background image

    The  Min  Frequency  adjustment  specifies  the 

frequency  used  when  the  0-10VDC  input  is  at 
0VDC.    The  Max  Frequency  adjustment  can  be 
used  to  ignore  the  upper  end  of  the  0-10VDC 
signal.   

C.  0-10VDC  signal  from  a  PLC  can  be  used  to 

remotely vary the output of the control instead of 
the keypad.  The “Amplitude Source” setting must 
be set to 0-10VDC to enable it.  After the 0-10VDC 
feature  is  selected,  the control  will  automatically 
turn ON whenever a signal is applied to the control 
(TB2- “+0-10” & “-“).  When the 0-10VDC signal 
has been removed, the amplitude setting resets to 
zero. 

D. When it is desirable to ignore the external speed 

control  inputs,  the  “Manual”  setting  can  be 

selected. 

 

12.  CFR Positive and Negative Gain

 

 

The CFR Positive and Negative Gain settings control 

the rate the feeder’s vibration level is corrected by the 

control.  When the vibration decreases below the set-

point, the “CFR Positive Gain” sets the rate at which 

the output gets boosted to compensate for a vibration 

decrease.  When the vibration increases, the “CFR 

Negative Gain” sets the rate at which the output gets 

lowered to compensate for a vibration increase.   If 

either  the  CFR  Positive or  Negative gain  is set  too 

low, it will take longer than desired to get back to the 

original feed rate.  If either gain is set too high, the 

control may over-shoot beyond the original feed rate.  

The CFR Positive and Negative Gain settings effect 

the control’s operation when the CFR sensor is used, 

and the “Amplitude Control” is set to “Auto Track.”   

 

13.  Current Limit Trip

 

The current limit comes preset to the maximum value 

to protect the control from damage.  The current limit 

can  be  lowered  to  protect  the  feeder’s  coil  from 

overheating. The load current can be viewed under 

the diagnostic menu. If a fault occurs, Press the 1/0 

button twice to clear the fault.  Control current ratings 

are below.  

 

 

Rating with 

120V input 

Rating with 

240V input 

120V Rating 

below 16Hz

VF3 

3 Amps 

2 Amps 

2 Amps 

VF9 

9 Amps 

7 Amps 

7 Amps 

VF12 

12 Amps 

9 Amps 

8 Amps 

VF18 

18 Amps 

12 Amps 

12 Amps

 

14.  Frequency Settings 

The  “Frequency”  menu  contains  the  portion  of  the 

menu  that  controls  the  frequency  settings.    The 

frequency can be adjusted from 5 to 140Hz (or 180Hz 

on VF-3 and VF-9 controls).  The spring/mass ratio 

of the vibratory bowl determines the natural vibrating 

(resonate)  frequency  of  the  bowl.  The  control’s 

output frequency needs to be adjusted to match the 

natural  frequency  of  the  bowl.    The  control  can  be 

manually  tuned  or  automatically  tuned.    The 

frequency setting is displayed as “F= 120.0Hz.” 

The “Frequency Mode” setting selects either manual 

frequency  adjustment  or  auto  tracking  frequency 

adjustment. 

A.  Manually  finding  the  resonate  frequency  of  the 

bowl is much like finding a station on the AM radio 

band.  Set  the  amplitude  to  about  30%.    Then 

adjust the frequency across its range.  The bowl 

should be expected to vibrate the parts at more 

than  one  spot  across  the  frequency  range.  The 

resonate frequency is the frequency with the most 

vibration.  Once the best feeding frequency range 

has  been  found,  fine  tune  the  frequency  for  the 

best parts movement.  To increase feeder stability 

for  parts  load  fluctuations,  adjust  the  frequency 

down  by  .2  or  .3Hz  so  that  the  feeder  becomes 

slightly over-tuned. 

B. “Auto Scan” scans to locate the bowl’s resonate 

frequency.  Once  auto  tracking  has  found  the 

resonate frequency, it can maintain the resonate 

frequency and amplitude of the feeder as the parts 

load changes.  The CFR sensor is needed in order 

for auto tracking to operate, and “Auto Tracking” 

needs  to  be  turned  on  under  both  “Amplitude 

Source” and “Frequency Mode” menus. 

To  show  when  frequency  “Auto  Tracking  is 

enabled, the normal display menu will show a bold 

F

.”    When  ”

=

”  is  shown  in  bold,  the  control  is 

locked onto the resonate frequency of the feeder. 

C. The frequency can be adjusted with an analog     0-

10 volt input when the amplitude source setting is 

set to 4-20mA.  

 

The Minimum frequency limit can protect the feeder 
from feeding at a low frequency if a spring or weld 
breaks.  The Min. or Max. frequency can block out 
undesirable  frequencies  during  Auto  Scan.  
To

 

avoid

 

coil  damage  and  blown  fuses

 

during  an 

Auto Tune scan, the Minimum frequency

 

should only 

be  adjusted  below  the  45Hz

 

default

 

when  the 

vibratory  feeder  has  been  specifically  designed  for 
operation below 45Hz.

 

  

15.  Resonate Threshold Level

 

The  “Resonate  Threshold  Level”  setting  sets  the 

minimum level of vibration that the control considers 

as  a  resonate  condition  during  an  Autoscan.    The 

setting should be reduced if a “Coarse Scan Error” is 

given  after  two  scan  attempts.  Adjustment  is  not 

normally needed except for some inlines. 

 

16.  Auto Track Dead Band

 

The “Auto Track Dead Band” setting controls how far 

the  resonant  frequency  of  the  vibratory  feeder  can 

deviate before the output frequency of the control is 

adjusted to follow it.   Decreasing the setting narrows 

the range, and increasing the setting makes the dead 

band  range  larger  before  a  reaction  takes  place.   

This setting normally doesn’t need to be changed. 

 

17. Setting the Time Delays 

The  ON  and  OFF  parts-sensor  time-delays  are  set 
independently for a period of 0-20 seconds.  The time 
delay  settings  can  be  adjusted  to  provide  the  best 
individual response  for  the  feeder.    The  time  delays 
can  be  found  under  the  timer  settings  menu.    The 
flashing “=” blinks every quarter second to show when 
either the ON or OFF delay timer is running.  

 

18.  Parts Sensor Settings

 

The “I/O Interface” menu contains the portion of the 

menu that controls the parts sensor polarity.  

The  control  comes  preset  to  “inverted”  sensor 

polarity.    Set  the  sensor  polarity  to  either  “Normal” 

(through  beam)  or  “Inverted”  (proximity  or  retro-

reflective).  The sensor input accepts a PNP sensor.

 

19.  Run Mode Settings

 

The “I/O Interface” menu contains the menu portion 

that controls the run mode and empty bowl logic. 

 A. The control comes preset for normal on/off parts 

sensor operation.  The following can be chosen: 

1) The “Constant On” feature can be used to keep 

the  bowl  running  while  the  Aux  output  switches 

power to a device (air valve, SSR, or relay).  

2) The “2-Speed” feature allows the bowl to keep 

some  vibration  going  to  either  trickle  parts  for 

weigh  counting  or  to  cut  down  the  time  to  full 

speed when a high feed rate is needed.  The parts 

sensor  switches  between  high  and  low  speed 

settings.  Low speed is set by “Min Amplitude.”   

3)  The  “high/low”  function  maintains  the  parts 
level between two parts sensors on the track. The 
second sensor (PNP) gets installed in place of the 
run jumper.        

 

           

 

10/21/20  Page 6 

Содержание 121-200-0762

Страница 1: ...for each unit The Autotune Series of Variable Frequency Feeder Cubes generates an output frequency for feeding that is independent from the power line frequency An optional vibration sensor P N 123 2...

Страница 2: ...RODIX INC rodix com Toll Free 800 562 1868 FAX 815 316 4701 Email custserve rodix com VF 3 3 cover docx 10 3 2020 Page 2 RODIX INC FEEDER CUBE VF Series Wiring Diagram...

Страница 3: ...ngs 2 Enter Restores User Settings 2 Restore Settings 3 Enter Restores User Settings 3 Save Settings 1 Enter Saves User Settings 1 Save Settings 2 Enter Saves User Settings 2 Save Settings 3 Enter Sav...

Страница 4: ...C installations to bypass the phase 2 fuse L2 To bypass remove the wire connecting TB1 L2 to the L2 fuse holder Remove the terminal end of the wire connecting the L2 fuse to the power switch and conne...

Страница 5: ...r setting is displayed in the following manner A 50 0 8 Limiting the Maximum Output of Control The Max Amplitude setting can be adjusted to keep a vibratory feeder from hammering or vibrating excessiv...

Страница 6: ...djust the frequency down by 2 or 3Hz so that the feeder becomes slightly over tuned B Auto Scan scans to locate the bowl s resonate frequency Once auto tracking has found the resonate frequency it can...

Страница 7: ...nown setting Once a feed system has been set up properly the setting should be manually saved into the Save Settings1 memory If an operator disturbs the settings the Restore Settings1 feature can rest...

Страница 8: ...ies Control 1 Add a wire from term TB2 1 of the VF master control to TB2 11 of the subordinate control 2 Add a wire from TB2 2 of the VF master control to TB2 12 of the subordinate control 3 Move the...

Страница 9: ...peration Note It takes 72 hours for the adhesive to fully cure at 70 F Alternatively 8 or M4 screws can be used to mount the sensor to the feeder The mounting holes are 1 375 3 49cm apart D ROUTE THE...

Страница 10: ...f a nearby noise source Solutions for Electrical Noise 1 Use shielded wires for all I O Input Output signals The I O signals may include 4 20mA input Run input Sensor input 0 5VDC input Interlock inpu...

Отзывы: