Command Reference
R&S
®
ZNA
1085
User Manual 1178.6462.02 ─ 12
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:TPORt:DEFault
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DELete:ALL
.............................. 1134
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:ACQuire
.......................... 1134
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine
............................1135
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine:TPORt
[SENSe<Ch>:]CORRection:COLLect:AUTO:SAVE
.......................................................... 1137
[SENSe<Chn>:]CORRection:PSTate?
............................................................................ 1137
[SENSe<Chn>:]CORRection:SSTate?
............................................................................ 1138
[SENSe:]CORRection:COLLect:AUTO:CKIT
<Characterization>
Generates a characterization (cal kit file) with the specified name containing the cal kit
data of the active calibration unit (selected via
). The cal kit file can be saved to a directory on the analyzer, to the
calibration unit's internal (flash) memory or to an SD card inserted at the calibration unit
(if available).
Note
that this command can only be executed if the number of cal unit ports is less or
equal to the number of test ports. Furthermore the command assumes the "canonical"
assignment of cal unit ports to test ports: cal unit port 1 assigned to test port 1, cal unit
port 2 assigned to test port 2 etc. The new command
allows for flexible assignments.
Setting parameters:
<Characterization>
Location of the created characterization / cal kit file (string
parameter, extension *.calkit):
If a path is specified, the file is saved to the analyzer's hard disk.
The default directory (
MMEMory:CDIRectory
) is
not
used.
If only the file name is specified, the file is saved to the calibra-
tion unit's internal (flash) memory. The factory calibration data on
the unit is not overwritten.
If the file name is prefixed by "SD:", the file is saved to the SD
card inserted at the calibration unit (if available/accessible).
Example:
CORR:COLL:AUTO:CKIT 'AutoCalChar'
Generate a cal kit file
AutoCalChar.calkit
for the active cali-
bration unit and store it in the calibration unit.
SENSe:CORRection:COLLect:AUTO:CKIT 'SD:test1'
Generate a cal kit file
test1.calkit
for the active calibration
unit and store it in the calibration unit.
Usage:
Setting only
Manual operation:
See
"File name / Comment (Optional)"
[SENSe:]CORRection:COLLect:AUTO:CKIT:PASSword
<Password>
Enters a password to enable a single password-protected action in the automatic cali-
bration or in the characterization wizard. If password protection has been activated
manually in the "Characterize Cal Unit" dialog, the password is required for any single
execution of one the following commands, provided that a cal kit (characterization) file
different from the active characterization is specified:
SCPI Command Reference