821 Sample Detector Trace 1 Trace 2 Trace 3 Trace 4 Trace 5 Trace 6 Selects the Sample detector Remote command DET SAMP se...

Страница 367: ...the selection of linear averaging The trace data is converted to linear values prior to averaging then averaged and reconverted to loga rithmic values After these conversions the data is displayed on...

Страница 368: ...oint defined with the Trace Math Position softkey The following subtractions can be performed T1 T1 T2 Subtracts trace 2 from trace 1 T1 T1 T3 Subtracts trace 3 from trace 1 T1 T1 T4 Subtracts trace 4...

Страница 369: ...e in W Remote command CALC MATH MODE POW see CALCulate n MATH MODE on page 816 Trace Math Position Opens an edit dialog box to define the zero point in of the diagram height The range of values extend...

Страница 370: ...n see Auto Select softkey If activated the Trace Detector setting is ignored Trace Detector Defines a specific trace detector If one of the following settings is defined the Detector Auto Select optio...

Страница 371: ...n page 370 4 To change the sweep count setting which also determines trace averaging press the Sweep Count softkey 5 To deactivate the reset of the traces in Min Hold and Max Hold mode after some spec...

Страница 372: ...r of sweeps is performed The average trace n is calculated at each measurement point according to Figure 6 17 Equation 1 with Avg average trace Curr current trace Until the first statistics cycle is c...

Страница 373: ...e bandwidth This ensures that the oscil lator step speed is conform to the hardware settling times and does not affect the pre cision of the measured power The number of measured values taken during a...

Страница 374: ...ion provided when you export a single trace Green font Information provided when you export all traces Black font Information provided regardless of the export mode Scale of the x axis Unit of the x a...

Страница 375: ...000 Reference position Header Blue font Information provided when you export a single trace Green font Information provided when you export all traces Black font Information provided regardless of the...

Страница 376: ...e Spectrum mode support a spectrogram result display Menus and softkeys The main menu of the Spectrogram result display is part of the Trace menu in Spec trum mode For more information see Some specia...

Страница 377: ...is updated continuously You can also restore the trace to a point that has already been recorded by selecting a specific frame see Select Frame on page 344 This is possible in Single Sweep mode or if...

Страница 378: ...ap The maximum value is on the right of the map You can also change the color scheme in use see Color Mapping The colors corresponding to the power levels however are always assigned automatically Mar...

Страница 379: ...instrument configuration does not provide all softkeys If a softkey is only available with a special option model or measurement mode this information is delivered in the corresponding softkey descrip...

Страница 380: ...the sweep When active the time stamp replaces the display of the frame number Remote command CALCulate n SGRam TSTamp STATe on page 826 Clear Spectrogram Spectrogram Resets the Spectrogram result disp...

Страница 381: ...r value is changed via the rotary knob Marker 1 Marker 2 Marker 3 Marker 16 Marker Norm Delta softkeys see Marker 1 Marker 2 Marker 3 Marker 16 Marker Norm Delta on page 177 If more measured values th...

Страница 382: ...position The link is off by default You can set the two markers on different traces to measure the difference e g between a max hold trace and a min hold trace or between a measurement and a ref eren...

Страница 383: ...y and level of the marker are displayed in relation to marker 1 in the marker field To change the marker type of a marker select the softkey for the corresponding marker then press the Marker 1 Marker...

Страница 384: ...er the index indicates the trace to which the marker is assigned Example M1 1 defines marker 1 on trace 1 For more than 2 markers a separate marker table is displayed beneath the diagram If applicable...

Страница 385: ...ncy time value is dis played Menu and softkey description Softkeys of the Marker To Menu on page 382 Further information Effect of Different Peak Excursion Settings Example on page 385 Tasks Searching...

Страница 386: ...tic peak search action for marker 1 at the end of each particular sweep This function may be used during adjustments of a device under test to keep track of the current peak marker position and level...

Страница 387: ...ne the lower limit press the Left Limit softkey To define the upper limit press the Right Limit softkey To define the threshold press the Threshold softkey To switch the search limits off press the Se...

Страница 388: ...herent noise of the instrument To avoid identifying noise peaks as maxima or minima enter a peak excursion value that is higher than the difference between the highest and the lowest value mea sured f...

Страница 389: ...methods apart from signal 1 signal 2 is detected as the signal level decreases at least by 29 85 dB to either side of this signal which is now greater than the peak excursion Next lower maximum Next r...

Страница 390: ...397 Measurement example for Phase Noise Auto Peak Search on page 399 Tasks Setting a Fixed Reference Point Phase Noise Measurement on page 396 Setting the Demodulation Mode and Duration on page 396 Pe...

Страница 391: ...ect one of 16 markers and define whether the marker is a nor mal or a delta marker see Marker 1 Marker 2 Marker 3 Marker 16 Marker Norm Delta on page 177 No indicates the number of the currently activ...

Страница 392: ...is acti vated its position becomes the reference point for the measurement Deltamarker 2 is activated and can be used to read out the phase noise value at a given frequency offset Remote command CALC...

Страница 393: ...s of the reference marker marker 1 as reference point for the delta markers level and frequency or time are set to fixed values and used as reference point Ref Fixed On Off Ref Fixed Switches the rela...

Страница 394: ...n on page 396 Remote command CALCulate n MARKer m FUNCtion DEModulation STATe on page 843 AM Marker Demod Sets AM as the output demodulation mode This is the default setting For more information see A...

Страница 395: ...elch level is indicated by a red line in the diagram Remote command SENSe DEMod SQUelch LEVel on page 617 n dB down Opens an edit dialog box to enter a value to define the level spacing of the two tem...

Страница 396: ...vel Response values are given In addition the peaks are indicated in the trace display A maximum of 50 entries are listed Remote command CALCulate n MARKer m FUNCtion FPEaks COUNt on page 848 CALCulat...

Страница 397: ...n MARKer m X SLIMits RIGHT on page 658 Threshold Marker Peak List Opens an edit dialog box to define the threshold line The threshold line represents the lower level limit for a Peak search and the up...

Страница 398: ...lyzer but only for Spec trum displays see Display Config on page 431 For more information see Performing Band Power Measurements on page 399 Select Marker No Band Power Opens a submenu to select one o...

Страница 399: ...ference point proceed as follows a Press the Ref Fixed softkey b Press the Ref Point Level softkey and enter a reference level value c If span 0 press the Ref Point Frequency span 0 Ref Point Time zer...

Страница 400: ...rrespective of the set span the frequency counter delivers a more exact result than a measurement performed with a marker This is due to the following The marker measures only the position of the poin...

Страница 401: ...e to stabilize the measured values With RMS detector used trace averaging must not be used since in this case it produces too low noise levels which cannot be corrected Instead the sweep time can be i...

Страница 402: ...the maximum of the displayed trace 6 PHASE NOISE 800 kHz The phase noise measurement is switched on The delta marker is positioned on the main marker and the measured phase noise value is displayed in...

Страница 403: ...mode for the marker trace is set to AutoSelect the RMS detector is used 1 In the MKR FUNC menu press Band Power 2 In the Band Power menu press Span and enter the width of the band around the marker fo...

Страница 404: ...upon an Existing Limit Line on page 195 Chapter 5 4 3 9 Activating Deactivating a Limit Line on page 196 6 4 Advanced Measurement Examples This chapter explains how to operate the R S ESRP in spectru...

Страница 405: ...ess the SETUP key c Press the Service softkey d Press the Input RF Cal softkey until RF is highlighted The internal signal path of the R S ESRP is switched back to the RF input in order to resume norm...

Страница 406: ...step size for the center frequency is now equal to the marker frequency 6 Set the center frequency to the second harmonic of the signal a Press the FREQ key b Press the UPARROW key once The center fr...

Страница 407: ...ppropriate Resolution Bandwidth A basic feature of a signal analyzer is being able to separate the spectral components of a mixture of signals The resolution at which the individual components can be...

Страница 408: ...me span The sweep time has to allow the resolution filters to settle during a sweep at all signal levels and frequencies to be displayed 6 4 3 2 Intermodulation Measurements If several signals are app...

Страница 409: ...lation using the 3rd order intercept TOI measurement func tion 1 Press the MEAS key and then the TOI softkey The R S ESRP activates four markers to measure the intermodulation distance Two markers are...

Страница 410: ...R S ESRP s intrinsic intermodulation products disappear below the noise floor 6 4 4 Measuring Signals in the Vicinity of Noise The minimum signal level a signal analyzer can measure is limited by its...

Страница 411: ...derably smaller than the resolution bandwidth noise spikes are suppressed i e the trace becomes much smoother The level of a sine wave signal is not influenced by the video bandwidth A sine wave signa...

Страница 412: ...b Press the SPAN key and enter 100 MHz Figure 6 25 Sine wave signal with low S N ratio The signal is measured with the auto peak detector and is completely hidden in the intrinsic noise of the R S ESR...

Страница 413: ...RF signal therefore can be more clearly distinguished from noise Figure 6 26 RF sine wave signal with low S N ratio if the trace is averaged 4 Instead of trace averaging a video filter that is narrow...

Страница 414: ...oftkey and enter 10 kHz The RF signal can be more clearly distinguished from noise Figure 6 27 RF sine wave signal with low S N ratio if a smaller video bandwidth is selected 5 By reducing the resolut...

Страница 415: ...ll be noisier Figure 6 28 Reference signal at a smaller resolution bandwidth 6 4 5 Noise Measurements Noise measurements play an important role in signal analysis Noise e g affects the sensitivity of...

Страница 416: ...state 2 Set the center frequency to 1 234 GHz and the span to 1 MHz a Press the FREQ key and enter 1 234 GHz b Press the SPAN key and enter 1 MHz 3 Switch on the marker and set the marker frequency to...

Страница 417: ...rected by 10 log RBWnoise with RBWnoise being the power bandwidth of the selected resolution filter RBW Detector selection The noise power density is measured in the default setting with the sample de...

Страница 418: ...wers By means of the following diagram the noise level of the DUT can be estimated from the level difference between the total and the internal noise level Figure 6 29 Correction factor for measured n...

Страница 419: ...R S ESRP is in its default state 2 Set the center frequency to 1 GHz and the span to 1 MHz 3 To obtain maximum sensitivity set RF attenuation on the R S ESRP to 0 dB 4 Set the Sweep Type to Sweep 5 Sw...

Страница 420: ...MHz channel bandwidth 6 Stabilize the measurement result by increasing the sweep time In the Ch Power ACLR menu press the Sweep Time softkey and enter 1 s The trace becomes much smoother because of th...

Страница 421: ...se softkey The R S ESRP activates phase noise measurement Marker 1 main marker and marker 2 delta marker are positioned on the signal maximum The posi tion of the marker is the reference level and fre...

Страница 422: ...el power could be measured at highest accuracy with a power meter its low selectivity means that it is not suitable for measuring adja cent channel power as an absolute value or relative to the transm...

Страница 423: ...ith a resolution bandwidth that is less than the channel bandwidth and integrates the level values of the trace versus the channel bandwidth This method is described in Measurement Example Measuring t...

Страница 424: ...el The spectrum is dis played in the upper part of the screen the numeric values of the results and the channel configuration in the lower part of the screen The various channels are rep resented by v...

Страница 425: ...2000 1x signal Measurement Example 2 Measuring Adjacent Channel Power of a W CDMA Uplink Signal Test setup Signal generator settings e g R S SMW Frequency 1950 MHz Level 4 dBm Modulation 3 GPP W CDMA...

Страница 426: ...idth and the detector are automatically set to the correct values The spectrum is displayed in the upper part of the screen and the channel power the level ratios of the adjacent chan nel powers and t...

Страница 427: ...d as channel filter Figure 6 34 Measuring the adjacent channel power of a W CDMA signal with the fast ACLR mode Optimum Level Setting for ACP Measurements on W CDMA Signals The dynamic range for ACPR...

Страница 428: ...B steps with the 5 dB RF attenuator the optimum 10 dB range spreads from 17 dBm to 22 dBm In this range the obtainable dynamic range with noise correction is 77 dB To set the attenuation parameter man...

Страница 429: ...that has a specified probability of being exceeded e g 0 01 To measure the amplitude distribution the R S ESRP has simple measurement func tions to determine both the APD Amplitude Probability Distri...

Страница 430: ...User Manual 1175 7445 02 08 Figure 6 36 Amplitude probability distribution of white noise 3 Switch to the CCDF display mode a Press the UP key b Press the CCDF softkey The CCDF display mode is switche...

Страница 431: ...race indicates the probability that a level will exceed the mean power The level above the mean power is plotted along the x axis of the graph The origin of the axis corresponds to the mean power leve...

Страница 432: ...UT OUTPUT Key on page 354 The Marker menu is identical to the one in Spectrum mode for display modes Magni tude Real Imag and Spectrum except for Marker Zoom see Display Config on page 431 For the oth...

Страница 433: ...ff 431 Data Acquisition 431 Sample Rate 431 Filter BW 431 Meas Time 431 Record Length 431 Display Config 431 C N C No 433 C N 434 C No 434 Channel Bandwidth 434 Adjust Settings 434 I Q Analyzer Starts...

Страница 434: ...as Time Data Acquisition I Q Analyzer Defines the I Q acquisition time By default the measurement time is calculated as the number of I Q samples Record Length divided by the sample rate If you change...

Страница 435: ...lyzer R S ESRP 432 User Manual 1175 7445 02 08 Magnitude Shows the values in time domain Spectrum Displays the frequency spectrum of the captured I Q samples Softkeys and Parameters of the I Q Analyze...

Страница 436: ...line Real Imag I Q Displays the I and Q values in separate diagrams Remote command CALCulate n FORMat on page 856 C N C No I Q Analyzer Opens a submenu to configure the carrier noise ratio measurement...

Страница 437: ...1 is activated The measurement is performed on the trace that marker 1 is assigned to To shift marker 1 and measure another trace use the Marker to Trace softkey in the Marker To menu Remote command...

Страница 438: ...Log 100 dB 436 Range Log 50 dB 436 Range Log 10 dB 436 Range Log 5 dB 437 Range Log 1 dB 437 Range Log Manual 437 Range Linear 437 Range Lin Unit 437 Unit 438 Preamp On Off 438 RF Atten Auto 438 Ref L...

Страница 439: ...he level display range to 100 dB Remote command Logarithmic scaling DISP WIND TRAC Y SPAC LOG see DISPlay WINDow n TRACe t Y SPACing on page 636 Display range DISP WIND TRAC Y 100DB see DISPlay WINDow...

Страница 440: ...mote command Logarithmic scaling DISP WIND TRAC Y SPAC LOG see DISPlay WINDow n TRACe t Y SPACing on page 636 Display range DISPlay WINDow n TRACe t Y SCALe on page 812 Range Linear Range Selects line...

Страница 441: ...Ampere Watt Remote command CALCulate n UNIT POWer on page 812 Preamp On Off Switches the preamplifier on and off Remote command INPut GAIN STATe on page 638 RF Atten Auto Sets the RF attenuation auto...

Страница 442: ...If activated the results are corrected by the instrument s inherent noise which increa ses the dynamic range ON A reference measurement of the instrument s inherent noise is carried out The noise pow...

Страница 443: ...lable Search Settings Opens a dialog box to define which data is used for marker search functions Note The search settings apply to all markers not only the currently selected one Search Real Marker s...

Страница 444: ...ered Once a measurement is completed another is started immediately Remote command TRIG SOUR IMM see TRIGger n SEQuence SOURce on page 807 External Trg Gate Source Defines triggering via a TTL signal...

Страница 445: ...triggering of the measurement using the second intermediate frequency For this purpose the R S ESRP uses a level detector at the second intermediate fre quency The available trigger levels depend on t...

Страница 446: ...lapsed Neg Edge triggering the sweep is continued on a 0 to 1 transition for the gate length duration after the gate delay time has elapsed Remote command TRIGger n SEQuence SLOPe on page 807 SENSe SW...

Страница 447: ...Holdoff Defines the value for the trigger holdoff The holdoff value in s is the time which must pass before triggering in case another trigger event happens This softkey is only available if IFPower R...

Страница 448: ...ed for accurate analysis by the R S ESRP For the I Q data acquisition digital decimation filters are used internally The passband of these digital filters determines the maximum usable I Q bandwidth I...

Страница 449: ...I Q Analyzer R S ESRP 446 User Manual 1175 7445 02 08 Figure 7 2 Relation between maximum usable bandwidth and sample rate RF input Working with I Q Data...

Страница 450: ...this case Menu and softkey description Chapter 8 1 Softkeys of the Tracking Generator Menu on page 447 Further information 8 1 Softkeys of the Tracking Generator Menu The Tracking Generator menu is d...

Страница 451: ...n for transmission and reflection measurement for tracking generators For details on the test setups see Chapter 8 4 2 Calibrating for transmission and reflection measurement on page 464 Calibrate Tra...

Страница 452: ...Normalization on page 465 Remote command DISPlay WINDow n TRACe t Y SCALe RPOSition on page 814 Reference Value Source Cal Opens an edit dialog box to enter a position value that shifts the reference...

Страница 453: ...modulation of the tracking generator output signal The modulation signal is applied to the TG IN I AM connector An input voltage of 1 V corresponds to 100 amplitude modulation Switching on an externa...

Страница 454: ...Displayed Information and Errors on page 471 The start power can be set between 30 dBm and 0 dBm The stop value can also be be set between 30 dBm and 0 dBm and may be smaller than the start value The...

Страница 455: ...ecording of measured values is possible For generators with a TTL interface a list of the frequencies to be set is entered into the generator before the beginning of the first sweep Then the sweep is...

Страница 456: ...SMA and R S SMU require the following firmware versions R S SMA V2 10 x or higher R S SMU V1 10 or higher Generator Interface Type Generator Min Freq Generator Max Freq Generator Min Power dBm Generat...

Страница 457: ...00 kHz 6 4 GHz 144 10 SML01 GPIB 9 kHz 1 1 GHz 140 13 SML02 GPIB 9 kHz 2 2 GHz 140 13 SML03 GPIB 9 kHz 3 3 GHz 140 13 SMR20 TTL 1 GHz 20 GHz 130 2 11 2 SMR20B111 TTL 10 MHz 20 GHz 130 2 13 2 SMR27 TTL...

Страница 458: ...U03 TTL 100 kHz 3 GHz 145 13 SMU03B31 TTL 100 kHz 3 GHz 145 19 SMU04 TTL 100 kHz 4 GHz 145 13 SMU04B31 TTL 100 kHz 4 GHz 145 19 SMU06 TTL 100 kHz 6 GHz 145 13 SMU06B31 TTL 100 kHz 6 GHz 145 19 SMW20 T...

Страница 459: ...ators The Tracking Generator Configuration dialog box is opened via the Source Config softkey in the Tracking Generator menu This dialox box allows you to define measurement settings as well as connec...

Страница 460: ...lt output power is 20 dBm The range is specified in the data sheet Remote command SOURce n POWer LEVel IMMediate AMPLitude on page 873 Power Offset Constant level offset for the tracking generator Val...

Страница 461: ...3 Result Frequency Start For reference only The start frequency for the generator calculated from the config ured generator frequency and the start value defined for the analyzer Result Frequency Stop...

Страница 462: ...is selected To acti vate the currently selected generator click the Source RF On softkey Source Frequency Coupling Enables or disables frequency coupling between the analyzer and the generator Auto De...

Страница 463: ...may differ from the input from the analyzer The RF frequency may be multiplied by a specified factor or a frequency offset can be added or both Note The input for the generator frequency is not valida...

Страница 464: ...MHz Numerator Denominator 1 FGeneratorStart 200 MHz FGeneratorStop 100 MHz If the offset is adjusted so that the sweep of the generator crosses the minimum gener ator frequency a message is displayed...

Страница 465: ...e allows for considerably higher measurement rates than pure GPIB control because the frequency stepping of the R S ESRP is directly cou pled with the frequency stepping of the generator Remote comman...

Страница 466: ...d for measuring the reference curve are included in the reference dataset Even with normalization switched on the instrument settings can be changed in a wide area without stopping the normalization T...

Страница 467: ...e status line Overloading can be avoided as follows Reducing the output level of the tracking generator Source Config on page 452 softkey in the Tracking Generator menu Increasing the reference level...

Страница 468: ...e R S ESRP is fed from the output of the DUT A calibration can be carried out to compensate for the effects of the test setup e g frequency response of connecting cables Figure 8 3 Test setup for tran...

Страница 469: ...Tracking Generator R S ESRP 466 User Manual 1175 7445 02 08 Figure 8 5 Normalized display CORR ON see SENSe CORRection STATe on page 870 Tracking Generator Functions...

Страница 470: ...RP 467 User Manual 1175 7445 02 08 Figure 8 6 Normalized measurement shifted with Reference Value Position 50 DISP WIND TRAC Y RPOS 10PCT see DISPlay WINDow n TRACe t Y SCALe RPOSition on page 814 Tra...

Страница 471: ...0 After the reference line has been shifted by entering 10 dB as the Reference Value on page 449 deviations from the nominal value can be displayed with high resolution e g 2 dB Div The absolute measu...

Страница 472: ...can be influenced by means of external signals input voltage range 1 V to 1 V Two BNC connectors at the rear panel are available as signal inputs Their function changes depending on the selected modul...

Страница 473: ...ator output signal The modulation frequency range is 1 kHz to 100 kHz the deviation can be set in 1 decade steps in the range of 100 Hz to 10 MHz at an input voltage of 1 V The phase deviation h shoul...

Страница 474: ...age 871 8 5 Displayed Information and Errors Diagram header In Tracking Generator measurement mode some additional information is displayed in the diagram header Label Description INT TG source power...

Страница 475: ...a cable damage or loose connec tion or wrong address Ext Generator Limits Exceeded The allowed frequency or power ranges for the generator were exceeded Reverse Sweep via min Ext Generator Frequency...

Страница 476: ...ed remotely the instrument only returns to the manual operation state when the alignment or test is completed The change to manual operation consists of Enabling the Front Panel Keys Returning to manu...

Страница 477: ...tkeys as well as a User Preference Setup softkey that allows you to define them Pressing one of the user definable soft keys has the same effect as the Save File Recall File function for a pre defined...

Страница 478: ...ct The selected file is displayed in the ApplicationManager dialog To store the softkey settings Click Save to store the user defined softkey definitions 9 3 Instrument Setup and Interface Configurati...

Страница 479: ...tics of a transducer factor 479 Managing transducer sets 481 Alignment 482 Self Alignment 482 Show Align Results 483 Touch Screen Alignment 483 General Setup 483 Configure Network 483 Network Address...

Страница 480: ...User Defined Colors 493 Set to Default 493 Print Colors 493 Select Print Color Set 494 Color On Off 494 Display Pwr Save On Off 494 System Info 494 Hardware Info 495 Versions Options 495 System Messag...

Страница 481: ...3 1 2 Rear Panel View on page 33 Remote command SENSe ROSCillator SOURce on page 902 Handle missing Ext Ref If an external reference is selected but none is available there are different ways the inst...

Страница 482: ...ed To actually include the transducer in the measurement you first have to select one and turn it on with the Active On Off softkey When the transducer factor is active all amplitude settings and outp...

Страница 483: ...ton to save and store the transducer factor on the internal hard disk of the R S ESRP Save Fac tor softkey Note that there are softkeys for several elements of the dialog box as mentioned in the legen...

Страница 484: ...tion TRANsducer SELect on page 921 SENSe CORRection TRANsducer COMMent on page 919 SENSe CORRection TRANsducer UNIT on page 921 SENSe CORRection TRANsducer SCALing on page 920 SENSe CORRection TRANsdu...

Страница 485: ...T SELect on page 924 SENSe CORRection TSET COMMent on page 923 SENSe CORRection TSET UNIT on page 924 SENSe CORRection TSET BREak on page 922 SENSe CORRection TSET RANGe range on page 923 SENSe CORRec...

Страница 486: ...hard disk If you notice that touching a specific point on the screen does not achieve the correct response you may try adjusting the alignment as well Using a finger or any other pointing device pres...

Страница 487: ...efer to Chapter 11 2 5 Setting Up a Network LAN Connection on page 592 DHCP On Off Network Address General Setup Switches between DHCP server available On or not available Off If a DHCP server is avai...

Страница 488: ...r accounts with administrator rights can reset the LAN configuration Remote command SYSTem LXI LANReset on page 958 GPIB General Setup Opens a submenu to set the parameters of the remote control inter...

Страница 489: ...umber of measurement points and available bandwidths to default R S ESRP values Remote command SYST COMP DEF see SYSTem COMPatible on page 888 Mode R S FSP Compatibility Mode GPIB General Setup Sets t...

Страница 490: ...ode GPIB General Setup Configures the internal IF gain settings in HP emulation mode due to the application needs This setting is only taken into account for resolution bandwidth 300 kHz NORM Optimize...

Страница 491: ...18 GPIB Commands of HP Models 856xE 8566A B 8568A B and 8594E on page 1007 Max 36 characters are allowed Remote command SYSTem REVision STRing on page 959 Tree Walking Compatibility Mode GPIB General...

Страница 492: ...FSV 7 1307 9002K07 100005 1 61 Remote command SYSTem FORMat IDENt on page 889 I O Logging On Off GPIB General Setup Activates or deactivates the SCPI log function All remote control commands received...

Страница 493: ...e the hardkeys and other hardware controls of the device are simulated on the screen This is the setting for working at a computer with Remote Desktop or at an external monitor Figure 9 1 Soft frontp...

Страница 494: ...ff Display Setup Activates deactivates the display of a diagram title if available and opens an edit dia log box to enter a new title for the active diagram Max 20 characters are allowed Remote comman...

Страница 495: ...ed Colors on page 493 Set to Default on page 493 Remote command DISPlay CMAP item HSL on page 946 Select Screen Color Set Screen Colors Display Setup Opens the Select Screen Color Set dialog box to se...

Страница 496: ...vailable if in the Select Color Set dialog box the User Defined Colors option is selected or the Color Setup dialog box is displayed For details on screen colors refer to Chapter 3 2 4 4 Setting the S...

Страница 497: ...witches from color printout to black and white printout and back All colored areas are printed in white and all colored lines in black This improves the contrast The default setting is color printout...

Страница 498: ...mation on open source code used by the R S ESRP firmware For details on options refer to Chapter 3 2 1 10 Checking the Supplied Options on page 43 Remote command IDN on page 884 OPT on page 885 SYSTem...

Страница 499: ...andwidth settings including filter type Limit lines Limit lines are only transferred if they are compatible to the current diagram config uration Remote command See Chapter 11 9 3 Coupling Parameters...

Страница 500: ...e to install options Application Setup Recovery Controls instrument behavior when switching between measurement applications e g from Spectrum to Analog Demod and back If this softkey is activated the...

Страница 501: ...to the RF input connector normal position This is the default setting Remote command DIAGnostic n SERVice INPut SELect on page 905 Calibration Frequency RF Input Source Service Opens an edit dialog bo...

Страница 502: ...c tions are only used by authorized personnel Remote command SYSTem PASSword CENable on page 907 Service Function Service Opens the Service Function dialog box to start special service functions For f...

Страница 503: ...s anticipated that these transducer values should remain valid after every recall of that save set Thus even if the transducer file is changed and the original save set file is recalled later the orig...

Страница 504: ...nu The following table shows all softkeys available in the Save Recall menu Save 502 Save File Recall File 502 Select Path 502 Select File 502 Edit File Name 502 Edit Comment 502 Select Items 502 Sele...

Страница 505: ...hapter 9 2 User Defined Menu USER key on page 474 Remote command MMEMory STORe n STATe on page 916 MMEMory STORe n STATe NEXT on page 917 MMEMory LOAD STATe on page 915 Select Path Save Opens the dire...

Страница 506: ...te command MMEMory CLEar STATe 1 on page 914 Recall Opens the Recall dialog box to load a settings file To navigate in the dialog box use the corresponding softkeys For details see also Chapter 10 1 2...

Страница 507: ...xports the currently selected trace indicated by the softkey label Note that the exported ASCII file has a slightly different structure compared to a single trace export The file consists of the heade...

Страница 508: ...single sweep mode For details see Chapter 10 1 3 Importing and Exporting I Q Data on page 509 Remote command MMEMory LOAD IQ STATe on page 857 File Manager Opens the File Manager dialog box and a sub...

Страница 509: ...mation Remote command MMEMory DELete on page 911 MMEMory RDIRectory on page 913 Sort Mode File Manager Opens a submenu to select the sorting mode for the displayed files The entry for the next higher...

Страница 510: ...page 956 Disconnect Network Drive Network Drive File Manager Opens the Disconnect Network Drive dialog box In the Drive list select the drive you want to disconnect and confirm with OK Remote command...

Страница 511: ...path or extension To set the focus on File Name field press the Edit File Name softkey In the Save dialog box the field already contains a suggestion for a new name the file name used in the last sav...

Страница 512: ...ons for further processing As opposed to storing trace data which may be averaged or restricted to peak values I Q data is stored as it was captured without further processing The data is stored as co...

Страница 513: ...tp www rohde schwarz com file RsIqTar xsd In particular the order of the XML elements must be respected i e iq tar uses an ordered XML schema For your own implementation of the iq tar file format make...

Страница 514: ...sample must be in the same format The for mat can be one of the following Complex Complex number in cartesian format i e I and Q values inter leaved I and Q are unitless Real Real number unitless Pola...

Страница 515: ...ou add user data as XML content PreviewData Optional contains further XML elements that provide a preview of the I Q data The preview data is determined by the routine that saves an iq tar file e g R...

Страница 516: ...and writing of streamed I Q data all data is interleaved i e complex values are interleaved pairs of I and Q values and multi channel signals contain interleaved complex sam ples for channel 1 channel...

Страница 517: ...rint menu It is possible that your instrument configuration does not provide all softkeys If a softkey is only available with a special option model or measurement mode this information is delivered i...

Страница 518: ...are selected via the Device 1 2 on page 515 softkey For further information refer to Chapter 3 2 4 6 Selecting and Configuring Printers on page 51 Remote command HCOPy DEVice LANGuage 1 2 on page 929...

Страница 519: ...x 120 characters are allowed 60 characters fit in one line In the first line at any point a manual line feed can be forced by entering Date and time are inserted automatically The comment is printed b...

Страница 520: ...Report Template The R S ESRP allows you to create test report templates for example if you handle different measurement tasks that require different information or a different layout in the test repo...

Страница 521: ...e Transducer on page 478 Scan Result List Adds the numerical scan results to the dataset Note that the test report will show only the results of the first 50000 measurement points The rest is cut off...

Страница 522: ...jpg and png format Similar to the alphanumeric lines in the header you can select the visibility for the logo as well By default the logo is not displayed in the report Remote command Title HCOPy TREP...

Страница 523: ...report configuration Before you save the configuration as a template enter a name for the template in the corresponding field Delete Deletes the selected test report template The table shows all avai...

Страница 524: ...Configuring and Printing the Test Report You can save the test report either as a pdf document or an rtf document in the corre sponding dropdown menu in the Test Report Content Selection dialog box Th...

Страница 525: ...ter descriptions of the variables as required For more information on changing the labels of the header see Designing the Header of the Test Report on page 518 Remote command New report HCOPy TREPort...

Страница 526: ...command Select a single dataset HCOPy TREPort TEST SELect on page 943 Select all datasets HCOPy TREPort TEST SELect ALL on page 943 Invert selection HCOPy TREPort TEST SELect INVert on page 943 Selec...

Страница 527: ...g with Test Reports 934 Display Configuration 944 Network Connection 955 Status Register 962 Remote Control Programming Examples 965 GPIB Commands of HP Models 856xE 8566A B 8568A B and 8594E 1007 11...

Страница 528: ...terface refer to Chapter 11 1 1 4 GPIB Interface IEC 625 IEEE 418 Bus Interface on page 530 VISA is a standardized software interface library providing input and output functions to communicate with i...

Страница 529: ...are used to communicate between the con troller and the instrument Interface messages can only be sent by instruments that have GPIB bus functionality For details see the sections for the required in...

Страница 530: ...ace card and protocols The network card can be oper ated with a 10 MHz Ethernet IEEE 802 3 or a 100 MHz Ethernet IEEE 802 3u inter face For remote control via a network the PC and the instrument must...

Страница 531: ...are connected to the network each instrument has its own IP address and associated resource string The controller identifies these instruments by means of the resource string For details on configurin...

Страница 532: ...ily require a VISA installation on the remote controller side It is available by default on all operating systems The simplest way to establish socket communication is to use the built in telnet pro g...

Страница 533: ...device online offline RSDLLTestSrq Checks whether a device has generated an SRQ RSDLLWaitSrq Waits until a device generates an SRQ RSDLLSwapBytes Swaps the byte sequence for binary numericdisplay only...

Страница 534: ...Messages Interface messages are transmitted to the instrument on the data lines with the atten tion line ATN being active LOW They are used for communication between the con troller and the instrumen...

Страница 535: ...Remote Control R S ESRP 532 User Manual 1175 7445 02 08 Figure 11 1 Communication lines used by the GPIB interface Remote Control Basics...

Страница 536: ...range 00 through 0F hex They only affect instruments addressed as listeners Command Effect on the instrument GET Group Execute Trigger Triggers a previously active instrument function e g a sweep The...

Страница 537: ...control and vice versa does not affect the other instrument set tings 2 During program execution send the SYSTem DISPlay UPDate ON command to activate the display of results The changes in the device...

Страница 538: ...ost cases one or more parameters The header and the parameters are separated by a white space ASCII code 0 to 9 11 to 32 decimal e g blank The headers may consist of several mnemonics keywords Queries...

Страница 539: ...ic_value numeric_value Long and short form 536 Numeric Suffixes 536 Optional Mnemonics 537 Long and short form The mnemonics feature a long form and a short form The short form is marked by upper case...

Страница 540: ...nstrument must recognize the long command to comply with the SCPI standard Some commands are considerably shortened by these optional mnemonics Example Definition HCOPy IMMediate Command HCOP IMM is e...

Страница 541: ...If the unit is missing the basic unit is used Allowed unit prefixes are G giga MA mega MOHM MHZ K kilo M milli U micro N nano Example SENSe FREQ STOP 1 5GHz SENSe FREQ STOP 1 5E9 Some settings allow...

Страница 542: ...after the quotation mark Example SENSe LIST FREQ MAXimum Returns the maximum numeric value as a result Boolean Parameters Boolean parameters represent two states The ON state logically true is represe...

Страница 543: ...ires a NL END message to terminate the data block This format is useful when the length of the transmission is not known or if speed or other considerations prevent segmentation of the data into block...

Страница 544: ...ollowing commands with identical meaning can be created SENS BAND RES 1 SENS BWID RES 1 Mnemonics in square brackets are optional and may be inserted into the header or omitted Example HCOPy IMMediate...

Страница 545: ...to SCPI the responses to queries are partly subject to stricter rules than in standard IEEE 488 2 The requested parameter is transmitted without a header Example HCOP PAGE ORI Response LAND Maximum va...

Страница 546: ...e command line even though they may be implemented as sequential commands are not necessarily serviced in the order in which they have been received In order to make sure that commands are actually ca...

Страница 547: ...el settings for adjacent channel power measure ments 11 1 5 1 Preventing Overlapping Execution To prevent an overlapping execution of commands one of the commands OPC OPC or WAI can be used All three...

Страница 548: ...rlapped command with OPC 3 Wait for a service request The service request indicates that the overlapped command has finished Event status register ESE 1 Set the OPC mask bit bit no 0 in the ESE ESE 1...

Страница 549: ...the messages the instrument returns to the controller It is not part of the status reporting system but determines the value of the MAV bit in the STB and thus is represented in the overview All stat...

Страница 550: ...filter When a bit of the CONDition part is changed from 0 to 1 the associated PTR bit decides whether the EVENt bit is set to 1 PTR bit 1 the EVENt bit is set PTR bit 0 the EVENt bit is not set This...

Страница 551: ...et to 1 as well This part can be written into and read by the user as required Its contents are not affected by reading Sum bit The sum bit is obtained from the EVENt and ENABle part for each register...

Страница 552: ...B and Service Request Enable Register SRE The STatus Byte STB is already defined in IEEE 488 2 It provides a rough over view of the instrument status by collecting the pieces of information of the low...

Страница 553: ...ailable The bit is set if a message is available in the output queue which can be read This bit can be used to enable data to be automatically read from the instrument to the controller 5 ESB bit Sum...

Страница 554: ...it is set if a device dependent error occurs The R S ESRP adds a number to the error queue The number is either in the range between 300 and 399 or greater than 0 If the number is positive it indicate...

Страница 555: ...Register The STATus QUEStionable register contains information about states that may occur if the R S ESRP is operated without meeting the specifications You can read out the state of the register wi...

Страница 556: ...10 Meaning of the bits used in the STATus QUEStionable FREQuency register Bit No Meaning 0 OVEN COLD This bit is set if the reference oscillator has not yet attained its operating temperature OCXO is...

Страница 557: ...set if limit line 8 is violated 8 to 14 Unused 15 This bit is always 0 STATus QUEStionable LMARgin Register This register contains information about the observance of limit margins You can read out th...

Страница 558: ...VERload This bit is set if an overload occurs at the RF input The R S ESRP displays the enhancement label OVLD 1 UNDerload This bit is set if an underload occurs at the RF input The R S ESRP displays...

Страница 559: ...subrange 6 has been reached 6 Range 7 This bit is set if subrange 7 has been reached 7 Range 8 This bit is set if subrange 8 has been reached 8 Range 9 This bit is set if subrange 9 has been reached 9...

Страница 560: ...TATus OPERation Register The STATus OPERation register contains information on current activities of the R S ESRP It also contains information on activities that have been executed since the last read...

Страница 561: ...has been activated 11 14 Not used 15 This bit is always 0 STATus QUEStionable Register The STATus QUEStionable register contains information on instrument states that do not meet the specifications Yo...

Страница 562: ...mation about the results of a limit check during ACLR measurements You can read out the register with STATus QUEStionable ACPLimit CONDition or STATus QUEStionable ACPLimit EVENt Table 11 17 Meaning o...

Страница 563: ...zer SRE 32 Set bit 5 in the SRE ESB 3 CALL InstrWrite analyzer INIT OPC Generate an SRQ after operation complete After its settings have been completed the instrument generates an SRQ The SRQ is the o...

Страница 564: ...The parallel poll itself is executed using PPE The parallel poll method is mainly used to find out quickly which one of the instruments connected to the controller has sent a service request To this...

Страница 565: ...roller program the error queue should be queried regu larly since faulty commands from the controller to the instrument are recorded there as well 11 1 6 5 Reset Values of the Status Reporting System...

Страница 566: ...settings Command sequence As a general rule send commands and queries in different program messages Other wise the result of the query may vary depending on which operation is performed first see also...

Страница 567: ...ted command scripts Single step mode Conditional execution based on the IDN and OPT strings You can use the IECWIN to try out the programming examples provided in the R S ESRP User Manuals Starting IE...

Страница 568: ...Interface 592 Setting Up a Network LAN Connection 592 How to Configure the LAN Using the LXI Web Browser Interface 598 Configuring the Network 600 11 2 1 Operation with Windows Remote Desktop In prod...

Страница 569: ...top program of the controller immediately No further configura tion is required However if the connection fails or other users need to connect this section provides basic instructions on setting up th...

Страница 570: ...ent is part of the operating system and can be accessed via Start All Applications Windows Accessories Remote Desktop Con nection 1 Press the Windows key on the external keyboard or the CTRL ESC key c...

Страница 571: ...the performance you can deactivate the Desktop background Show contents of window while dragging and Menu and window animation options 7 Open the Local Resources tab for enabling printers local drives...

Страница 572: ...ull screen mode is to select Disconnect from the Start menu 11 2 1 3 Starting and Terminating Remote Operation Setting up a connection to the R S ESRP 1 In the Remote Desktop Connection dialog box see...

Страница 573: ...knob are mapped to buttons 8 To deactivate or activate the Soft Front Panel press the F6 key After the connection is established the R S ESRP screen is displayed in the Remote Desktop application win...

Страница 574: ...ent Virtual Network Computing VNC allows you to control the R S ESRP the VNC server from a remote computer the VNC client for example to monitor the instru ments in a production line The handling of a...

Страница 575: ...f Introduction to Remote Control The following programming examples are structured hierarchically i e the later exam ples are based on the preceding ones This allows you to easily assemble a well func...

Страница 576: ...called In addition the VPPTYPE BAS file must be added to the project This file contains constants and definitions for error handling timeout values etc The modules visa32 bas and vpptype bas can be f...

Страница 577: ...this will raise an error if status is not VI_SUCCESS CALL CheckError vi status End Sub Public Sub InstrRead ByVal vi As Long Response As String _ ByVal count As Long retCount As Long Dim status As Lo...

Страница 578: ...Desc returns a text description of the status code returned by viOpenDefaultRM status viOpenDefaultRM defaultRM status viStatusDesc defaultRM status Response Open the connection to the device and stor...

Страница 579: ...tch off screen display END SUB REM Configuring the Power Save Function for the Display During remote control operation it is often unnecessary to display the measurement results on screen Although the...

Страница 580: ...ment settings PUBLIC SUB ReadSettings Dim retCount as Long CFfrequency SPACE 20 Provide text variable 20 characters CALL InstrWrite analyzer FREQ CENT Request center frequency CALL InstrRead analyzer...

Страница 581: ...ommand REM must not be carried out until a full sweep has been completed CALL InstrWrite analyzer INIT CONT OFF REM First method Using WAI CALL InstrWrite analyzer ABOR INIT IMM WAI REM Second method...

Страница 582: ...ion ERROR END SUB REM 11 2 3 2 Detailed Programming Examples In the following sections typical programming examples for setting measurement parameters and functions general settings printout and data...

Страница 583: ...Sub SetupInstrument Default setting f the R S FSV CALL SetupStatusReg Configure status registers CALL InstrWrite analyzer RST Reset instrument CALL InstrWrite analyzer SYST DISP UPD ON ON screen displ...

Страница 584: ...MS Detector Trace1 CALL InstrWrite analyzer DET2 AUTO ON Detector Trace2 CALL InstrWrite analyzer DET3 AUTO ON Detector Trace3 CALL InstrWrite analyzer DET4 AUTO ON Detector Trace4 CALL InstrWrite ana...

Страница 585: ...sweep with sync CALL InstrWrite analyzer CALC MARK MAX X Y Marker to peak read frequency and level CALL InstrRead analyzer result 100 retCount Debug Print Marker 1 result CALL InstrWrite analyzer CAL...

Страница 586: ...Write analyzer INIT WAI Sweep with sync END SUB REM Frequency Counting The following example is based on a signal at 100 MHz with a level of 30 dBm Again the instrument s default setting can be used S...

Страница 587: ...ird harmonic The default setting of the R S ESRP for measurements SetupInstrument is also used as a starting point here after which adaptations for the measurement are carried out REM Public Sub RefFi...

Страница 588: ...he phase noise at a 10 kHz offset from the carrier signal REM Public Sub Noise Dim retCount as Long Default setting of the R S FSV CALL SetupStatusReg Configure status register CALL InstrWrite analyze...

Страница 589: ...ength specification is evaluated and used to calculate the x axis values In ASCII format merely the list of level values is output Binary data is read out in three steps 1 The number of digits in the...

Страница 590: ...tart frequency CALL InstrRead analyzer startFreq 100 retCount startFreq Val startFreq CALL InstrWrite analyzer FREQ SPAN Read span CALL InstrRead analyzer span 100 retCount span Val span Reading out i...

Страница 591: ...he state OFF for the sake of completeness REM Public Sub StoreSettings This subroutine selects the settings to be stored and creates the data record TEST1 in the directory C R_S Instr user It uses the...

Страница 592: ...is then set after each RST presetting and each time the instrument is started For demonstration purposes the command RST is carried out again REM Public Sub StartupRecallSettings Resetting the R S FSV...

Страница 593: ...creen display on Measurement settings CALL InstrWrite analyzer FREQ CENT 100MHz SPAN 10MHz Frequency setting CALL InstrWrite analyzer DISP TRAC Y RLEV 10dBm Reference level CALL InstrWrite analyzer IN...

Страница 594: ...sableEvent vi VI_EVENT_SERVICE_REQ VI_QUEUE Disable subsequent events IF NOT statusSRQ 0 THEN CALL Srq If SRQ not detected Subroutine for evaluation Printout in WMF format BMP format to file CALL Inst...

Страница 595: ...M GPIB ADDR 20 Setting the default ID response string In the GPIB menu press the ID String Factory softkey to select the default response to the IDN command Setting a user defined ID response string I...

Страница 596: ...iguration 11 2 5 1 Connecting the Instrument to the Network There are two methods to establish a LAN connection to the instrument A non dedicated network Ethernet connection from the instrument to an...

Страница 597: ...u ment can be obtained in different ways If the network supports dynamic TCP IP configuration using the Dynamic Host Configuration Protocol DHCP all address information can be assigned automati cally...

Страница 598: ...e Subnet Mask softkey and enter the subnet mask for example 255 255 255 0 The subnet mask consists of four number blocks separated by dots Every block contains a maximum of 3 numbers If you have enter...

Страница 599: ...nvalid IP address or subnet mask the message out of range is displayed in the status line If the settings are correct the configuration is saved and you are prompted to restart the instrument 8 Confir...

Страница 600: ...Windows operating system help 11 2 5 3 Using Computer Names In a LAN that uses a DNS server Domain Name System server each PC or instru ment connected in the LAN can be accessed via an unambiguous co...

Страница 601: ...ork Rohde Schwarz highly recommends the use of the firewall on your instrument Rohde Schwarz instruments are shipped with the Windows firewall enabled and preconfigured in such a way that all ports an...

Страница 602: ...tion provides all mandatory LAN parameters Advanced LAN Configuration provides LAN settings that are not declared manda tory by the LXI standard Ping Client provides the ping utility to verify the con...

Страница 603: ...instrument and another device The ping command uses the ICMP echo request and echo reply pack ets to determine whether the LAN connection is functional Ping is useful for diagnos ing IP network or ro...

Страница 604: ...nnection either temporarily or perma nently no special measures are necessary Windows automatically detects the inter ruption of the network connection and does not set up the connection when the inst...

Страница 605: ...Enter the user account name 4 Enter the old password 5 Enter the new password in the upper text line and repeat it in the following line 6 Press Enter The new password is now active Logging on to the...

Страница 606: ...a different user name activate the Connect using a different user name option The Map Network Drive dialog box is extended by the User name and Pass word fields 8 Enter your user name and password 9...

Страница 607: ...6 Select Share to confirm the settings 7 Select Done to close the dialog box The drive is shared and can be accessed by the selected users 11 3 Selecting the Operating Mode INSTrument CREate NEW 604 I...

Страница 608: ...d name INSTrument CREate NEW on page 604 Remote program example Chapter 11 17 16 Usage of Four Spectrum Instances on page 1005 Parameters Mode RECeiver Receiver mode SANalyzer Spectrum mode Spectrum 2...

Страница 609: ...h the suffix n XTIM SPECtrum Selects the IF analysis result display XTIM BARGraph Selects the bargraph Note that the Bargraph always has to be in window 1 XFRequency SPECtrum Selects the scan result d...

Страница 610: ...r off Parameters State ON OFF RST OFF Example DISP BARG PHOL ON Manual operation See Bargraph Maxhold on page 134 DISPlay BARGraph PHOLd RESet This command resets the maxhold value of the numeric indi...

Страница 611: ...the Max Peak detector QPEak Selects the Quasipeak detector RMS Selects the RMS detector RST AVERage Example DET REC POS AVER QPE Selects the peak average and quasipeak detectors Manual operation See B...

Страница 612: ...Rec Frequency on page 135 See Hold Final Measurement on page 137 INITiate n CONMeas This command resumes a scan that was interrupted by a transducer stop at the current reciever frequency If the scan...

Страница 613: ...u restart the measurement In single sweep mode you can synchronize to the end of the measurement with OPC OPC or WAI In continuous sweep mode synchronization to the end of the mea surement is not poss...

Страница 614: ...Event SENSe DETector t FMEasurement Detector This command selects the detector for the final measurement Parameters Detector AVERage Selects the Average detector CAVerage Selects the CISPR Average det...

Страница 615: ...ion on the TRACe DATA command and a detailed description of the characteristics of that command Basically the command queries the results of the current measurement The command supports various SCPI p...

Страница 616: ...le the scan is running The amount of returned results depends on the scan settings The type of returned data is as follows 4 byte trace status bit 0 to 9 represent the subscan bit 10 represents the la...

Страница 617: ...mit line value at corresponding fre quency If no limit line is active the delta value is set to 0 0 The trace has to be active for the command to work Example TRAC PLIST3 Querying results for the fina...

Страница 618: ...election ALL Selects all active traces for export to an ASCII file The trace parameter for the MMEMory STORe n TRACe command is ignored SINGle Only a single trace is selected for export namely the one...

Страница 619: ...ers ResultType TRACE1 TRACE3 Returns the level values displayed on the corresponding trace Example TRAC IF TRACE1 Queries the level for each trace point of trace 1 11 4 1 6 Demodulating Analog Signals...

Страница 620: ...e SCPI confirmed Manual operation See Squelch on page 139 See Squelch Level on page 392 SENSe DEMod SQUelch STATe State This command turns selective demodulation at the marker position on and off A vi...

Страница 621: ...e history buffer Suffix n 1 4 Selects the measurement window Example CALC SGR CLE Resets the result display and clears the memory Usage Event Manual operation See Spectrogram Clear on page 345 See Cle...

Страница 622: ...ing on page 377 DISPlay WINDow SGRam COLor STYLe ColorScheme This command sets the color scheme for the spectrogram Parameters ColorScheme HOT Uses a color range from blue to red Blue colors indicate...

Страница 623: ...sed on Parameters Trace Number of the trace The range depends on the result display Scans support six traces IF Analysis supports three traces Example CALC SGR TRAC 2 Assigns trace two to the spectrog...

Страница 624: ...tains the data for every frame in the history buffer The data corresponding to a particular frame begins with information about the frame number and the time that frame was recorded Note that dependin...

Страница 625: ...SGRam FRAMe Frame Time This command positions the marker on a particular frame The command is available for the spectrogram Suffix n 1 4 Selects the measurement window m 1 16 Selects the marker Parame...

Страница 626: ...rent peak level of the spectro gram The possible position depends on the marker search area The command is available for the spectrogram For more information see CALCulate n MARKer m SGRam SARea on pa...

Страница 627: ...s and positions marker 2 on the peak level of the y axis CALC MARK2 SGR Y MAX ABV Positions marker 2 on the next higher level found in the diagram area above the current marker position Usage Event CA...

Страница 628: ...ARKer m SGRam Y MAXimum PEAK This command positions the specified marker on the current peak level of the vertical axis The search is performed over all frames It does not change the horizontal posi t...

Страница 629: ...marker The command is available for the spectrogram Suffix n 1 4 Selects the measurement window m 1 16 Selects the marker Example CALC MARK2 SGR Y MIN Activates and positions marker 2 on the minimum...

Страница 630: ...kers Commands useful to work with markers in the spectrogram described elsewhere CALCulate n DELTamarker m MAXimum LEFT on page 666 CALCulate n DELTamarker m MAXimum NEXT on page 667 CALCulate n DELTa...

Страница 631: ...marker Parameters Frame Time Frame if time stamp is off Defines the distance of the deltamarker in frames and in relation to the position of marker 1 Note that in receiver mode it is not possible to...

Страница 632: ...possible position depends on the deltamarker search area The command is available for the spectrogram For more information see CALCulate n DELTamarker m SGRam SARea on page 628 Suffix n 1 4 Selects th...

Страница 633: ...rogram Suffix n 1 4 Selects the measurement window m 1 16 Selects the marker Example CALC DELT2 SGR Y MAX Activates and positions delta marker 2 on the peak level of the y axis CALC DELT2 SGR Y MAX AB...

Страница 634: ...available for the spectrogram Suffix n 1 4 Selects the measurement window m 1 16 Selects the marker Example CALC DELT2 SGR Y MAX Activates and positions delta marker 2 on the peak level of the y axis...

Страница 635: ...GRam Y MINimum BELow This command positions the specified delta marker on the next minimum level of the vertical axis The search includes only frames below the current marker position It does not chan...

Страница 636: ...ed delta marker on the current minimum level of the vertical axis The search is performed over all captured frames It does not change the horizontal position of the delta marker The command is availab...

Страница 637: ...Se FREQuency CENTer Frequency This command defines the receiver frequency for measurements in the frequency or time domain Parameters Frequency 0 to fmax Range 0 Hz to fmax RST fmax 2 Example FREQ CEN...

Страница 638: ...SCAN on how to define the frequency Time domain scans are available with options R S ESRP B50 and R S ESRP K53 RST TDOMain Example FREQ MODE TDOM Selects time domain scans Manual operation See Time Do...

Страница 639: ...Z DBPT DBPT_MHZ DBUV_M DBUV_MMHZ DBUA_M DBUA_MMHZ RST dB V Example CALC UNIT POW DBM Selects the unit dBm Manual operation See Unit on page 158 See dBx MHz on page 158 DISPlay WINDow n TRACe t Y SCALe...

Страница 640: ...ut ATTenuation Attenuation This command defines the attenuation level at the RF input To protect the input mixer attenuation levels of 10 dB or less are possible only if you have turned the input prot...

Страница 641: ...INP GAIN AUTO ON Includes the preamplifier into the auto range function Manual operation See Auto Preamp On Off on page 158 INPut GAIN STATe State This command turns the preamplifier on and off For th...

Страница 642: ...0 The available bandwidths depend on the current frequency span RST 3 kHz Example BAND IF 10KHZ Defines a resolution bandwidth of 10 kHz Manual operation See IF Analysis RBW on page 161 SENSe BANDwidt...

Страница 643: ...th BWIDth RESolution TYPE FilterType This command selects the type of resolution filter For detailed information on filters see Chapter 6 2 6 3 Selecting the Appropriate Filter Type on page 337 and Ch...

Страница 644: ...gles the RF input of the R S ESRP between AC and DC coupling Parameters CouplingType AC DC RST AC Example INP COUP DC Manual operation See Input AC DC on page 163 INPut TYPE Input The command selects...

Страница 645: ...on page 165 SENSe SWEep COUNt CURRent This command queries the number of scans that have started in single scan mode with more than one scan SENSe SWEep COUNt 1 Example SWE COUNt 64 sets scan count t...

Страница 646: ...0 RST 1000 Example TRAC POIN LIM 8000 Transfers a maximum of 8000 measurement values per trace with a single query 11 4 6 2 Scan Table SENSe SCAN range BANDwidth RESolution 643 SENSe SCAN range INPut...

Страница 647: ...INP ATT 30dB Defines an attenuation level of 30 dB for the 4th scan range SENSe SCAN range INPut ATTenuation AUTO State This command turns auto ranging in a particular scan range on and off Suffix ran...

Страница 648: ...arameters Input INPUT1 Selects RF input 1 INPUT2 Selects RF input 2 RST INPUT1 Example SCAN4 INP TYPE INPUT2 Selects RF input 2 as the signal source for the 4th scan range SENSe SCAN RANGes COUNt Rang...

Страница 649: ...le SCAN4 STEP 1MHz Defines a step size of 1 MHz for the 4th scan range SENSe SCAN range STOP Frequency This command defines the stop frequency of a particular scan range Suffix range 1 10 Selects the...

Страница 650: ...get the best measurement results RST LINear Example SWE SPAC LOG Selects logarithmic frequency steps Manual operation See Step Mode on page 166 11 4 6 3 Peak Search CALCulate n MARKer m PEXCursion 647...

Страница 651: ...ns have been fulfilled for that frequency Note that the frequency has to be in the displayed frequency range Suffix n irrelevant Parameters Frequency Numeric value in Hz The range depends on the measu...

Страница 652: ...ular number of peaks over the complete scan range RST PEAK Example CALC PEAK METH SUBR Divides the scan range into smaller subranges for the peak search Manual operation See Peak Search Mode Peaks Sub...

Страница 653: ...er of peaks per subrange depends on the number of subranges you have defined Range 1 to depends on the number of subranges RST 1 Example CALC PSE METH SUBR CALC PSE SUBR 20 CALC PSE SUBR PCO 5 Looks f...

Страница 654: ...ame String containg the file name The extension of the file is dat Example MMEM STOR PEAK D TEST DAT Manual operation See Peak List Export on page 171 11 4 6 5 Trace Configuration Commands useful to c...

Страница 655: ...p mode If calculation of average values is active selection between logarithmic and linear aver aging is possible For more detail see SENSe AVERage n TYPE on page 820 Suffix n window For applications...

Страница 656: ...mand turns the 150 kHz highpass filter of the ENV216 network for the final measurement on and off The command is available for the R S ENV216 network Parameters State ON OFF RST ON Example FME LISN FI...

Страница 657: ...re controllable FOURphase R S ESH2 Z5 four phases and protective earth are controlla ble OFF Turns off remote control of the LISN TWOPhase R S ESH3 Z5 two phases and protective earth are controllable...

Страница 658: ...network used for the premeasurement on and off It also selects the type of network in use Parameters Type ENV216 R S ENV 216 two phases and highpass are controllable ENV432 R S ENV 432 four phases ar...

Страница 659: ...INFo STATe 659 CALCulate n MARKer m COUPled STATe State This command couples or decouples the receiver frequency to the current marker fre quency Parameters State ON OFF RST OFF Example CALC MARK COUP...

Страница 660: ...the new receiver or center frequency and the span is adjusted according to the zoom factor Note that you should perform a complete measurement with synchronization to the end of the measurement This i...

Страница 661: ...rder Example CALC MARK X SLIM ON Switches the search limit function on CALC MARK X SLIM LEFT 10MHz Sets the left limit of the search range to 10 MHz Manual operation See Left Limit on page 182 See Lef...

Страница 662: ...N OFF RST OFF Example CALC MARK X SLIM ON Switches on search limitation Manual operation See Search Limits on page 182 See Search Lim Off on page 183 See Limits On Off on page 272 DISPlay WINDow n MIN...

Страница 663: ...mum LEFT This command positions a marker to the next smaller trace maximum on the left of the current position i e in descending X values If no next smaller maximum value is found on the trace level s...

Страница 664: ...uced Suffix n Selects the measurement window m depends on mode Selects the marker Example CALC MARK2 MAX Positions marker 2 to the maximum value of the trace Usage Event Manual operation See Peak on p...

Страница 665: ...T This command positions ae marker to the next higher trace minimum If no next higher minimum value is found on the trace level spacing to adjacent values peak excursion an execution error error code...

Страница 666: ...w m Selects the marker Example CALC MARK2 MIN Positions marker 2 to the minimum value of the trace CALC MARK2 MIN RIGH Positions marker 2 to the next higher minimum value to the right of the current v...

Страница 667: ...er m X Position This command positions a marker on a particular coordinate on the x axis If marker 2 3 or 4 is selected and used as delta marker it is switched to marker mode Suffix n Selects the meas...

Страница 668: ...his query returns the Real Q value of the marker first then the Imag I value Example INIT CONT OFF Switches to single sweep mode CALC MARK2 ON Switches marker 2 INIT WAI Starts a sweep and waits for t...

Страница 669: ...arker Suffix n Selects the measurement window m 1 irrelevant Parameters State ON OFF RST OFF Example CALC DELT LINK ON Manual operation See Link Mkr1 and Delta1 on page 379 CALCulate n DELTamarker m M...

Страница 670: ...tions the delta marker to the current trace maximum If necessary the corresponding delta marker is activated first Suffix n Selects the measurement window m Selects the marker Example CALC DELT3 MAX S...

Страница 671: ...1 CALCulate n DELTamarker m MINimum NEXT This command positions the delta marker to the next higher trace minimum The corre sponding delta marker is activated first if necessary If no next higher mini...

Страница 672: ...delta marker 1 to the next higher minimum value to the right of the current value Manual operation See Next Mode on page 181 CALCulate n DELTamarker m MODE Mode This command selects the delta marker...

Страница 673: ...ns delta marker 3 to trace 2 Manual operation See Marker to Trace on page 178 See Marker Wizard on page 178 See Marker Configuration on page 279 CALCulate n DELTamarker m X Position This command posit...

Страница 674: ...16 Marker Norm Delta on page 177 CALCulate n DELTamarker m Y This command queries the measured value of a delta marker The corresponding delta marker is activated if necessary The output is always a r...

Страница 675: ...s 674 11 4 8 1 Using Display Lines CALCulate n DLINe k 672 CALCulate n DLINe k STATe 672 CALCulate n DLINe k Position This command defines the vertical position of a display line Suffix n Selects the...

Страница 676: ...indow k 1 2 Selects the frequency line Parameters Frequency Range 0 Hz to fmax RST STATe to OFF Example CALC FLIN2 120MHz Manual operation See Frequency Line 1 Frequency Line 2 on page 189 CALCulate n...

Страница 677: ...CALCulate n LIMit k TRACe 676 CALCulate n LIMit k UNIT 676 CALCulate n LIMit k ACTive This command queries the names of all active limit lines Suffix n irrelevant k irrelevant Return values LimitLine...

Страница 678: ...e limit line Example CALC LIM1 COPY 2 Copies limit line 1 to line 2 CALC LIM1 COPY FM2 Copies limit line 1 to a new line named FM2 Manual operation See Copy to on page 187 CALCulate n LIMit k DELete T...

Страница 679: ...page 186 CALCulate n LIMit k UNIT Unit This command defines the unit of a limit line Upon selection of the unit DB the limit line is automatically switched to the relative mode For units different fr...

Страница 680: ...UPPer SHIFt 684 CALCulate n LIMit k UPPer SPACing 685 CALCulate n LIMit k UPPer STATe 685 CALCulate n LIMit k UPPer THReshold 685 CALCulate n LIMit k CONTrol DATA LimitLinePoints This command defines...

Страница 681: ...fix n Selects the measurement window k 1 8 Selects the limit line Parameters Mode ABSolute Limit line is defined by absolute physical values Hz or s RELative Limit line is defined by relative values r...

Страница 682: ...it line Parameters Distance Numeric value that defines the distance of the shift The unit depends on the scale of the x axis Example CALC LIM2 CONT SHIF 50KHZ Shifts all reference values of limit line...

Страница 683: ...unit CALC LIM2 LOW Outputs the lower limit values of limit line 2 separated by a comma Manual operation See Edit Value on page 187 CALCulate n LIMit k LOWer MARGin Margin This command defines an area...

Страница 684: ...defines the offset RST 0 Example CALC LIM2 LOW OFFS 3dB Shifts limit line 2 by 3 dB upwards Manual operation See Y Offset on page 188 CALCulate n LIMit k LOWer SHIFt Distance This command moves a com...

Страница 685: ...Lines on page 828 Suffix n irrelevant k 1 8 Selects the limit line Parameters State ON OFF RST OFF Example CALC LIM4 LOW STAT ON Switches on limit line 4 lower limit CALCulate n LIMit k LOWer THReshol...

Страница 686: ...IMit k UNIT RST LIMit STATe is set to OFF Example CALC LIM2 UPP 10 0 0 10 5 Defines 5 upper limit values for limit line 2 in the preset unit CALC LIM2 UPP Outputs the upper limit values for limit line...

Страница 687: ...does not actually change the limit line defi nition points Suffix n irrelevant k 1 8 Selects the limit line Parameters Offset Numeric value that defines the offset RST 0 Example CALC LIM2 UPP OFFS 3dB...

Страница 688: ...TATe State This command turns a lower limit line on and off Before you can use this command you have to select a limit line with CALCulate n LIMit k NAME The limit check is turned on separately For mo...

Страница 689: ...current limit check for all limit lines in all win dows Suffix n irrelevant k irrelevant Example CALC LIM CLE Deletes the result of the limit check Usage Event CALCulate n LIMit k FAIL This command q...

Страница 690: ...l operation See Select Traces to check on page 186 See Deselect All on page 186 11 5 Remote Commands in Spectrum Analyzer Mode The following remote commands configure and perform spectrum measurements...

Страница 691: ...specified power measurement type in the window specified by the suffix n The channel spacings and channel bandwidths are configured in the SENSe POWer subsystem Note If CPOWer is selected the number o...

Страница 692: ...9 See Ch Power ACLR on page 212 See C N on page 227 See C No on page 227 See OBW on page 230 CALCulate n MARKer m FUNCtion POWer RESult ResultType This command queries the result of the performed powe...

Страница 693: ...ndWidth Returns the occupied bandwidth as well as the position and level of the temporary markers T1 and T2 used to calculate the occupied bandwidth The syntax of the result is Occupied bandwidth T1 x...

Страница 694: ...rithmic scaling RANGE LOG the power is output in dBm with linear scaling RANGE LIN dB or LIN the power is output in W If SENSe POWer ACHannel MODE is set to REL the adjacent alternate channel power is...

Страница 695: ...ge 230 CALCulate n MARKer m FUNCtion POWer MODE Mode This commands defines the method by which the channel power values are calculated from the current trace in the window specified by the suffix n Su...

Страница 696: ...he channel bandwidths and the chan nel spacings selected in the active power measurement If necessary adjacent chan nel power measurement is switched on prior to the adjustment To obtain correct resul...

Страница 697: ...eference level to the measured channel power Manual operation See Adjust Ref Lvl on page 222 See Adjust Ref Lvl span 0 on page 231 TRACe n DATA Trace Data ResultType This command queries current trace...

Страница 698: ...ol lowing order No StartFreq StopFreq RBW PeakFreq PowerAbs PowerRel PowerDelta Limit Check Unused1 Unused2 No range number StartFreq StopFreq start and stop frequency of the range RBW resolution band...

Страница 699: ...ement window m Selects the marker Setting parameters Standard Note that predefined standards go without quotes while user standards have to be in quotes user standard Predefined standard Predefind sta...

Страница 700: ...DMA2000 Deletes the CDMA2000 standard Manual operation See Delete on page 221 CALCulate n MARKer m FUNCtion POWer STANdard SAVE Standard This command saves a cutomized ACLR standard The numeric suffix...

Страница 701: ...nnel bandwidth of the adjacent channel of the radio transmission system If the bandwidth of the adjacent channel is changed the band widths of all alternate adjacent channels are automatically set to...

Страница 702: ...adjacent channels are not influenced by this modification With SENSe source POWer HSPeed set to ON steep edged channel filters are available For further information on filters refer to Chapter 6 2 6...

Страница 703: ...annel Parameters Name RST TX 1 12 Example POW ACH NAME CHAN3 XYZ Defines the name of the third transmission channel as XYZ Manual operation See Names on page 216 SENSe POWer ACHannel SPACing ACHannel...

Страница 704: ...from 40 kHz to 100 kHz In consequence the spacing between the TX channel and all higher alternate adja cent channels is increased by the factor 100 40 2 5 ALT2 150 kHz ALT3 200 kHz ALT4 250 kHz Manual...

Страница 705: ...CHannel Alpha This command defines the roll off factor for the weighting filter of the adjacent channel Parameters Alpha numeric value RST 0 22 Example POW ACH FILT ALPH ACH 0 35 Manual operation See...

Страница 706: ...ATe ACHannel State This command activates the weighting filter for the adjacent channel Parameters State ON OFF RST OFF Example POW ACH FILT ACH ON Manual operation See Weighting Filter on page 216 SE...

Страница 707: ...ENSe POWer ACHannel REFerence TXCHannel AUTO 704 SENSe POWer ACHannel REFerence TXCHannel MANual 705 SENSe POWer ACHannel REFerence AUTO ONCE This command sets the reference value to the currently mea...

Страница 708: ...ference channel in relative adjacent channel power measurements The command is available only for multicarrier channel and adjacent channel power measurements with span 0 CALCulate n MARKer m FUNCtion...

Страница 709: ...evant Parameters LowerLimit UpperLimit first value 200DBM to 200DBM limit for the lower and the upper adjacent channel RST 200DBM Example CALC LIM ACP ACH ABS 35DBM 35DBM Sets the absolute limit value...

Страница 710: ...lute Limit on page 218 CALCulate n LIMit k ACPower ACHannel RELative LowerLimit UpperLimit This command defines the relative limit of the upper lower adjacent channel for adja cent channel power measu...

Страница 711: ...annel to 30 dB below the channel power CALC LIM ACP ACH ABS 35DBM 35DBM Sets the absolute limit value for the power in the lower and upper adjacent channel to 35 dBm CALC LIM ACP ON Switches on global...

Страница 712: ...lt in the adjacent channels Manual operation See Limit Checking on page 217 CALCulate n LIMit k ACPower ALTernate Channel ABSolute LowerLimit UpperLimit This command defines the absolute limit value f...

Страница 713: ...rnate channel Parameters State ON OFF RST OFF Example CALC LIM ACP ALT2 30DB 30DB Sets the relative limit value for the power in the lower and upper second alternate adjacent channel to 30 dB below th...

Страница 714: ...rst value 0 to 100dB limit for the lower and the upper alternate adjacent channel RST 0 DB Example CALC LIM ACP ALT2 30DB 30DB Sets the relative limit value for the power in the lower and upper second...

Страница 715: ...N Switches on the check of absolute limit values for the lower and upper second alternate adjacent channel INIT WAI Starts a new measurement and waits for the sweep end CALC LIM ACP ALT2 RES Queries t...

Страница 716: ...result in the second alternate adjacent channels Usage Query only CALCulate n LIMit k ACPower STATe State This command switches on and off the limit check for adjacent channel power mea surements The...

Страница 717: ...tput channel power is displayed in dBm RST OFF Example CALC MARK FUNC POW RES PHZ ON Output of results referred to the channel bandwidth For details on a complete measurement example refer to CALCulat...

Страница 718: ...tion Mode This command turns noise cancellation on and off If noise cancellation is on the R S ESRP performs a reference measurement to deter mine its inherent noise and subtracts the result from the...

Страница 719: ...1 REV CDMA IS95C Class 1 reverse CDMA 2000 CDMA 2000 TD SCDMA FWD TD SCDMA forward TD SCDMA REV TD SCDMA reverse WLAN 802 11A WLAN 802 11A WLAN 802 11B WLAN 802 11B WiMAX WiMAX WIBRO WIBRO RFID 14443...

Страница 720: ...peration See Power Bandwidth span 0 on page 231 11 5 1 4 Measuring Spurious Emissions The following commands configure and perform spurious emission measurements Controlling the Measurement 717 Managi...

Страница 721: ...RANGe range DELete 720 SENSe LIST RANGe range DETector 720 SENSe LIST RANGe range FILTer TYPE 720 SENSe LIST RANGe range FREQuency STARt 721 SENSe LIST RANGe range FREQuency STOP 721 SENSe LIST RANGe...

Страница 722: ...z Manual operation See VBW on page 266 SENSe LIST RANGe range BREak State This command configures the sweep behavior The suffixes are irrelevant Suffix range 1 20 range Parameters State ON The R S ESR...

Страница 723: ...on page 268 SENSe LIST RANGe range DETector Detector This command sets the detector for the specified range For details refer to Detector on page 266 Suffix range 1 20 range Parameters Detector NEGati...

Страница 724: ...If you set a span that is smaller than the overall span of the ranges the mea surement includes only the ranges that lie within the defined span and have a minimum span of 20 Hz The first and last ra...

Страница 725: ...rameters Value Refer to the data sheet RST 0 dB Example LIST RANG3 INP ATT 10 Sets the attenuation of range 3 to 10 dB Manual operation See RF Attenuator on page 266 SENSe LIST RANGe range INPut ATTen...

Страница 726: ...s setting is independent of the defined limit check type Suffix range 1 20 range Parameters Limit 400 to in 400 dBm RST 13 dBm Example LIST RANG1 LIM ABS STAR 10 Sets an absolute limit of 10 dBm at th...

Страница 727: ...e 267 SENSe LIST RANGe range POINts Value This command sets the number of sweep points for the specified range Suffix range 1 20 range Parameters Value For details on possible values refer to chapter...

Страница 728: ...ic sweep time setting for the speci fied range Suffix range 1 20 range Parameters State ON OFF RST ON Example LIST RANG3 SWE TIME AUTO OFF Deactivates the sweep time auto mode for range 3 Manual opera...

Страница 729: ...t the values peak values including the delta to a limit you have to switch on the limit again This command is only for FSP compatibility and not necessary to use on the R S ESRP Suffix n irrelevant Ex...

Страница 730: ...be stored in the list Manual operation See Peaks per Range on page 269 Retrieving Results The following commands analyze and retrieve measurement results for spurious emis sion measurements CALCulate...

Страница 731: ...directory include the relative path Example ESP PRES WCDMA 3GPP DL PowerClass_31_39 xml Selects the PowerClass_31_39 xml XML file in the C R_S instr sem_std WCDMA 3GPP DL directory ESP PRES W CDMA 3GP...

Страница 732: ...e INIT ESP Starts a Spectrum Emission Mask measurement Manual operation See Meas Start Stop on page 243 SENSe SWEep MODE Mode This command selects the spurious emission and spectrum emission mask meas...

Страница 733: ...um RANGe range LIMit source RELative STOP 736 SENSe ESPectrum RANGe range LIMit source STATe 737 SENSe ESPectrum RANGe range RLEVel 737 SENSe ESPectrum RANGe range SWEep TIME 737 SENSe ESPectrum RANGe...

Страница 734: ...ted see SENSe ESPectrum HighSPeed on page 730 Suffix range 1 20 range Parameters Value Refer to the data sheet RST 30 0 kHz Example ESP RANG2 BAND RES 5000 Sets the RBW for range 2 to 5 kHz Manual ope...

Страница 735: ...20 range Example ESP RANG4 DEL Deletes range 4 Manual operation See Delete Range on page 238 SENSe ESPectrum RANGe range FILTer TYPE Type This command sets the filter type for the specified range Note...

Страница 736: ...Range Set tings on page 245 Suffix range 1 20 range Parameters Frequency numeric value RST 250 0 MHz range 1 2 52 MHz range 2 2 52 MHz range 3 Example ESP RANG1 STAR 100000000 Sets the start frequenc...

Страница 737: ...d activates or deactivates the automatic RF attenuation setting for the specified range Note that this parameter can not be set for all ranges if Fast SEM mode is activated see SENSe ESPectrum HighSPe...

Страница 738: ...Range on page 238 SENSe ESPectrum RANGe range LIMit source ABSolute STARt Level This command sets an absolute limit value at the start frequency of the specified range Different from manual operation...

Страница 739: ...fix range 1 20 range Parameters Limit 400 to in 400 dBc RST 50 dBc Example ESP RANG3 LIM REL STAR 20 Sets a relative limit of 20 dBc at the start frequency of the range Manual operation See Rel Limit...

Страница 740: ...ck fails when one of the limits is violated RST REL Example ESP RANG3 LIM STAT AND Sets for all ranges the combined absolute relative limit check Manual operation See Limit Check 1 4 on page 237 SENSe...

Страница 741: ...weep time setting for the speci fied range Note that this parameter can not be set for all ranges if Fast SEM mode is activated see SENSe ESPectrum HighSPeed on page 730 Suffix range 1 20 range Parame...

Страница 742: ...m RTYPe on page 740 Parameters Bandwidth minimum span value span of reference range RST 3 84 MHz Example ESP RTYP CPOW Sets the power reference type to channel power ESP BWID 1 MHZ Sets the Tx bandwid...

Страница 743: ...the current position number of the reference range SENSe ESPectrum RTYPe Type This command sets the power reference type Parameters Type PEAK CPOWer PEAK Measures the highest peak within the referenc...

Страница 744: ...and 200 in ascending order separated by commas 200 Example CALC LIM ESP LIM 50 50 70 Defines the following power classes 200 50 50 50 50 70 70 200 Query CALC LIM ESP LIM Response 200 50 50 70 200 CALC...

Страница 745: ...ossible to use power classes for which limits are defined Also either only one power class at a time or all power classes together can be selected Suffix n irrelevant k irrelevant Class 1 4 the power...

Страница 746: ...limit always ends at 200 dBm i e the upper limit of the last power class can not be set If more than one power class is in use the upper limit must equal the lower limit of the next power class Suffix...

Страница 747: ...a value Depending on the entered value one of the predefined limit lines is selected Suffix n 1 4 window k irrelevant Parameters Power 33 28 0 33 P 33 28 28 P 33 0 P 28 RST 0 Example CALC LIM ESP VAL...

Страница 748: ...n page 269 See ASCII File Export on page 269 CALCulate n ESPectrum PSEarch PEAKsearch IMMediate This command starts the list evaluation Suffix n Selects the measurement window Example CALC ESP PSE Usa...

Страница 749: ...n page 686 CALCulate n MARKer m FUNCtion POWer RESult on page 689 TRACe n DATA on page 694 11 5 1 6 Performing EMI Measurements The following commands configure and perform EMI measurements Configurin...

Страница 750: ...LCulate n MARKer m TRACe on page 664 Remote commands exclusive to EMI measurements CALCulate n DELTamarker m FUNCtion FMEasurement DETector 747 CALCulate n MARKer m FUNCtion FMEasurement DETector 748...

Страница 751: ...relevant m marker Parameters Detector OFF turns the marker off AVER average detector CAV CISPR Average detector CRMS RMS Average detector POS maximum peak detector QPE quasipeak detector RST OFF Manua...

Страница 752: ...e n MARKer m FUNCtion FMEasurement LIMit k CONDition 750 CALCulate n MARKer m FUNCtion FMEasurement LIMit k DELTa 751 CALCulate n MARKer m FUNCtion FMEasurement RESult 751 CALCulate n DELTamarker m FU...

Страница 753: ...the second limit line Usage Query only Manual operation See Marker Table on page 281 CALCulate n DELTamarker m FUNCtion FMEasurement RESult Result This command queries the result of the EMI measureme...

Страница 754: ...This command queries the vertical distance from the marker position to the limit line The unit is dB If the marker has been assigned to a different trace than the limit line the command returns 200 S...

Страница 755: ...nitiates a search for the signals required for the AM depth measure ment Note that the command does not perform a new measurement but looks for the sig nals on the current trace Suffix n Selects the m...

Страница 756: ...level The level value of marker 1 is regarded as the carrier level On activating the function marker 2 and marker 3 are automatically set as delta markers symmetrically to the car rier to the adjacen...

Страница 757: ...results described elsewhere TRACe n DATA on page 694 CALCulate n MARKer m FUNCtion HARMonics BANDwidth AUTO 754 CALCulate n MARKer m FUNCtion HARMonics DISTortion 755 CALCulate n MARKer m FUNCtion HA...

Страница 758: ...Switches to single sweep mode CALC MARK FUNC HARM NHARM 3 Sets the number of harmonics to be measured to 3 CALC MARK FUNC HARM ON Activates the harmonic distortion measurement INIT WAI Starts a sweep...

Страница 759: ...c Distortion on page 201 See Harmonic Distortion on page 311 CALCulate n MARKer m FUNCtion HARMonics NHARmonics NoHarmonics This command sets the number of harmonics to be measured Suffix n Selects th...

Страница 760: ...rmined is defined by the last span If you perform the measurement in the time domain the current center frequency is used as the frequency of the first harmonic Thus the frequency search is bypassed T...

Страница 761: ...mmand initiates a search for signals in the current trace to determine the third intercept point No new measurement is performed Only the currently available trace selected for the TOI measurement is...

Страница 762: ...rement To get a valid result you have to perform a complete measurement with synchroniza tion to the end of the measurement before reading out the result This is only possible for single sweeps Suffix...

Страница 763: ...amplitude distribution APD On activating this function the CCDF measurement is switched off Suffix n irrelevant Parameters State ON OFF RST OFF Example CALC STAT APD ON Switches on the APD measuremen...

Страница 764: ...ge 664 TRACe n DATA on page 694 CALCulate n STATistics CCDF X Trace 761 CALCulate n STATistics RESult Trace 762 CALCulate n STATistics CCDF X Trace Probability This command queries the results of the...

Страница 765: ...or The required result is selected via the following parameters Example CALC STAT RES2 ALL Reads out the three measurement results of trace 2 Example of answer string 5 56 19 25 13 69 i e mean power 5...

Страница 766: ...Gate Ranges on page 291 SENSe SWEep EGATe TRACe k STARt range Value This command defines the starting point for the range to be traced using gated trigger ing Suffix k 1 6 trace range 1 3 range Param...

Страница 767: ...ace 1 at 5 ms Manual operation See Gate Ranges on page 291 Scaling the Diagram The following commands configure the diagram for statistical measurements CALCulate n STATistics PRESet 764 CALCulate n S...

Страница 768: ...d mean power for the CCDF measurement In addition the probability scale for the number of test points is adapted Subsequent commands have to be synchronized with WAI OPC or OPC to the end of the auto...

Страница 769: ...TAT SCAL X RLEV 60dBm Manual operation See x Axis Ref Level on page 288 CALCulate n STATistics SCALe Y LOWer Value This command defines the lower limit for the y axis of the diagram in statistical mea...

Страница 770: ...surements of the time domain power CALCulate n MARKer m FUNCtion SUMMary AOFF 767 CALCulate n MARKer m FUNCtion SUMMary AVERage 767 CALCulate n MARKer m FUNCtion SUMMary MODE 768 CALCulate n MARKer m...

Страница 771: ...ALCulate n MARKer m FUNCtion SUMMary MODE Mode This command selects absolute or relative power measurement in zero span The reference power for relative measurement is defined with CALCulate n MARKer...

Страница 772: ...r m FUNCtion SUMMary AVERage or CALCulate n MARKer m FUNCtion SUMMary PHOLd is switched on the cur rent value is the accumulated measurement value at the time considered Suffix n Selects the measureme...

Страница 773: ...SUMMary RMS AVERage RESult 774 CALCulate n MARKer m FUNCtion SUMMary RMS PHOLd RESult 775 CALCulate n MARKer m FUNCtion SUMMary RMS RESult 776 CALCulate n MARKer m FUNCtion SUMMary RMS STATe 776 CALC...

Страница 774: ...chroniza tion to the end of the measurement before reading out the result This is only possible for single sweeps Suffix n Selects the measurement window m Selects the marker Return values MeanPower M...

Страница 775: ...is positioned In order to evaluate another trace marker 1 must be positioned on another trace with CALCulate n MARKer m TRACe Suffix n Selects the measurement window m Selects the marker Parameters S...

Страница 776: ...ed previously using CALCulate n MARKer m FUNCtion SUMMary PHOLd To get a valid result you have to perform a complete measurement with synchroniza tion to the end of the measurement before reading out...

Страница 777: ...See Peak on page 272 CALCulate n MARKer m FUNCtion SUMMary PPEak STATe State This command turns the measurement of the positive peak time domain power on and off Note The measurement is performed on t...

Страница 778: ...ARKer m FUNCtion SUMMary RMS PHOLd RESult This command queries the maximum RMS time domain power The query is only pos sible if the peak hold function has been activated previously using CALCulate n M...

Страница 779: ...RK FUNC SUMM RMS ON Switches on the function INIT WAI Starts a sweep and waits for the end CALC MARK FUNC SUMM RMS RES Outputs the result Usage Query only Manual operation See RMS on page 272 CALCulat...

Страница 780: ...sweep mode CALC MARK FUNC SUMM SDEV ON Switches on the function CALC MARK FUNC SUMM AVER ON Switches on the calculation of average INIT WAI Starts a sweep and waits for the end CALC MARK FUNC SUMM MEA...

Страница 781: ...ngle sweeps Suffix n Selects the measurement window m Selects the marker Return values StandardDeviation Standard deviation of the signal during the measurement time Example INIT CONT OFF Switches to...

Страница 782: ...trigger signal or by the video signal An individ ual trigger event is required for each burst to be measured If an external trigger signal is used the threshold is fixed to TTL level while with a vid...

Страница 783: ...er SENSe MPOWer RESult LIST This command queries the results of the pulse power measurement This command may be used to obtain measurement results in an asynchronous way using the service request mech...

Страница 784: ...If you use the command repeatedly the R S ESRP only changes those settings that you have actually changed before initiating the measurement Thus measurement times are kept as low as possible If you s...

Страница 785: ...vel The trigger level is available for the video trigger In that case the level is a percentage of the diagram height In case of an external trigger the R S ESRP uses a fix TTL level Range 0 PCT to 10...

Страница 786: ...on trigger FREE RUN The results are output as a list in the order of the entered frequency points The num ber of results per test point depends on the number of concurrently active measure ments peak...

Страница 787: ...e command returns an error Return values PowerLevel Power level for each frequency included in the measurement The command returns up to 3 power levels for each frequency depending on the number of ev...

Страница 788: ...unction is only available in REMOTE operation It is deactivated when switching the instrument back to LOCAL Note that using the command as a query initiates the measurement and returns the results if...

Страница 789: ...th SENSe LIST POWer SET on page 787 The result is a list of floating point values separated by commas The unit depends on CALCulate n UNIT POWer Example SENSe LIST POWer 935 2 MHz 0 dBm 10 dB OFF NORM...

Страница 790: ...Freq MHz Ref Level dBm RF Att dB el Att dB Filter type RBW VBW Meas Time us TRG Level reserve d 1 935 2 0 10 OFF Normal 1 MHz 3 MHz 440 0 2 935 4 0 10 10 Channel 30 kHz 100 kHz 440 0 3 935 6 0 10 20 C...

Страница 791: ...an 0 Hz sweep time to 30 s span 10 Hz 31 25 ns to 30 s RST 0 s GateLength Defines the gate length for gated measurements Setting 0 seconds turns gated measurements off Other values activate the gated...

Страница 792: ...ich action is taken when the probe s micro button is pressed Parameters Mode RSINgle A single sweep is performed NOACtion No action is taken RST OFF Example PROB SET STAT ON Manual operation See Probe...

Страница 793: ...Configuration 811 11 5 2 1 Frequency Settings The following commands select the frequency characteristics the R S ESRP uses for the measurement Defining Frequency Characteristics Commands useful to d...

Страница 794: ...s the coupling of the step size to the span SENSe FREQuency CENTer STEP LINK CouplingType This command couples and decouples the center frequency step size to the span or the resolution bandwidth Para...

Страница 795: ...uency mode between span 0 and zero span in the Spectrum mode Parameters Mode CW Selects the time domain span 0 In the time domain define the measurement frequency with SENSe FREQuency CENTer FIXed Sel...

Страница 796: ...Se FREQuency STOP Frequency This command defines the stop frequency for measurements in the frequency domain Parameters Frequency min span to fmax RST fmax Example FREQ STOP 2000 MHz Manual operation...

Страница 797: ...dth 10 Hz to MAX span RST span 10 on activating the function Example CALC MARK FUNC STR BAND 1 MHZ Sets the search bandwidth to 1 MHz CALC MARK FUNC STR BWID 1 MHZ Alternative command for the same fun...

Страница 798: ...MARKer m FUNCtion STRack TRACe TraceNumber This command defines the trace on which the signal is tracked Suffix n Selects the measurement window m Selects the marker Parameters TraceNumber Range 1 to...

Страница 799: ...adjusted when the signal level falls below 18 dBm Manual operation See Lower Level Hysteresis on page 329 SENSe ADJust CONFiguration HYSTeresis UPPer Threshold This command defines an upper threshold...

Страница 800: ...manual definition of the measurement duration ADJ CONF LEV DUR 5 Specifies the duration manually SENSe ADJust CONFigure TRIG State This command defines the behaviour of the measurement when adjusting...

Страница 801: ...ful to configure the RBW described elsewhere SENSe BANDwidth BWIDth RESolution on page 639 SENSe BANDwidth BWIDth RESolution TYPE on page 640 SENSe BANDwidth BWIDth RESolution AUTO 798 SENSe BANDwidth...

Страница 802: ...T filter Example BAND FFT NARR Select narrow partial span for FFT filter Manual operation See Auto on page 330 See Narrow on page 330 SENSe BANDwidth BWIDth RESolution RATio Ratio This command defines...

Страница 803: ...deo BW Auto on page 333 See Default Coupling on page 336 SENSe BANDwidth BWIDth VIDeo RATio Ratio This command defines the ratio between video bandwidth Hz and resolution band width Hz Note that the r...

Страница 804: ...n page 609 INITiate n IMMediate on page 610 INITiate n CONTinuous on page 609 INITiate n CONMeas 801 SENSe SWEep COUNt 802 SENSe SWEep POINts 802 SENSe SWEep TIME 803 SENSe SWEep TIME AUTO 803 SENSe S...

Страница 805: ...calculating the average or maximum value If the values 0 or 1 are set one sweep is performed Parameters NumberSweeps 0 to 32767 RST 0 GSM 200 PHN 1 Example SWE COUN 64 Sets the number of sweeps to 64...

Страница 806: ...command automatically sets the sweep time to 32 ms In analyzer mode this command controls the automatic coupling of the sweep time to the frequency span and bandwidth settings If SENSe SWEep TIME is...

Страница 807: ...ands configure triggered or gated measurements Triggering Measurements TRIGger n SEQuence HOLDoff TIME 804 TRIGger n SEQuence IFPower HOLDoff 805 TRIGger n SEQuence IFPower HYSTeresis 805 TRIGger n SE...

Страница 808: ...IQ mode measurements Suffix n irrelevant Parameters Value RST 150 ns Example TRIG SOUR IFP Sets the IF power trigger source TRIG IFP HOLD 200 ns Sets the holding time to 200 ns Manual operation See T...

Страница 809: ...ample TRIG LEV IFP 30DBM Manual operation See Trg Gate Level on page 348 See Trigger Level on page 443 TRIGger n SEQuence LEVel RFPower TriggerLevel This command defines the power level at the third i...

Страница 810: ...e n IQ EGATe TYPE on page 865 SENSe SWEep EGATe TYPE and an IFP trigger the holdoff time for the IFP trigger see TRIGger n SEQuence HOLDoff TIME on page 804 is ignored for frequency sweep FFT sweep ze...

Страница 811: ...Source on page 346 See Free Run on page 346 See External on page 346 See Video on page 347 See RF Power on page 347 See IF Power BB Power on page 347 See Time on page 348 TRIGger n SEQuence TIME RINTe...

Страница 812: ...ed Parameters State ON OFF RST OFF Example SWE EGAT ON Switches on the external gate mode SWE EGAT TYPE EDGE Switches on the edge triggered mode SWE EGAT HOLD 100US Sets the gate delay to 100 s SWE EG...

Страница 813: ...le SWE EGAT POL POS Manual operation See Trg Gate Polarity on page 348 See Trigger Polarity on page 443 SENSe SWEep EGATe SOURce Source This command selects the signal source for gated measurements If...

Страница 814: ...rements EDGE The gate is edge triggered After detection of the set gate signal edge the gate remains open until the gate delay SENSe SWEep EGATe HOLDoff has expired RST EDGE Example SWE EGAT TYPE EDGE...

Страница 815: ...ing with DISPlay WINDow n TRACe t Y SPACing on page 636 Suffix n window For applications that do not have more than 1 measure ment window the suffix n is irrelevant t irrelevant Parameters Range Range...

Страница 816: ...he reference level With the reference level offset 0 the value range of the reference level is modified by the offset The unit depends on the setting defined with CALCulate n UNIT POWer Suffix n irrel...

Страница 817: ...tion See Ref Level Position on page 326 See Reference Value Position on page 449 INPut ATTenuation AUTO State This command automatically couples the input attenuation to the reference level state ON o...

Страница 818: ...DATA 818 FORMat DEXPort DSEParator 818 MMEMory STORe n TRACe 819 SENSe AVERage n COUNt 819 SENSe AVERage n STATe Trace 820 SENSe AVERage n TYPE 820 SENSe WINDow DETector trace FUNCtion 821 SENSe WIND...

Страница 819: ...ects linear averaging for trace math calculations Manual operation See Lin on page 365 See Log on page 366 See Power on page 366 CALCulate n MATH POSition Position This command defines the position of...

Страница 820: ...atically to avoid false results e g with span changes For applications that require no reset after parameter changes the automatic reset can be switched off Suffix n window For applications that do no...

Страница 821: ...SCII data are transmitted in plain text separated by commas REAL REAL data are transmitted as 32 bit IEEE 754 floating point numbers in the definite length block format RST ASCII Example FORM REAL 32...

Страница 822: ...ation of the path complies with DOS conventions Example MMEM STOR TRAC 3 TEST ASC Stores trace 3 in the file TEST ASC Manual operation See ASCII Trace Export on page 175 SENSe AVERage n COUNt NoMeasur...

Страница 823: ...ST OFF Example AVER OFF Switches off the average calculation for trace 1 AVER STAT3 ON Switches on the average calculation for trace 3 SENSe AVERage n TYPE FunctionType This command selects the type o...

Страница 824: ...OS Sets the detector to positive peak Manual operation See Trace 1 6 on page 175 See Auto Peak on page 363 See Positive Peak on page 363 See Negative Peak on page 363 See Sample on page 363 See RMS on...

Страница 825: ...you use it as a setting command it transfers trace data from an external source to the R S ESRP The data format depends on FORMat DATA Parameters Trace Selects the trace to write data to TRACE1 TRACE6...

Страница 826: ...n bandwidth PeakFreq frequency of the peak in a range PowerAbs absolute power of the peak in dBm PowerRel power of the peak in relation to the channel power in dBc PowerDelta distance from the peak to...

Страница 827: ...ful to work with markers in the spectrogram described elsewhere CALCulate n SGRam CLEar IMMediate on page 618 CALCulate n SGRam HDEPth on page 620 CALCulate n SGRam TSTamp DATA on page 620 CALCulate n...

Страница 828: ...ts for the end of the sweep CALC SGR CONT ON Repeats the single sweep measurement without deleting the results of the last measurement Manual operation See Continue Frame On Off on page 344 CALCulate...

Страница 829: ...nuous sweep CALC SGR FRAM SEL 25 Selects frame number 25 Manual operation See Select Frame on page 344 CALCulate n SGRam TSTamp STATe State This command activates and deactivates the time stamp If the...

Страница 830: ...sewhere CALCulate n MARKer m MAXimum LEFT on page 660 CALCulate n MARKer m MAXimum NEXT on page 660 CALCulate n MARKer m MAXimum PEAK on page 661 CALCulate n MARKer m MAXimum RIGHt on page 661 CALCula...

Страница 831: ...r m SGRam SARea on page 628 CALCulate n DELTamarker m SGRam XY MAXimum PEAK on page 629 CALCulate n DELTamarker m SGRam XY MINimum PEAK on page 629 CALCulate n DELTamarker m SGRam Y MAXimum ABOVe on p...

Страница 832: ...nd off Suffix n irrelevant Line 1 2 Selects the time line Parameters State ON OFF RST OFF Example CALC TLIN2 STAT ON Manual operation See Time Line 1 Time Line 2 on page 189 Using Limit Lines For comm...

Страница 833: ...X SLIMits RIGHT on page 658 CALCulate n MARKer m X SLIMits STATe on page 659 CALCulate n MARKer m Y on page 664 CALCulate n MARKer m FUNCtion REFerence 830 CALCulate n MARKer m LOEXclude 831 CALCulat...

Страница 834: ...race level spacing to adjacent values peak excursion an execution error error code 200 is produced Suffix n Selects the measurement window m Selects the marker Parameters State ON OFF RST OFF Example...

Страница 835: ...CALCulate n MARKer m X SLIMits STATe Suffix n irrelevant m irrelevant Parameters State ON OFF RST OFF Example CALC MARK X SLIM ZOOM ON Switches the search limit function on CALC MARK X SLIM RIGH 20MH...

Страница 836: ...m X Suffix n Selects the measurement window m Selects the marker Parameters Probability Range 0 to 100 Default unit PCT Example CALC1 MARK Y PERC 95PCT Positions marker 1 to a probability of 95 Manua...

Страница 837: ...e on page 179 Using Delta Markers Commands useful to control delta markers described elsewhere CALCulate n DELTamarker m AOFF on page 666 CALCulate n DELTamarker m LINK on page 666 CALCulate n DELTama...

Страница 838: ...n See Marker Wizard on page 178 Controlling Fixed Reference Markers CALCulate n DELTamarker m FUNCtion FIXed RPOint MAXimum PEAK 835 CALCulate n DELTamarker m FUNCtion FIXed RPOint X 836 CALCulate n D...

Страница 839: ...ef Point Time zero span on page 389 CALCulate n DELTamarker m FUNCtion FIXed RPOint Y RefPointLevel This command defines the vertical position of the fixed delta marker reference point The coordinates...

Страница 840: ...the CALCulate n DELTamarker m FUNCtion FIXed RPOint X commands and CALCulate n DELTamarker m FUNCtion FIXed RPOint Y independently of the position of marker 1 and of a trace It applies to all delta ma...

Страница 841: ...measurement before reading out the result This is only possible for single sweeps Suffix n Selects the measurement window m irrelevant Parameters State ON OFF RST OFF Example INIT CONT OFF Switches to...

Страница 842: ...ency counter marker 1 Suffix n Selects the measurement window m irrelevant Parameters Resolution 0 001 0 01 0 1 1 10 100 1000 10000 Hz RST 0 1 Hz Example CALC MARK COUN RES 1kHz Sets the resolution of...

Страница 843: ...marker position The result can be queried with CALCulate n MARKer m FUNCtion NOISe RESult Suffix n Selects the measurement window m Selects the marker Parameters State ON OFF RST OFF Example CALC MAR...

Страница 844: ...ge Query only Manual operation See Phase Noise On Off on page 389 CALCulate n DELTamarker m FUNCtion PNOise STATe State This command turns the phase noise measurement at the delta marker position on a...

Страница 845: ...odulation HOLDoff 842 CALCulate n MARKer m FUNCtion DEModulation SELect 843 CALCulate n MARKer m FUNCtion DEModulation STATe 843 CALCulate n MARKer m FUNCtion DEModulation CONTinuous State This comman...

Страница 846: ...ALC MARK FUNC DEM SEL FM Manual operation See AM on page 391 See FM on page 391 CALCulate n MARKer m FUNCtion DEModulation STATe State This command turns the audio demodulator on and off when the meas...

Страница 847: ...on NDBDown TIME The bandwidth between the markers can be queried with CALCulate n MARKer m FUNCtion NDBDown RESult Suffix n Selects the measurement window m irrelevant Parameters Distance Distance of...

Страница 848: ...surements Suffix n Selects the measurement window m Selects the marker Example INIT CONT OFF Switches to single sweep mode CALC MARK FUNC NDBD ON Switches on the n dB down function INIT WAI Starts a s...

Страница 849: ...al operation See n dB down on page 392 CALCulate n MARKer m FUNCtion NDBDown STATe State This command turns the n dB Down marker function on and off Suffix n Selects the measurement window m irrelevan...

Страница 850: ...Culate n MARKer m PEXCursion on page 647 MMEMory STORe PEAKlist on page 651 CALCulate n MARKer m FUNCtion FPEaks ANNotation LABel STATe 847 CALCulate n MARKer m FUNCtion FPEaks COUNt 848 CALCulate n M...

Страница 851: ...with Position of a peak on the x axis CALCulate n MARKer m FUNCtion FPEaks X Position of a peak on the y axis CALCulate n MARKer m FUNCtion FPEaks Y Number of peaks in the list CALCulate n MARKer m FU...

Страница 852: ...und CALC MARK FUNC FPE X Queries the frequencies span 0 or time span 0 of max ima found CALCulate n MARKer m FUNCtion FPEaks LIST SIZE MaxNoPeaks This command defines the maximum number of peaks the m...

Страница 853: ...creasing position on the x axis Y Sorts the peaks according to decreasing position on the y axis Example CALC MARK FUNC FPE SORT Y Sets the sort mode to decreasing y values Manual operation See Sort M...

Страница 854: ...Queries the number of maxima found CALC MARK FUNC FPE X Queries the frequencies span 0 or time span 0 of the maxima found 107 5E6 153 8E6 187 9E6 frequencies in increasing order 2 05E 3 2 37E 3 3 71e...

Страница 855: ...e 853 CALCulate n DELTamarker m FUNCtion BPOWer MODE 854 CALCulate n DELTamarker m FUNCtion BPOWer RESult 854 CALCulate n DELTamarker m FUNCtion BPOWer SPAN 854 CALCulate n DELTamarker m FUNCtion BPOW...

Страница 856: ...nt m irrelevant Parameters Span Numeric value that defines the span in Hz The maximum span depends on the marker position and R S FSV model RST 5 of current span Example CALC MARK FUNC BPOW SPAN 20 Ma...

Страница 857: ...OW MODE DENS CALCulate n DELTamarker m FUNCtion BPOWer RESult This command queries the results of the band power measurement Suffix n irrelevant m irrelevant Return values Power Signal power over the...

Страница 858: ...vant m irrelevant Parameters State ON OFF When switched to on if no marker is active yet marker 1 is acti vated Otherwise the currently active marker is used as a delta band power marker all other mar...

Страница 859: ...tude Magnitude FREQuency Spectrum VECTor IQ Vector Manual operation See I Q Analyzer on page 430 See Display Config on page 431 CALCulate n MARKer m SEARch MarkRealImag This command selects the trace...

Страница 860: ...AD IQ STATe 1 FileName This command loads the I Q data from the specified iq tar file Note switch to single sweep mode INIT CONT OFF before importing I Q data as otherwise the instrument will continue...

Страница 861: ...r conditions and the record length If this command is omitted the current instrument settings are used for the correspond ing parameters This command switches to IQ mode automatically see also TRACe n...

Страница 862: ...ST 128 Example TRAC IQ SET NORM 10 MHz 32 MHz EXT POS 0 2048 Reads 2048 I Q values starting at the trigger point sample rate 32 MHz trigger External slope Positive TRAC IQ SET NORM 10 MHz 4 MHz EXT PO...

Страница 863: ...are scaled linearly in Volts and correspond to the voltage at the RF or digital input of the instrument The number of the returned values is 2 the number of samples Note Using the command with the RST...

Страница 864: ...MPatible IQBLock IQPair RST IQBL For maximum performance the formats Compatible or IQPair should be used Furthermore for large amounts of data the data should be in binary format to improve performanc...

Страница 865: ...ds out the averaged data Usage SCPI confirmed TRACe n IQ AVERage STATe State The command enables averaging of the recorded I Q data provided that I Q data acquisition was previously enabled with TRACe...

Страница 866: ...in unit Volt and correspond to the voltage at the RF or digital input of the instrument The format of the output buffer corresponds to the TRACe n IQ DATA command The command requires all response da...

Страница 867: ...TA MEM 2048 1024 Reads 1024 I Q data from half of the recorded data TRAC IQ DATA MEM 100 512 Reads 512 I Q data starting at the trigger point Pretrigger Samples was 100 Usage Query only TRACe n IQ EGA...

Страница 868: ...samples defined by TRACe n IQ SET sample rate defined by TRACe n IQ SRATe Range 1 Max samples RST 100 Example TRAC IQ EGAT LENG 2000 TRACe n IQ EGATe NOFgateperiods Number Defines the number of gate...

Страница 869: ...number of samples for RF input is 200 MS For details see Chapter 7 5 Working with I Q Data on page 444 Suffix n irrelevant Parameters State ON OFF ON all functions of the IQ Analyzer are available OFF...

Страница 870: ...le Suffix n irrelevant Example TRAC IQ TPIS Usage Query only Manual operation See Trigger Offset on page 349 11 6 2 I Q Gating Sometimes in spectrum analysis measurements are based on a certain length...

Страница 871: ...les The number of complex samples to be recorded prior to the trigger event can be selected see TRACe n IQ SET on page 858 for all available trigger sources except for Free Run SCPI command TRACE IQ E...

Страница 872: ...ge SCPI confirmed Manual operation See Source RF Internal On Off on page 448 SENSe CORRection COLLect ACQuire MeasType This command initiates a reference measurement calibration The reference mea sure...

Страница 873: ...ents TRANsmission Selects transmission measurements RST TRANsmission Example CORR METH TRAN Sets the type of measurement to transmission Manual operation See Calibrate Transmission on page 448 See Cal...

Страница 874: ...acking generator option R S FSV B9 Suffix n 1 4 Selects the measurement window Parameters State ON OFF RST OFF Example SOUR2 AM STAT ON Switches on the external amplitude modulation of the tracking ge...

Страница 875: ...quency The possible value range is 100 Hz to 10 MHz in steps of 10 Hz Range 100 Hz to 10 MHz Increment 10 Hz RST 100 Hz Example SOUR1 FM DEV 1 MHz Sets the maximum frequency deviation of the tracking...

Страница 876: ...Hz to 200 MHz RST 0 Hz Example SOUR FREQ OFFS 10MHz Defines a frequency offset of the tracking generator of 10 MHz Usage SCPI confirmed Manual operation See Frequency Offset on page 458 SOURce n POWer...

Страница 877: ...er sweep is on the R S ESRP enters the time domain span 0 Hz During the sweep time the power at the internal tracking generator is changed linearly from start power to stop power The start and stop po...

Страница 878: ...nerator POWer LEVel 879 SOURce n EXTernal generator ROSCillator SOURce 879 SOURce n EXTernal generator STATe 880 SYSTem COMMunicate GPIB RDEVice GENerator generator ADDRess 881 SYSTem COMMunicate RDEV...

Страница 879: ...Source frequency of the external tracking generator RST RST value Example SOUR EXT2 FREQ 10MHz Manual operation See Fixed Generator Frequency on page 459 SOURce n EXTernal generator FREQuency COUPling...

Страница 880: ...meters Value numeric value RST 1 Example SOUR EXT FREQ NUM 4 SOUR EXT FREQ DEN 3 Sets a multiplication factor of 4 3 i e the transmit frequency of the generator is 4 3 times the analyzer frequency Mod...

Страница 881: ...or the External Tracking Generator option R S FSV B10 Suffix n irrelevant generator 1 2 tracking generator For Noise Figure measurements K30 1 only m 1 2 offset for Noise Figure measurements K30 only...

Страница 882: ...Level This command sets the output power of the selected generator This command is only valid for the External Tracking Generator option R S FSV B10 Suffix n irrelevant generator 1 2 tracking generat...

Страница 883: ...cillator Mode A NF Manual operation See Reference on page 462 SOURce n EXTernal generator STATe State This command activates or deactivates the external generator indicated by the suffix generator Thi...

Страница 884: ...OMM GPIB RDEV GEN1 ADDR 15 Mode A NF SYSTem COMMunicate RDEVice GENerator generator INTerface Type Defines the interface used for the connection to the external tracking generator This command is only...

Страница 885: ...ly Parameters Type GPIB TTL GPIB GPIB connection without TTL synchronization for all generators of other manufacturers and some Rohde Schwarz devices TTL GPIB connection with TTL synchronization if av...

Страница 886: ...etters Many common commands are related to the Status Reporting System Available common commands CAL 883 CLS 884 ESE 884 ESR 884 IDN 884 IST 884 OPC 885 OPT 885 PCB 885 PRE 885 PSC 885 RST 886 SRE 886...

Страница 887: ...tents of the event status enable register in decimal form Parameters Value Range 0 to 255 ESR Event status read Returns the contents of the event status register in decimal form and subsequently sets...

Страница 888: ...ress Pass control back Indicates the controller address to which remote control is returned after termination of the triggered action Setting parameters Address Range 0 to 30 Usage Setting only PRE Va...

Страница 889: ...dicated value This command deter mines under which conditions a service request is triggered Parameters Contents Contents of the service request enable register in decimal form Bit 6 MSS mask bit is a...

Страница 890: ...servicing of the subsequent commands until all preceding commands have been executed and all signals have settled see also command synchronization and OPC Usage Event 11 9 System Configuration Genera...

Страница 891: ...parameters of the current instrument setting are passed between applications e g center frequency level settings Note that although the default value is OFF this parameter is not reset using the PRESE...

Страница 892: ...n Off on page 488 SYSTem FIRMware UPDate This command starts a firmware update using the msi files in the selected directory The default path is D FW_UPDATE The path is changed via the MMEMory COMMent...

Страница 893: ...STem PRESet This command initiates an instrument reset The effect of this command corresponds to that of the PRESET key with manual oper ation or to the RST command For details on preset settings refe...

Страница 894: ...loudspeaker to mute SYSTem TIME Time This command sets the internal clock The sequence of entry is hour minute second Only user accounts with administrator rights are able to set the time Parameters...

Страница 895: ...N OFF RST OFF Example SYST DISP FPAN ON Manual operation See Soft Frontpanel on page 490 SYSTem DEVice ID Returns the unique ID of the R S ESRP instrument Example SYST DEV ID 1307 9002K07 100123 aZ Us...

Страница 896: ...47 11 Example DIAG SERV BIOS Returns the BIOS version Usage Query only DIAGnostic n SERVice HWINfo This command queries the hardware information Suffix n irrelevant Return values HardwareInfo componen...

Страница 897: ...was defined using the ID String User softkey Example SYST IDEN FACT IDN Response Rohde Schwarz R S ESRP 7 1307 9002K07 101768 1 05 Usage Event SCPI confirmed Manual operation See ID String Factory on...

Страница 898: ...495 SYSTem ERRor NEXT This command queries the earliest error queue entry and deletes it Positive error numbers indicate device specific errors negative error numbers are error messages defined by SC...

Страница 899: ...command turns synchronization of the resolution bandwidth and filter type between measurement channels on and off Parameters State ALL 1 Turns on synchronization NONE 0 Turns on synchronization RST NO...

Страница 900: ...n of the preamplifier configuration between mea surement channels on and off Parameters State ALL 1 Turns on synchronization NONE 0 Turns off synchronization RST NONE Example INST COUP GAIN ALL Synchr...

Страница 901: ...etween measurement channels on and off Note that it is not possible to turn off the preselector in receiver mode Thus the prese lector is always turned on when you enter receiver mode Parameters State...

Страница 902: ...ce NSOurce 899 INPut PRESelection STATe 899 INPut ip UPORt STATe 900 INPut ip UPORt VALue 900 OUTPut IF SOURce 900 OUTPut TRIGger 901 OUTPut UPORt STATe 901 OUTPut UPORt VALue 901 DIAGnostic n SERVice...

Страница 903: ...vant Parameters State ON 1 User port is switched to INPut OFF 0 User port is switched to OUTPut RST 1 Manual operation See User Port on page 163 INPut ip UPORt VALue This command queries the control l...

Страница 904: ...the user ports for the AUX PORT connector This 9 pole SUB D male connector is located on the rear panel of the R S ESRP Parameters State ON 1 User port is switched to OUTPut OFF 0 User port is switche...

Страница 905: ...d the reference signal must be connected to the rear panel of the instrument Parameters Frequency Range 1 MHz to 20 MHz Example ROSC EXT FREQ 13 Sets the frequency to 13 MHz SOUR EXT ROSC EXT FREQ 13...

Страница 906: ...available reference type Usage Query only 11 9 6 Calibrating the R S ESRP CALibration ABORt 903 CALibration ALL 903 CALibration RESult 904 CALibration STATe 904 DIAGnostic n SERVice INPut PULSed CFReq...

Страница 907: ...ing and then a service request is generated Usage Query only SCPI confirmed CALibration RESult This command returns the results collected during calibration Return values CalibrationData String contai...

Страница 908: ...es a calibration signal frequency of 128 MHz Usage SCPI confirmed Manual operation See Calibration Frequency RF on page 498 DIAGnostic n SERVice INPut SELect Signal This command selects the source of...

Страница 909: ...FUNction LASTresult 907 DIAGnostic n SERVice SFUNction RESults DELete 907 DIAGnostic n SERVice SFUNction RESults SAVE 907 SYSTem PASSword RESet 907 SYSTem PASSword CENable 907 DIAGnostic n SERVice SFU...

Страница 910: ...n irrelevant Example DIAG SERV SFUN RES DEL Usage Event DIAGnostic n SERVice SFUNction RESults SAVE FileName This command saves the results of the most recent service function you have used in a file...

Страница 911: ...Rectory 913 MMEMory CATalog This command reads the indicated directory Query parameters Path path DOS Path name The path name should be conform to DOS conventions and may also include the drive name A...

Страница 912: ...ile name suffix Length of file in bytes Return values Result total number of bytes the directory contains File name1 suffix1 Length of file 2 in bytes File name 2 suffix2 Length of file2 in bytes File...

Страница 913: ...from 23 05 MMEMory MMEMory STORe1 STATe 1 ACP_T As a result in the selection list for recall settings the comment ACP measurement with Standard Tetra from 23 05 is added to the ACP entry Usage SCPI co...

Страница 914: ...name in single quotes selects the file to be transferred Block binary data block with the following structure digit for the length of the length information length information number of bytes for the...

Страница 915: ...meters FileSource FileDestination file_source file_destination file_name DOS file name The file name includes indication of the path and may also include the drive The path name complies with DOS conv...

Страница 916: ...ntions Example MMEM NAME C R_S instr user PRINT1 BMP Selects the file name Usage Event SCPI confirmed Manual operation See Configuring and Printing the Test Report on page 521 MMEMory RDIRectory Direc...

Страница 917: ...tions Example MMEM CLE STAT 1 TEST Usage Event Manual operation See Delete File on page 503 MMEMory LOAD AUTO Mode This command defines which settings file is automatically loaded after the device is...

Страница 918: ...ry depending on previous actions in the GUI or in previous scripts even if the script starts with the RST command It is therefore recommended that you use the appropriate MMEMory SELect ITEM command b...

Страница 919: ...xport on page 505 MMEMory STORe IQ STATe 1 FileName This command stores the complex I Q data to the specified iq tar file in 32 bit float ing point format Parameters FileName Complete file name includ...

Страница 920: ...l File on page 502 11 11 Using Transducers Before making any changes to a transducer factor or set you have to select one by name with SENSe CORRection TRANsducer SELect Compared to manual configurati...

Страница 921: ...SENSe CORRection TRANsducer SCALing 920 SENSe CORRection TRANsducer SELect 921 SENSe CORRection TRANsducer STATe 921 SENSe CORRection TRANsducer UNIT 921 SENSe CORRection TRANsducer VIEW 922 SENSe CO...

Страница 922: ...mand queries all transducer factors stored on the R S ESRP Parameters FileInformation String containg the file information NameFileN SizeFileN NameFileN Name of the file SizeFileN Size of the file in...

Страница 923: ...Nsducer UNIT on page 921 By default it is dB Example See Designing a transducer factor on page 917 Manual operation See Defining characteristics of a transducer factor on page 479 SENSe CORRection TRA...

Страница 924: ...turns a transducer factor on and off Before you can use the command you have to select a transducer factor with SENSe CORRection TRANsducer SELect You can turn on up to 8 transducer factors at the sa...

Страница 925: ...n use the command you have to select a transducer set with SENSe CORRection TSET SELect Parameters State ON OFF RST OFF Example See Managing a transducer set on page 918 Manual operation See Managing...

Страница 926: ...e Managing transducer sets on page 481 SENSe CORRection TSET RANGe range Frequency Frequency FileName This command selects a set of transducer factors used for a particular frequency range Ranges 1 to...

Страница 927: ...he command you have to select a transducer set with SENSe CORRection TSET SELect Parameters State ON OFF RST OFF Example See Managing a transducer set on page 918 Manual operation See Using transducer...

Страница 928: ...2 Limit lines CMAP13 Limit and margin check PASS CMAP14 Limit and margin check FAIL CMAP15 Softkey text CMAP16 Softkey background CMAP17 Selected field text CMAP18 Selected field background CMAP19 Sof...

Страница 929: ...rgraph CISPR Average CMAP49 Bargraph RMS Average these settings can only be defined via the theme see Chapter 11 14 Display Configuration on page 944 and are thus ignored in the SCPI command HCOPy ABO...

Страница 930: ...optimized color set 3 user defined color set 4 current screen colors without any changes setting for hardcop ies Example HCOP CMAP DEF2 Selects the optimized color set for the color settings of a prin...

Страница 931: ...information see Table 11 25 Parameters Color BLACk BLUE BROWn GREen CYAN RED MAGenta YELLow WHITe DGRAy LGRAy LBLUe LGREen LCYan LRED LMAGenta Example HCOP CMAP2 PDEF GRE Usage SCPI confirmed Manual...

Страница 932: ...rns color printing on and off Parameters State ON OFF ON Color output OFF Black and white output RST OFF Example HCOP DEV COL ON Usage SCPI confirmed Manual operation See Color On Off on page 494 HCOP...

Страница 933: ...rt Example HCOP DEV LANG WMF Usage SCPI confirmed Manual operation See Device Setup on page 515 See Configuring and Printing the Test Report on page 521 HCOPy ITEM ALL This command includes all screen...

Страница 934: ...Tamp STATe 1 2 State This command includes or excludes the time and date in the printout The suffix 1 2 selects the printing device Parameters State ON OFF RST OFF Example HCOP TDST STAT OFF Manual op...

Страница 935: ...ge 514 See Configuring and Printing the Test Report on page 521 HCOPy IMMediate 1 2 NEXT This command initiates a print job If you are printing to a file the file name depends on MMEMory NAME This com...

Страница 936: ...d by quotation marks is output for the next query Further queries are answered by a query error The SYSTem COMMunicate PRINter ENUMerate FIRSt command should be sent previously to return to the beginn...

Страница 937: ...ox It also restores the default names of the header titles in the Header tab of the Test Report Settings dialog box Example HCOP TREP ITEM DEF Restores the default test report configuration Usage Even...

Страница 938: ...rogram Screenshot of the spectrogram SRESults Numerical results for the scan TDSTamp Date and time of the measurement x TRANsducer Characteristics of the transducer Table 11 27 Available items in spec...

Страница 939: ...ow often the items in the report header are displayed in the document Suffix line 1 7 Selects the header line Parameters Repitition ALWays The selected header line is displayed at the top of every pag...

Страница 940: ...ice under Test Renames the third title into Device under Test HCOP TREP ITEM HEAD LINE3 TEXT Some Device Labels the third title as Some Device Manual operation See Designing the Header of the Test Rep...

Страница 941: ...format you can use the following file types bmp jpg png For reports in rtf format you can use the following file types bmp jpg Example HCOP TREP ITEM LOGO C aPicture jpg Includes a picture at the top...

Страница 942: ...HCOPy TREPort PAGesize Size This command selects the size of the test report document Parameters Size A4 Document pages have an A4 size US Document pages have a US letter size RST A4 Example HCOP TREP...

Страница 943: ...test report template Parameters Template String containing the name of the template Example HCOP TREP ITEM TEMP DEL myTemplate Deletes a test report template Usage Event HCOPy TREPort ITEM TEMPlate L...

Страница 944: ...DE Mode Selects the output of the HCOPy IMMediate 1 2 command Parameters Mode SCReen HCOPy IMMediate prints the current screen contents REPort HCOPy IMMediate generates a measurement report RST SCReen...

Страница 945: ...selected with HCOPy TREPort ITEM SELect on page 934 Example HCOP TREP NEW Creates a test report Usage Event Manual operation See Adding and Removing Datasets on page 522 HCOPy TREPort TEST REMove Dat...

Страница 946: ...aset as shown in the Test Report Con tent Selection dialog box If the index number is greater than the number of available data sets the command does nothing State ON OFF Example HCOP TREP TEST SEL 2...

Страница 947: ...gacy command Replaced by Comment HCOPy TREPort ITEM HEADer STATe HCOPy TREPort ITEM SELect Items now selected by the command parameter HCOPy TREPort ITEM DIAGram STATe HCOPy TREPort ITEM SELect Items...

Страница 948: ...y WINDow n SUBWindow m SELect 950 DISPlay WINDow n TEXT STATe 950 DISPlay WINDow n TEXT DATA 951 DISPlay WINDow n TIME 951 DISPlay WINDow n TIME FORMat 951 INITiate n DISPlay 952 DISPlay ANNotation FR...

Страница 949: ...0 8 1 0 Changes the grid color Usage SCPI confirmed Manual operation See Screen Colors on page 492 See Select Object on page 492 See User Defined Colors on page 493 DISPlay CMAP item PDEFined Color T...

Страница 950: ...Off on page 491 DISPlay PSAVe HOLDoff Minutes This command sets the holdoff time for the power save mode of the display The avail able value range is 1 to 60 minutes the resolution 1 minute The entry...

Страница 951: ...P SBAR OFF Manual operation See Status Bar on page 491 DISPlay SKEYs STATe State This command turns the softkey bar in the display on and off Parameters State ON OFF RST ON Example DISP SKEY OFF DISPl...

Страница 952: ...N OFF FRAMe ON Touch screen function is active for entire screen OFF Touch screen is deactivated for entire screen FRAMe Touch screen is deactivated for the diagram area of the screen but active for t...

Страница 953: ...ll screen size DISP WIND2 SIZE LARG Displays the marker table in full screen size Mode all DISPlay WINDow n SUBWindow m SELect This command selects which window screen is active for applications with...

Страница 954: ...Parameters Comment RST empty Example DISP WIND TEXT Noise Measurement Defines the screen title Manual operation See Screen Title On Off on page 491 DISPlay WINDow n TIME State This command turns the d...

Страница 955: ...ISP OFF Sets the display behavior to OFF INIT WAI Starts the measurement with display switched off 11 14 2 Selecting Items to Store and Restore MMEMory SELect ITEM ALL 952 MMEMory SELect ITEM DEFault...

Страница 956: ...t maximum This number includes the activated limit lines and if available the de activated limit lines last used Therefore the combination of the non activated restored limit lines depends on the sequ...

Страница 957: ...OFF i e no traces is stored Example MMEM SEL TRAC ON Manual operation See Select Items on page 502 MMEMory SELect ITEM TRANsducer ALL State This command includes all transducer factors and sets in the...

Страница 958: ...ng Network Connections MMEMory NETWork DISConnect 955 MMEMory NETWork MAP 955 MMEMory NETWork UNUSeddrives 956 MMEMory NETWork USEDdrives 956 SYSTem COMMunicate GPIB SELF ADDRess 957 SYSTem COMMunicat...

Страница 959: ...FF ON Reconnects at logon with the same user name OFF Does not reconnect at logon Usage Event Manual operation See Map Network Drive on page 507 MMEMory NETWork UNUSeddrives This command lists all unu...

Страница 960: ...binary code 0AH used for LF might be included in the binary data block and therefore should not be interpreted as a terminator in this particular case This can be avoided by changing the receive term...

Страница 961: ...ption Also the instrument description can be changed using this command Only user accounts with administrator rights are able to use LXI functionality Parameters String Example SYST LXI MDES Manual op...

Страница 962: ...age 488 SYSTem REVision STRing Name Sets the response to the REV queryto the defined string HP emulation only see Chapter 11 18 GPIB Commands of HP Models 856xE 8566A B 8568A B and 8594E on page 1007...

Страница 963: ...A 8566B 8568A 8568A_DC 8568B 8568B_DC 8591E 8594E 71100C 71200C 71209A PSA89600 PSA ESA RST SCPI Example SYST LANG 8560E Sets the system language to 8560E to simulate the HP analyzer Manual operation...

Страница 964: ...rmed Manual operation See IF Gain Norm Puls on page 487 SYSTem TWALking Mode Defines how remote control programs are evaluated in particular multiple commands in one command line Note that the default...

Страница 965: ...s QUEStionable FREQuency Register on page 553 STATus QUEStionable LIMit Register on page 553 STATus QUEStionable LMARgin Register on page 554 STATus QUEStionable POWer Register on page 555 STATus QUES...

Страница 966: ...SCPI If the error queue is empty the error number 0 No error is returned This command is identical to the SYSTem ERRor NEXT on page 895 command Example STAT QUES Usage Query only SCPI confirmed 11 16...

Страница 967: ...us QUEStionable LIMit n ENABle SumBit STATus QUEStionable LMARgin n ENABle SumBit STATus QUEStionable POWer ENABle SumBit STATus QUEStionable TRANsducer ENABle SumBit These commands control the ENABle...

Страница 968: ...t STATus QUEStionable LMARgin n PTRansition SumBit STATus QUEStionable POWer PTRansition SumBit STATus QUEStionable TRANsducer PTRansition SumBit These commands control the Positive TRansition part of...

Страница 969: ...17 2 Using Marker and Delta Marker 974 11 17 2 1 Shape Factor Measurement using n dB down 974 11 17 2 2 Measuring the Third Order Intercept Point 975 11 17 2 3 Measuring the AM Modulation Depth 976 11...

Страница 970: ...e analyzer SRE 168 Enable service request for STAT OPER STAT QUES and ESR register CALL InstrWrite analyzer ESE 60 Set event enable bit for command execution device dependent and query error CALL Inst...

Страница 971: ...r VI_EVENT_SERVICE_REQ SRQWaitTimeout VI_NULL VI_NULL Wait for service request user inputs are not possible during the waiting time IF status VI_SUCCESS THEN CALL Srq If SRQ is recognized subroutine f...

Страница 972: ...ard and Mouse A frequent problem with remote control programs using Visual Basic is to insert waiting times without blocking the keyboard and the mouse If the program is to respond to user inputs also...

Страница 973: ...oll read status byte IF STB 0 THEN This instrument has bits set in the STB SRQFOUND 1 IF STB AND 16 0 THEN CALL Outputqueue IF STB AND 4 0 THEN CALL ErrorQueueHandler IF STB AND 8 0 THEN CALL Question...

Страница 974: ...outine for the output queue Public Sub Outputqueue Reading the output buffer Try Dim result As String mbSession ReadString Console WriteLine Contents of Output Queue result Catch exp As Exception Cons...

Страница 975: ...SPACE 20 Preallocate blanks to text variable CALL InstrWrite analyzer STATus OPERation EVENt CALL InstrRead analyzer Oper Debug Print Operation Status Oper END SUB REM REM Subroutine for evaluating Q...

Страница 976: ...STOP Stop software END IF IF VAL Esr AND 64 0 THEN Debug Print User request IF VAL Esr AND 128 0 THEN Debug Print Power on END SUB REM REM Subroutine for evaluating the Event Status Register Public S...

Страница 977: ...ep Set frequency CALL InstrWrite analyzer FREQ SPAN 1 MHz Span CALL InstrWrite analyzer BAND RES 30 kHz Resolution bandwidth CALL InstrWrite analyzer INIT WAI Perform sweep with sync Measure 60 dB val...

Страница 978: ...e formula The frequency is set so that the examined mixture products are displayed in the diagram Otherwise the default setting of the R S ESRP is used for measurements SetupInstrument REM Public Sub...

Страница 979: ...T OFF Single sweep CALL InstrWrite analyzer INIT WAI Perform sweep with sync CALL InstrWrite analyzer CALC MARK PEXC 6 DB Peak excursion CALL InstrWrite analyzer CALC MARK STAT ON Marker 1 on CALL Ins...

Страница 980: ...alyzer CALC LIM5 COMM Upper limit Define comment CALL InstrWrite analyzer CALC LIM5 TRAC 1 Assign trace CALL InstrWrite analyzer CALC LIM5 CONT DOM FREQ Define x axis range CALL InstrWrite analyzer CA...

Страница 981: ...VI_SUCCESS THEN CALL Srq If SRQ is recognized subroutine for evaluation Read out result IF status 1 THEN CALL InstrWrite analyzer STAT QUES LIM1 COND Read out STAT QUES LIMit register CALL InstrRead a...

Страница 982: ...ite analyzer INIT WAI Perform sweep with sync CALL InstrWrite analyzer CALC MARK FUNC POW RES ACP Query result CALL InstrRead analyzer result Read out result Print Result CP ACP low ACP up Alt low Alt...

Страница 983: ...Perform sweep with sync CALL InstrWrite analyzer CALC LIM ACP ACH RES Query result of limit check CALL InstrRead analyzer result Read out result Print Result Limit Check result END SUB REM Example of...

Страница 984: ...surement POW ACH REF AUTO ONCE Defines the measured channel power as the reference value for relative power mea surements If the channel power only is to be measured all commands relating to adjacent...

Страница 985: ...LL InstrRead analyzer result Print result END SUB REM 11 17 6 Time Domain Power Measurement In the following example the mean carrier power of a signal with 300 kHz bandwidth at 100 MHz is to be deter...

Страница 986: ...ed The R S ESRP offers two test functions for this task which can be used depending on the signal characteristics In the following the two methods are presented by means of two examples 11 17 7 1 Powe...

Страница 987: ...er FREQ CENT 935 2 MHz SPAN 0Hz Frequency setting CALL InstrWrite analyzer DISP WIND TRAC Y RLEV 10 dBm Set reference level to 10 dB CALL InstrWrite analyzer INP ATT 30 dB Set input attenuation to 30...

Страница 988: ...lses must be above the trig ger threshold The function is therefore particularly suitable for re measuring DUTs already adjusted and whose output power is within the specified range The measurement is...

Страница 989: ...r cmd cmd 8 Number of bursts CALL InstrWrite analyzer cmd CALL InstrRead analyzer result Read results Print result END SUB REM 11 17 8 Fast Level Measurement Using Frequency Lists A typical task for t...

Страница 990: ...Harmonic The frequencies are selected in ascending order to minimize system inherent waiting times resulting from frequency changes At each frequency point the peak power and the rms power are measur...

Страница 991: ...st setup must be taken into account in all power measurements to avoid any measurement errors being intro duced from sources other than the DUT The R S ESRP offers the possibility of defining a freque...

Страница 992: ...and read it from the I Q memory 1 Data is output in the form of voltage values referred to the analyzer input Data can be read in binary or ASCII format In binary format the length information carrie...

Страница 993: ...CALL InstrWrite analyzer TRAC IQ SET NORM 10 MHz 32 MHz IMM POS 0 130560 Read out in binary format CALL InstrWrite analyzer FORMAT REAL 32 Set binary format CALL InstrWrite analyzer TRAC IQ DATA Measu...

Страница 994: ...of digits of length information Dim fileBytes As Long Length of file with trace data in bytes result Space 100 Buffer for simple results Default setting of status register CALL SetupStatusReg Configur...

Страница 995: ...ta fileBytes FileLen H work vb TEST1 R S FSV DFL Determine length of file fileSize Str fileBytes digits Len fileSize 1 Determine number of digits of fileSize Right fileSize digits length information F...

Страница 996: ...CALL InstrWrite analyzer SENSe1 SWEep EGATe ON Switch on the external gate mode CALL InstrWrite analyzer SENSe1 SWEep EGATe SOURce EXTernal Set enternal gate source CALL InstrWrite analyzer SENSe1 SWE...

Страница 997: ...200 US Rem Set time interval Setting the span CALL InstrWrite analyzer SENSe1 FREQuency SPAN 8 MHZ Set the span to 8 MHz Inserting new ranges Rem Enlarge number of ranges to 5 by adding one at the en...

Страница 998: ...nstrWrite analyzer SENSe1 ESPectrum RANGe2 BANDwidth VIDeo 300 kHZ CALL InstrWrite analyzer SENSe1 ESPectrum RANGe2 SWEep TIME 50 ms CALL InstrWrite analyzer SENSe1 ESPectrum RANGe2 RLEVel 10 DBM CALL...

Страница 999: ...alyzer SENSe1 ESPectrum RANGe5 BANDwidth RESolution 1 MHZ CALL InstrWrite analyzer SENSe1 ESPectrum RANGe5 FILTer TYPE CFILter CALL InstrWrite analyzer SENSe1 ESPectrum RANGe5 BANDwidth VIDeo 3 MHZ CA...

Страница 1000: ...ite analyzer SENSe1 LIST RANGe1 BANDwidth RESolution CALL InstrWrite analyzer SENSe1 LIST RANGe1 BANDwidth VIDeo CALL InstrWrite analyzer SENSe1 LIST RANGe1 SWEep TIME AUTO CALL InstrWrite analyzer SE...

Страница 1001: ...LL InstrWrite analyzer SENSe1 LIST RANGe1 BREak OFF CALL InstrWrite analyzer SENSe1 LIST RANGe1 LIMit STARt 20 CALL InstrWrite analyzer SENSe1 LIST RANGe1 LIMit STOP 20 CALL InstrWrite analyzer SENSe1...

Страница 1002: ...lyzer result REM 11 17 14 Averaging I Q Data The R S ESRP has averaging capability also for I Q measurements i e I Q data can be averaged over several test runs This is subject to the following condit...

Страница 1003: ...e are different modes to define the gate areas Edge triggered recording Level triggered recording For more information see IQ Gating in the Remote Commands chapter under Trace IQ subsystem Measurement...

Страница 1004: ...s displayed Edges are only visible around the IFP trigger point otherwise only the signal level is recorded Figure 11 9 Recorded signal using IQ gating in level mode Measurement example Edge mode In e...

Страница 1005: ...gnal in edge mode and displays only the signal without gaps and noise Used values sample rate 32 MHz number of IQ samples 128000 delay 11020 samples length 15500 samples trigger mode EXT Program TRACe...

Страница 1006: ...program records the signal in edge mode and displays only the noise Used values sample rate 32 MHz number of IQ samples 128000 delay 0 samples length 9000 samples trigger mode EXT Program TRACe1 IQ ON...

Страница 1007: ...and the distance between two periods gap In this example a power ramp is recorded triggered by the IFP on the highest edge see Figure 11 13 As a result only the levels of the individual ramp areas of...

Страница 1008: ...ACe1 IQ EGATe GAP 3800 Result Figure 11 14 Result periodic power ramp levels 11 17 16 Usage of Four Spectrum Instances As of R S ESRP firmware version 1 50 you can open up to four tabs for spectrum mo...

Страница 1009: ...ite analyzer CALC MARK FUNC POW SEL OBW select OBW meas CALL InstrWrite analyzer CALC MARK FUNC POW PRES FW3Gppcdma select WCDMA BTS standard Setup the Spectrum 4 Tab CCDF measurement CALL InstrWrite...

Страница 1010: ...4E 8565E 8566A 8566B 8568A 8568B and 8594E Despite the differences in system architecture and device features the supported com mands have been implemented in a way to ensure a sufficiently high degre...

Страница 1011: ...vailable Command sets A and B differ in the rules regarding the command structure 8568A_DC Uses DC input coupling by default if supported by the instrument 8568B Command sets A and B are available Com...

Страница 1012: ...GHz 0 dBm AC 8563E 601 0 Hz 26 5 GHz 0 dBm AC 8564E 601 0 Hz 40 GHz 0 dBm AC 8565E 601 0 Hz 50 GHz 0 dBm AC 8594E 401 0 Hz 3 GHz 0 dBm AC Note The stop frequency indicated in the table may be limited...

Страница 1013: ...HP 8566B HP 8568B HP 8594E available ADJCRT 2 ADJCRT Adjust CRT HP 856xE HP 8566B HP 8568B HP 8594E available ADJIF 2 ADJIF Auto adjust IF HP 856xE HP 8566B HP 8568B HP 8594E available AMB AMB ON OFF...

Страница 1014: ...94E available AXB AXB Exchange trace A and B HP 856xE HP 8566B HP 8568B HP 8594E available B1 B1 Clear Write B HP 8566A HP 8568A available B2 B2 Max Hold B HP 8566A HP 8568A available B3 B3 View B HP...

Страница 1015: ...P 8594E available CF CF numeric_value HZ KHZ MHZ GHZ CF UP CF DN CF Center Frequency HP 8566A HP 8568A HP 856xE HP 8566B HP 8568B HP 8594E available CHANPWR CHANPWR TRA TRB numeric_value Channel Power...

Страница 1016: ...8566A HP 8568A available D1 2 D1 Display Size normal HP 8566A HP 8568A available DA 2 DA Display address available DEMOD 1 DEMOD ON OFF AM FM AF Demodulator HP 856xE HP 8566B HP 8568B HP 8594E availa...

Страница 1017: ...HP 8566B HP 8568B HP 8594E available DONE DONE DONE Done query HP 856xE HP 8566B HP 8568B HP 8594E available DW 2 DW Write to display and increment address available E1 E1 Peak Search HP 8566A HP 8568...

Страница 1018: ...R ERR Error queue query HP 856xE HP 8566B HP 8568B HP 8594E not yet availa ble EX EX Exchange trace A and B HP 8566A HP 8568A available FA FA numeric_value HZ KHZ MHZ GHZ FA UP FA DN FA Start Frequenc...

Страница 1019: ...HP 856xE HP 8566B HP 8568B HP 8594E available GATECTL 1 GATECTL EDGE LEVEL GATECTL HP 856xE HP 8566B HP 8568B HP 8594E available GD 1 GD numeric_value US MS SC GD DN GD UP GD HP 856xE HP 8566B HP 856...

Страница 1020: ...HZ KHZ MHZ GHZ KS DN KS UP KS Marker Frequency Coun ter Resolution HP 8566A HP 8568A available KS KS Manual Peaking HP 8566A HP 8568A available KS KS Lock register HP 8566A HP 8568A available KS KS U...

Страница 1021: ...HP 8566A HP 8568A available KSQ 2 KSQ Band lock off HP 8566A HP 8568A available KST KST Fast Preset HP 8566A HP 8568A available KSV KSV numeric_value HZ KHZ MHZ GHZ KSV Frequency Offset HP 8566A HP 85...

Страница 1022: ...KSp KSp Character display on HP 8566A HP 8568A available KSr KSr Create service request HP 8566A HP 8568A available KSt 2 KSt Band lock on HP 8566A HP 8568A available KSv 2 KSv Signal ident on HP 856...

Страница 1023: ...B DM LG Amplitude Scale Log HP 856xE HP 8566B HP 8568B HP 8594E available LL 2 LL Plot command HP 8566A HP 8568A available LN LN Amplitude Scale Lin HP 856xE HP 8566B HP 8568B HP 8594E available M1 M1...

Страница 1024: ...E available MKA MKA numeric_value MKA Marker Amplitude HP 856xE HP 8566B HP 8568B HP 8594E available MKACT MKACT 1 MKACT Select the active marker HP 856xE HP 8566B HP 8568B HP 8594E not available MKBW...

Страница 1025: ...KFCR Frequency Counter Res olution HP 856xE HP 8566B HP 8568B HP 8594E available MKMIN MKMIN Marker Min HP 856xE HP 8566B HP 8568B HP 8594E available MKN MKN MKN numeric_value HZ KHZ MHZ GHZ MKN DN MK...

Страница 1026: ...6B HP 8568B HP 8594E available MKRL MKRL Ref Level Marker Level HP 856xE HP 8566B HP 8568B HP 8594E available MKSP MKSP Deltamarker to span HP 856xE HP 8566B HP 8568B HP 8594E available MKSS MKSS CF S...

Страница 1027: ...r Track On HP 8566A HP 8568A available MXMH MXMH TRA TRB Maximum Hold HP 856xE HP 8566B HP 8568B HP 8594E available NORMALIZE NORMALIZE Normalize trace HP 856xE HP 8566B HP 8568B HP 8594E not availabl...

Страница 1028: ...A Queries amplitude at the spot frequency HP 856xE available PH_DRIFT 0 for stable signals 1 for drifty HP 856xE available PH_RLVL Reference level for the log plot HP 856xE available PH_SMTHV Trace sm...

Страница 1029: ...TRA ALLDSP GRT Plot command HP 856xE HP 8566B HP 8568B HP 8594E available PP PP Preselector Peaking HP 8566A HP 8568A available PRINT 1 PRINT PRINT 1 0 Hardcopy HP 856xE HP 8566B HP 8568B HP 8594E av...

Страница 1030: ...8594E available RC1 6 RC1 6 Recall Last State HP 8566A HP 8568A available RCLS RCLS numeric_value Recall State Register HP 856xE HP 8566B HP 8568B HP 8594E available RCLT RCLT TRA TRB num ber Recall...

Страница 1031: ...856xE HP 8566B HP 8568B HP 8594E available S1 S1 Continuous Sweep HP 8566A HP 8568A available S2 S2 Single Sweep HP 8566A HP 8568A available SADD add a limit line segment HP 856xE available SAVES SAVE...

Страница 1032: ...8568B HP 8594E not available SRCPOFS 1 SRCPOFS numeric_value DB DM SRCPOFS DN SRCPOFS UP SRCPOFS Source Power Offset HP 856xE HP 8566B HP 8568B HP 8594E not available SRCPWR 1 SRCPWR numeric_value DB...

Страница 1033: ...trace amplitudes HP 8566B HP 8568B HP 8594E available SV1 6 SV1 6 Save State HP 8566A HP 8568A available SWPCPL 2 SWPCPL SA SR SWPCPL Sweep Couple HP 856xE HP 8566B HP 8568B HP 8594E available SWPOUT...

Страница 1034: ...HP 8568A available TDF TDF P TDF M TDF B TDF A TDF I Trace Data Format HP 856xE HP 8566B HP 8568B HP 8594E available TH TH numeric_value DB DM TH DN TH UP TH ON TH OFF TH AUTO TH Threshold HP 856xE H...

Страница 1035: ...E available UR 2 UR Plot Command HP 8566A HP 8568A available VARDEF VARDEF Variable definition arrays are not supported HP 8566B HP 8568B HP 8594E available VAVG VAVG VAVG TRA TRB TRC Video Averaging...

Страница 1036: ...en models A and models B The command structure for models A is as follows command command code SPC data step SPC delimiter command code delimiter data Value SPC units code SPC delimiter SPC data step...

Страница 1037: ...equired for that purpose ERR Deletes the error bit in the status register but always returns a 0 in response FA Default value range step size FB Default value range step size ID M2 Default value range...

Страница 1038: ...shows the default settings obtained after a change of the GPIB lan guage and for the commands IP KST and RESET Model of Trace Points Start Freq Stop Freq Ref Level Input Coupling 8566A B 1001 2 GHz 2...

Страница 1039: ...the number of test points trace points corre sponds to that of the 85xx families 1001 for 8566A B and 8568A B 601 for 8560E to 8565E 401 for 8594E 11 18 7 Trace Data Output Formats All formats are su...

Страница 1040: ...ESRP reports any key pressed on the front panel rather than only the unit keys if bit 1 was enabled Another difference is the behavior of bit 6 when using the STB query On the HP ana lyzers this bit...

Страница 1041: ...f electric shock If moisture enters the casing for example if you clean the instrument using a moist cloth contact with the instrument can lead to electric shock Before cleaning the instru ment other...

Страница 1042: ...ents and connectors from being damaged The antistatic packing foil avoids any undesired electrostatic charging to occur If you do not use the original packaging provide for sufficient padding to preve...

Страница 1043: ...d out by means of query SYSTem ERRor A short explanation of the error cause is given The left hand col umn contains the associated error code Additionally option specific warnings and error messages d...

Страница 1044: ...Printer Not Available This message is displayed when the selected printer is not included in the list of available output devices A possible cause is that the required printer driver is missing or inc...

Страница 1045: ...ktop 1049 Operation with a VNC Client 1053 Starting and Terminating Remote Operation 1054 Deactivating the R S ESRP via Remote Operation 1055 14 1 Configuring the Network After network support has bee...

Страница 1046: ...s the Network Address softkey The submenu is displayed 4 Press the Computer Name softkey and enter the computer name 5 If you enter an invalid name the error message out of range is displayed in the s...

Страница 1047: ...ment 14 1 3 Operating the Instrument Without a Network To operate the instrument without a network connection either temporarily or perma nently no special measures are necessary Windows automatically...

Страница 1048: ...sers 3 Select Add someone else to this PC 4 In the Microsoft account dialog box enter the new user name and password 5 Select OK 6 Select Finish The new user is created 14 1 5 Changing the User Passwo...

Страница 1049: ...tion perform the following steps 1 Press the Windows key or the CTRL ESC key combination on your keyboard to access the operating system of the R S ESRP see also Chapter 3 2 5 4 Accessing the Start Me...

Страница 1050: ...L key on the front panel of the R S ESRP 2 Press the File Manager softkey 3 Press the More softkey 4 Press the Network Drive softkey The Map Network Drive dialog box is displayed 5 Press the Drive lis...

Страница 1051: ...ed 6 In the Drive list select the drive you want to disconnect 7 Confirm with OK 14 1 9 Sharing Directories only with Microsoft Networks Sharing directories makes data available for other users This i...

Страница 1052: ...described in the Windows documentation 14 2 1 Configuring the R S ESRP for Remote Operation To avoid problems use a fixed IP address When a DHCP server is used a new IP address is assigned each time...

Страница 1053: ...Remote Desktop Client The Windows Remote Desktop Client is part of the operating system and can be accessed via Start All Applications Windows Accessories Remote Desktop Con nection 1 Press the Windo...

Страница 1054: ...the options are activated or deactivated 6 To improve the performance you can deactivate the Desktop background Show contents of window while dragging and Menu and window animation options 7 Open the...

Страница 1055: ...S ESRP activate the Printers option Do not change the remaining settings 10 Open the Display tab The options for configuring the R S ESRP screen display are displayed 11 Under Remote desktop size you...

Страница 1056: ...both on the instrument and with the client simultaneously Using the Remote Desktop turns off the display and viewing the display contents is only possible on the controlling computer The complete min...

Страница 1057: ...Select the Connect button The connection is set up 6 If the Drives option is activated on the Local Resources tab a warning is dis played indicating that the drives are enabled for access from the R S...

Страница 1058: ...the Remote Desktop window The connection to the R S ESRP is terminated any time is possible On the R S ESRP a user logs on The connection to the controller is terminated as a result A message is dis...

Страница 1059: ...R S ESRP A safety query is displayed to warn you that the instrument cannot be reactivated via remote operation and asks you whether you want to continue the shutdown process 3 Respond to the safety q...

Страница 1060: ...ion 639 SENSe BANDwidth BWIDth RESolution AUTO 640 SENSe BANDwidth BWIDth RESolution AUTO 798 SENSe BANDwidth BWIDth RESolution FFT 799 SENSe BANDwidth BWIDth RESolution RATio 799 SENSe BANDwidth BWID...

Страница 1061: ...t GAIN STATe 734 SENSe ESPectrum RANGe range INSert 735 SENSe ESPectrum RANGe range LIMit source ABSolute STARt 735 SENSe ESPectrum RANGe range LIMit source ABSolute STOP 735 SENSe ESPectrum RANGe ran...

Страница 1062: ...RANGe range POINts 724 SENSe LIST RANGe range RLEVel 724 SENSe LIST RANGe range SWEep TIME 725 SENSe LIST RANGe range SWEep TIME AUTO 725 SENSe LIST RANGe range TRANsducer 725 SENSe LIST RANGe range...

Страница 1063: ...SENSe SCAN range BANDwidth RESolution 643 SENSe SCAN range INPut ATTenuation 644 SENSe SCAN range INPut ATTenuation AUTO 644 SENSe SCAN range INPut GAIN AUTO 644 SENSe SCAN range INPut GAIN STATE 645...

Страница 1064: ...DELTamarker m FUNCtion FIXed STATe 837 CALCulate n DELTamarker m FUNCtion FMEasurement DETector 747 CALCulate n DELTamarker m FUNCtion FMEasurement LIMit k CONDition 749 CALCulate n DELTamarker m FUN...

Страница 1065: ...ate n FEED 606 CALCulate n FLINe k 673 CALCulate n FLINe k STATe 673 CALCulate n FORMat 856 CALCulate n LIMit k ACPower ACHannel ABSolute 706 CALCulate n LIMit k ACPower ACHannel ABSolute STATe 706 CA...

Страница 1066: ...ATe 685 CALCulate n LIMit k UPPer THReshold 685 CALCulate n LIMit k UPPer DATA 683 CALCulate n MARKer m AOFF 660 CALCulate n MARKer m COUNt 838 CALCulate n MARKer m COUNt FREQuency 838 CALCulate n MAR...

Страница 1067: ...846 CALCulate n MARKer m FUNCtion NOISe RESult 839 CALCulate n MARKer m FUNCtion NOISe STATe 840 CALCulate n MARKer m FUNCtion POWer MODE 692 CALCulate n MARKer m FUNCtion POWer PRESet 696 CALCulate n...

Страница 1068: ...0 CALCulate n MARKer m MAXimum RIGHt 661 CALCulate n MARKer m MAXimum PEAK 661 CALCulate n MARKer m MINimum AUTO 831 CALCulate n MARKer m MINimum LEFT 662 CALCulate n MARKer m MINimum NEXT 662 CALCula...

Страница 1069: ...CALCulate n SGRam HDEPth 620 CALCulate n SGRam TSTamp DATA 620 CALCulate n SGRam TSTamp STATe 826 CALCulate n SGRam STATe 826 CALCulate n STATistics APD STATe 760 CALCulate n STATistics CCDF X Trace 7...

Страница 1070: ...SBAR STATe 948 DISPlay SKEYs STATe 948 DISPlay TBAR STATe 948 DISPlay THEMe CATalog 948 DISPlay THEMe SELect 949 DISPlay TOUChscreen STATe 949 DISPlay WINDow SGRam COLor DEFault 618 DISPlay WINDow SGR...

Страница 1071: ...y TREPort ITEM HEADer LINE line CONTrol 936 HCOPy TREPort ITEM HEADer LINE line TEXT 937 HCOPy TREPort ITEM HEADer LINE line TITLe 937 HCOPy TREPort ITEM LOGO 938 HCOPy TREPort ITEM LOGO CONTrol 938 H...

Страница 1072: ...COUPle BANDwidth 896 INSTrument COUPle CENTer 896 INSTrument COUPle DEMod 897 INSTrument COUPle GAIN 897 INSTrument COUPle LIMit 897 INSTrument COUPle MARKer 898 INSTrument COUPle PRESel 898 INSTrume...

Страница 1073: ...RIGger 901 OUTPut UPORt STATe 901 OUTPut UPORt VALue 901 OUTPut STATe 869 PROBe ID PARTnumber 789 PROBe ID SRNumber 789 PROBe SETup MODE 789 PROBe SETup NAME 789 PROBe SETup STATe 789 PROBe STATe 790...

Страница 1074: ...n PTRansition 965 STATus QUEStionable LIMit n EVENt 963 STATus QUEStionable LMARgin n CONDition 964 STATus QUEStionable LMARgin n ENABle 964 STATus QUEStionable LMARgin n NTRansition 964 STATus QUESt...

Страница 1075: ...Tem IDENtify STRing 894 SYSTem IFGain MODE 961 SYSTem KLOCk 890 SYSTem LANGuage 960 SYSTem LXI INFo 957 SYSTem LXI LANReset 958 SYSTem LXI MDEScription 958 SYSTem LXI PASSword 958 SYSTem PASSword RESe...

Страница 1076: ...TPISample 867 TRACe n IQ STATe 866 TRACe n POINts 643 TRACe n DATA MEMory 823 TRACe n DATA X 824 TRIGger n SEQuence HOLDoff TIME 804 TRIGger n SEQuence IFPower HOLDoff 805 TRIGger n SEQuence IFPower...

Страница 1077: ...7 Amplitude Menu 322 Amplitude distribution 282 AMPT key 322 Analysis Bandwidth definition 445 Annotation 492 AP trace information 65 Application cards 24 Application notes 24 APX tracking generator 4...

Страница 1078: ...e 540 CommandS Programming examples 965 Common commands Syntax 535 Compatible mode hardware setting 63 Computer name Changing 597 1043 Concluding entry 71 CONDition 547 Connector AC power supply 33 AF...

Страница 1079: ...labels 63 Entry Canceling 71 Concluding 71 error messages device specific 1040 Error messages 1040 Error queue 550 Error queues recommendations 563 ESD 37 ESE Event Status Enable register 551 ESR Even...

Страница 1080: ...Hardware information dialog box 495 Hardware settings Display 63 Harmonic Distortion Measurement 308 Results 311 Harmonics Measurement 201 311 HiSLIP Protocol 528 Resource string 527 HOME key 118 HP e...

Страница 1081: ...onfiguring 598 Connector 34 Interface 527 IP address 527 Remote control interface 524 RSIB protocol 529 VISA 527 VXI protocol 528 LAN interface 1042 level axis 325 438 display range 288 289 294 295 32...

Страница 1082: ...example amplitude distribution 426 harmonics 402 high sensitivity harmonics 310 intermodulation 301 modulated signals 419 noise 412 signals in the vicinity of noise 407 spectra of complex signals 404...

Страница 1083: ...y 318 gate signal 350 reference level 325 438 Offset Limit line 188 Trigger 349 443 Offset hardware setting 63 On screen keyboard 70 Online help Working with 20 Operating system 53 Login 54 service pa...

Страница 1084: ...rement 996 time domain power measurement 982 transducers 988 writing files 992 Programming examples Service request 967 Protocol RSIB 529 VXI 528 PTRansition 547 Q QP trace information 65 Queries 526...

Страница 1085: ...RM trace information 65 RMS VBW 333 370 RMS detector 370 RMS value 272 Rotary knob 72 RSIB interface functions 530 protocol 529 RUN CONT key 312 RUN SINGLE key 312 S SA trace information 65 Safety ins...

Страница 1086: ...ion Short 448 Calibrate Transmission 448 Calibration Frequency RF 904 CCDF remote control 760 762 Center Mkr Freq 383 Center Mkr Freq remote control 656 CF Stepsize remote control 634 Chan Pwr Hz 218...

Страница 1087: ...ptions remote control 893 Interpolation Lin Log remote control 920 LAN Reset 958 Last Span 156 Left Limit 182 394 Left Limit remote control 658 Limit Chk On Off remote control 708 712 713 Limit Line S...

Страница 1088: ...ncy 389 390 Ref Point Frequency remote control 836 Ref Point Level 389 390 Ref Point Level remote control 836 Ref Point Time remote control 836 Ref Value remote control 857 Ref Value Position remote c...

Страница 1089: ...g Gate Polarity Pos Neg 348 Trg Gate Polarity Pos Neg remote control 807 810 Trg Gate Source remote control 805 806 810 Trigger Holdoff remote control 804 Trigger Out Low High remote control 901 Trigg...

Страница 1090: ...ode Lvl Edge 350 Gate Settings 350 Gated Trigger 349 General setup 483 GPIB 485 GPIB address 485 GPIB language 486 GPIB terminator LFEOI EOI 489 Handle missing Ext ref 478 Hardware info 495 Harmonic D...

Страница 1091: ...l 392 Std Dev 272 Subnet mask 484 Sweep 330 334 342 Sweep Count remote control 802 Sweep List 234 265 Sweep repeat On Off 487 Sweep Time remote control 803 Sweep Type 329 334 342 Sweeptime Auto 334 34...

Страница 1092: ...0 Common commands 883 STB 545 Step size Center frequency 316 stop frequency 318 storing 1039 Storing Instrument configuration 114 Measurement data 115 String in remote commands 539 Suffixes 536 supply...

Страница 1093: ...king generator 465 Tree walking Softkey 488 TRG enhancement label 63 64 TRIG key 345 trigger external 346 441 level 348 slope 348 443 Trigger Event remote 886 External gate 349 Gated sweep 350 Holdoff...

Страница 1094: ...ng for trigger Status register 557 Web browser Configuration interface 598 weighting filter 216 White papers 24 White space 540 Windows 7 53 Access 56 WordPad Accessing 57 Y Y Axis Max I Q Analyzer 43...

Отзывы: