Preparing for use
R&S
®
RTM3000
19
Getting Started 1335.9084.02 ─ 06
Measuring accessories
Use only probes and measuring accessories that comply with IEC
61010-031.
Signal input and output levels
Information on signal levels is provided in the data sheet. Keep the signal levels
within the specified ranges to avoid damage to the product and connected devi-
ces.
Preventing electrostatic discharge (ESD)
Electrostatic discharge is most likely to occur when you connect or disconnect a
DUT.
►
NOTICE!
Risk of electrostatic discharge. Electrostatic discharge can damage
the electronic components of the product and the device under test (DUT).
Ground yourself to prevent electrostatic discharge damage:
a) Use a wrist strap and cord to connect yourself to ground.
b) Use a conductive floor mat and heel strap combination.
During operation, if the firmware observes a serious unexpected disturbance (e.g.
due to ESD), it resets some hardware components and initiates a new alignment
to ensure proper instrument functioning. Then it restores the user settings to the
state before the disturbance.
3.6
Connecting to power
For safety information, see
"Connecting to power and grounding"
The R&S
RTM3000 can be used with different AC power voltages and adapts
itself automatically to it.
The nominal ranges are:
●
100 V to 240 V AC at 50 Hz to 60 Hz, with maximal 10% voltage fluctuation on
line
●
1.6 A to 0.7 A
●
max. 160 W
Connecting to power