When you switch between remote control languages, the following settings or
changes are made:
SCPI:
The instrument performs a PRESET.
8566A/B, 8568A/B, 8594E; FSEA, FSEB, FSEM; FSEK:
–
The instrument performs a PRESET.
–
The following instrument settings are changed:
Table 11-29: Instrument settings for emulation of 8566A/B, 8568A/B, 8594E; FSEA, FSEB, FSEM;
FSEK instruments
Model
# of Trace
Points
Start Freq.
Stop Freq.
Ref Level
Input Coupling
8566A/B
1001
2 GHz
22 GHz
0 dBm
AC
8568A/B
1001
0 Hz
1.5 GHz
0 dBm
AC
8560E
601
0 Hz
2.9 GHz
0 dBm
AC
8561E
601
0 Hz
6.5 GHz
0 dBm
AC
8562E
601
0 Hz
13.2 GHz
0 dBm
AC
8563E
601
0 Hz
26.5 GHz
0 dBm
AC
8564E
601
0 Hz
40 GHz
0 dBm
AC
8565E
601
0 Hz
50 GHz
0 dBm
AC
8594E
401
0 Hz
3 GHz
0 dBm
AC
Note
: The stop frequency indicated in the table may be limited to the corresponding
frequency of the R&S
ESRP, if required.
11.18.2
Command Set of Models 8560E, 8561E, 8562E, 8563E, 8564E,
8565E, 8566A/B, 8568A/B, 8591E, 8594E, 71100C, 71200C, and
71209A
As with the original units, the R&S
ESRP includes the command set of the A models in
the command set of the B models.
The HP model 8591E is compatible to HP model 8594E, the HP models 71100C,
71200C, and 71209A are compatible to HP models 8566A/B.
Command
Supported subset
Function
Corresp. HP-
Models
Status
A1
A1
Clear/Write A
HP 8566A/
HP 8568A
available
A2
A2
Max Hold A
HP 8566A/
HP 8568A
available
GPIB Commands of HP Models 856xE, 8566A/B, 8568A/B and 8594E