Command Reference
R&S
®
ZNA
1309
User Manual 1178.6462.02 ─ 12
Manual operation:
See
SYSTem:PRESet:MODE
<PresetMode>
Allows you to select between different preset modes.
GLOBal and RFOFf presets can only be performed from the analyzer GUI. FILE pre-
sets can also be performed via remote commands, if
is set to
ON
.
Parameters:
<PresetMode>
NORMal | FILE | GLOBal | RFOFf
NORMal
Factory preset
FILE
GLOBal
Factory GUI reset + reset RF connection configuration
between VNA and external devices
RFOFf
Factory
Manual operation:
See
"Normal / RF Off / Normal, GUI, Ext Setup / <File Name>"
SYSTem:PRESet:REMote[:STATe]
<Boolean>
Defines the behavior of the preset commands
*RST
Parameters:
<Boolean>
OFF (0)
*RST
and
SYSTem:PRESet
perform a factory preset.
ON (1)
*RST
and
SYSTem:PRESet
restore the user preset file selected
using
, if this file is available and
valid. Otherwise, they perform a factory preset.
Manual operation:
See
SYSTem:PRESet:SCOPe
<Scope>
Specifies whether the preset actions affect all open recall sets, or only the active recall
set.
command. The
*RST
com-
mand always resets all open recall sets.
Parameters:
<Scope>
ALL | SINGle
SCPI Command Reference