Remote Control - SCPI
R&S
®
QAR
99
User Manual 1178.7200.02 ─ 09
Example:
//Query the threshold value for the bumper type
RR_L
:MEAS:THR? 'RR_L'
//Set a new threshold value for the bumper type
RR_L
:MEAS:THR 'RR_L', 1.0
6.2.5.2
Configuration
Remote commands described elsewhere:
●
●
●
●
●
●
6.2.6
System
............................................................................................100
................................................................................................101
SYSTem:ERRor:CODE?
This command queries system errors.
Return values:
<ErrorCode>
Example:
//Query system error code
SYST:ERR:CODE?
Usage:
Query only
Code
Message
Description
0
Ok
No error
101
Calibration data missing
Restore calibration data in
C:\CalDataK1
102
Could not load QAR configuration
Reinstall software
103
Invalid operation mode
Reinstall software
104
No configuration found for operation
mode '<opmode>'
Reinstall software
Remote Commands