Command Reference
R&S
®
ZNA
1310
User Manual 1178.6462.02 ─ 12
ALL
All open recall sets are deleted and a new recall set"Set1" is cre-
ated according to the current preset mode.
See
SINGle
The active setup is reset according to the current preset mode;
the name of the active setup and the parameters of all other set-
ups remain unchanged.
Example:
See
Manual operation:
See
SYSTem:PRESet:STARt
<Boolean>
Defines whether the analyzer firmware always starts with the configured system preset
(factory or user defined) or with the previous state. This is a global setting.
Parameters:
<Boolean>
ON
The FW always starts with the configured system preset.
See
OFF
The analyzer firmware recalls the previous state, i.e. the settings
that were persisted during the previous shutdown.
Manual operation:
See
SYSTem:PRESet:USER[:STATe]
<Boolean>
Determines whether a factory or a user-defined preset is performed.
This setting applies to the analyzer GUI and to the remote commands
*RST
and
is set to
ON
.
Parameters:
<Boolean>
OFF (0)
A factory preset is performed.
ON (1)
A user-defined preset is performed, if the setup file selected
using
erwise, a factory preset is performed.
SCPI Command Reference