Phoromat 2000 Operation Manual
- 25 -
2.5 Standard Program Refraction
The Phoromat 2000 includes two standard programs (Fixed Programs 1/2) for full-
corrected far value and prescription.
See “7.
4
Standard Programs” for all test items in the standard programs.
2.5.1 Fixed Program 1
1. Select
‘
Fixed Program 1
’
.
Click
‘
Select
’
button of
‘
Fixed Program 1
’
on the view of selecting program.
2. Press .
Program A starts.
3. Load the AR data
2
1)
Measure the patient’s eye with an auto
Ref/Keratometer and print the result.
2) Press
Confirm that the data No. is the same as the print
data No. of the auto Ref/Keratometer.
See “2.2.1 From an auto
Ref/Kera
tometer ”
3) Press .
The AR data appears in the subjective and AR
data fields.
4. Press .
It is possible to measure the unaided visual
acuity for right eye.
Instruct the patient to look at the presented chart
through the measuring windows of the refractor.
5. Measure the unaided visual acuity.
Refer to Steps 3 to 7 in
“4.1 Unaided Visual Acuity Test”
6. Press .
2
When the data has already been entered from an auto Ref/Keratometer or a lensmeter before Program A, Steps 3, 7,
and 8 are skipped automatically.
Содержание Phoromat 2000
Страница 45: ...Phoromat 2000 Operation Manual 37 3 2...
Страница 99: ......
Страница 100: ......
Страница 101: ......
Страница 102: ......
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......
Страница 108: ......
Страница 109: ......
Страница 110: ......
Страница 111: ......
Страница 112: ......
Страница 113: ......
Страница 114: ......
Страница 115: ......
Страница 116: ......
Страница 207: ......
Страница 215: ...11 Specifications 12 Components 13 EMC ELECTROMAGNETIC COMPATIBILITY...
Страница 229: ...9 Types of Chart 9 1 A Type...