Command Reference
R&S
®
ZNA
880
User Manual 1178.6462.02 ─ 12
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<HorizontalPos>
LEFT | MID | RIGHt
Horizontal position
<VerticalPos>
TOP | MID | BOTTom
Vertical position
Example:
See
Manual operation:
See
8.3.1.19
CALCulate:SMOothing...
The
CALCulate:SMOothing...
commands provide the settings for trace smoothing.
CALCulate<Chn>:SMOothing[:STATe]
..............................................................................880
CALCulate<Chn>:SMOothing:APERture
.......................................................................... 880
CALCulate<Chn>:SMOothing[:STATe]
<Boolean>
Enables or disables smoothing for trace no.
<Chn>
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON | OFF - smoothing on or off.
*RST:
OFF
Example:
See
CALCulate<Chn>:SMOothing:APERture
Manual operation:
See
CALCulate<Chn>:SMOothing:APERture
<SmoothAperture>
Defines how many measurement points are averaged to smooth the trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<SmoothAperture>
Smoothing aperture. A smoothing aperture of n % means that
the smoothing interval for each sweep point i with stimulus value
x
i
is equal to [x
i
- span*n/200, x
i
+ span*n/200], and that the
result of i is replaced by the arithmetic mean value of all mea-
surement points in this interval.
Range:
0.05% to 100%.
*RST:
1
Default unit: %
SCPI Command Reference