![Rohde & Schwarz 1322.8003K06 Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/rohde-and-schwarz/1322-8003k06/1322-8003k06_getting-started_1477773025.webp)
Preparing for Use
R&S
®
FSPN
25
Getting Started 1179.4370.02 ─ 01
To suppress electromagnetic radiation during operation:
●
Use high-quality shielded cables, for example, double-shielded RF and LAN
cables.
●
Always terminate open cable ends.
●
Ensure that connected external devices comply with EMC regulations.
Preventing electrostatic discharge (ESD)
Electrostatic discharge is most likely to occur when you connect or disconnect a
DUT.
►
NOTICE!
Risk of electrostatic discharge. Electrostatic discharge can damage
the electronic components of the product and the device under test (DUT).
Ground yourself to prevent electrostatic discharge damage:
a) Use a wrist strap and cord to connect yourself to ground.
b) Use a conductive floor mat and heel strap combination.
Signal input and output levels
Information on signal levels is provided in the data sheet. Keep the signal levels
within the specified ranges to avoid damage to the R&S
FSPN and connected
devices.
Considerations for Test Setup