Remote control commands
R&S
®
NRPxxTWG
38
User Manual 1178.8371.02 ─ 08
Usage:
Event
7.4
Preparing for the measurement
Before starting a measurement, you need to do the following:
Selecting the reference source
............................................................................... 38
............................................................................. 38
7.4.1
Selecting the reference source
The
ROSCillator
subsystem contains commands for configuring the reference
source.
[SENSe<Sensor>:]ROSCillator:SOURce
<source>
Sets the source of the reference oscillator. Refers typically to a precision, stabilized
time base.
Parameters:
<source>
INTernal | EXTernal | HOST
INTernal
Internal precision oscillator
EXTernal | HOST
External signal supplied at the host interface connector.
*RST:
If the sensor boots or reboots, the source is set to
INTernal. If the sensor is reset, the source setting is
kept unchanged.
Example:
ROSC:SOUR INT
7.4.2
Selecting a measurement mode
► Before starting a measurement, select the measurement mode using:
The available measurement modes and how to configure them are described in
Chapter 7.7, "Configuring the measurement modes"
[SENSe<Sensor>:]FUNCtion
<function>
Sets the measurement mode.
Preparing for the measurement