Command Reference
R&S
®
ZNA
1086
User Manual 1178.6462.02 ─ 12
[SENSe]:CORRection:COLLect:AUTO
[SENSe]:CORRection:COLLect:AUTO:CKIT
[SENSe]:CORRection:COLLect:AUTO:PORTs
[SENSe]:CORRection:COLLect:AUTO:PORTs:TYPE
[SENSe]:CORRection:COLLect:AUTO:TYPE
Tip:
You have to send
[SENSe:]CORRection:COLLect:AUTO:CKIT:PASSword
repeatedly if your command script uses several password-protected commands. You
do not need a password to perform automatic calibrations using the active cal unit
characterization.
Setting parameters:
<Password>
Password (string parameter), as defined in manual control.
Example:
SENSe:CORRection:COLLect:AUTO:CKIT:PASSword
'My_password'
Enter a password
My_password
(assuming that password pro-
tection has been activated manually).
SENSe:CORRection:COLLect:AUTO '', 1, 2
Perform an automatic 2-port calibration at test ports 1 and 2
using the calibration unit's default calibration kit file and auto-
matic port assignment.
SENSe:CORRection:COLLect:AUTO:CKIT:PASSword
'My_password'
Re-enter the password.
SENSe:CORRection:COLLect:AUTO '', 1, 2
Repeat the calibration.
Usage:
Setting only
Manual operation:
See
[SENSe:]CORRection:COLLect:AUTO:CKIT:PORTs
<Characterization>,
<TestPort1>, <CalUnitPort1>, <TestPort2>...
Generates a characterization (cal kit file) with the specified name containing the cal kit
data of the active calibration unit (
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
). The cal kit file can be saved to a directory on the analyzer, to the calibration unit's
internal (flash) memory or to an SD card inserted at the calibration unit (if available).
[SENSe:]CORRection:COLLect:AUTO:CKIT
, but with flexible port
assignment.
Setting parameters:
<Characterization>
Location of the created characterization / cal kit file (string
parameter, extension *.calkit):
If a path is specified, the file is saved to the analyzer's hard disk.
The default directory (
MMEMory:CDIRectory
) is
not
used.
If only the file name is specified, the file is saved to the calibra-
tion unit's internal (flash) memory. The factory calibration data on
the unit is not overwritten.
SCPI Command Reference