F.  Maintenance 

 

1.  After each use,  please  rinse suction bottle with clean water rather than any  chemical  detergents which 

may cause cracks or damage to the structure of the suction bottle.   

2.  Rinse the internal part of silicone tube

tube connecting adaptor

handle and all fitting adaptors by 

sucking 30cc water then sterilize with an autoclave and wipe with a dry cloth after each use. 

3.  This bottle is made of PC material and available for autoclave at 121°C for 30 min. 

4.  The rack and liquid collection tray are not available for autoclave; please clean with water and sterilize 

with 75% alcohol. 

5.  If SS tube is blocked, please soak it in water for 30 min. and clean the blocked tube with L type cleaning 

rod.   

6.  It is strongly recommended to replace internal rubber adaptor

O-ring and SS spring of the handle with 

new parts every year at least to keep good air tightness. Please replace with new parts by referring to 

<pic 10>.   

Note: O-ring kit (199100-53) available for order 

7.  Disc filter and silicone tubes are consumables, please replace at least every half year to ensure sufficient 

air flow.   

 

G.  Troubleshooting 

 

 

Symptoms 

Possible causes and Solution 

Device fails to start 

1. 

Check the power supply conform to power specified on main unit   

2. 

Blown fuse

 replace fuse if it is burnt out   

3. 

Check if the power switch is in the ON position 

4. 

Vacuum still exists in system

 release vacuum and restart 

5. 

Defective switch

 contact your distributor for repair 

Vacuum fails to adjust 

1. 

Faulty gauge

 contact your distributor for repair 

2. 

Loose regulator

 tighten the regulator by yourself or contact your 

distributor for repair     

3. 

Faulty regulator 

 contact your distributor for repair 

Low filtration speed 

1. 

Regulator improperly set

 set regulator vacuum level 

following instruction manual 

2. 

Air leak

 Check the tubing and connected accessories for possible 

leaks.    (Attach, tighten or replace) 

3. 

Check if the suction bottle is overflowing 

4. 

Check if the PTFE Disc filter is good part 

Weak aspiration 

1. 

Adjust vacuum to proper level or change vacuum source 

2. 

Air leakage from tubing 

 change silicone tube 

3. 

Air leakage form handle 

 change rubber adaptor or O-ring 

4. 

Handle or adaptor is broken 

 contact your distributor for repair 

Содержание BioVac 225

Страница 1: ...BioVac 225 BioVac 225 Plus BioVac 225 BioDolphin Portable suction system ...

Страница 2: ...ce Before using the instrument please ensure that the power supply voltage matches the voltage rating for the instrument Please ensure that all connections are well sealed The disc filter needs to be replaced periodically or immediately once it is contaminated or saturated The built in thermal protector will shut down the instrument automatically when it gets too hot Please check the fuse if the i...

Страница 3: ...C Portable Suction System Diagram pic 1 BioVac 225 pic 2 BioVac 225 Plus pic 3 BioVac 225 BioDolphin ...

Страница 4: ...5 BioDolphin Item Description 1 BioVac 225 suction unit 2 Disc filter 3 Suction bottle 4 Tube connection adaptor 5 2m silicone tube 6 Handle 7 1 channel tip adaptor with ejector 8 8 channel tip adaptor with ejector 9 1 channel tip adaptor 150 mm 10 Repair kit of handle Item Description 1 BioVac 225 suction unit 2 Disc filter 3 Suction bottle 4 Tube connection adaptor 5 2m silicone tube 6 BioDolphi...

Страница 5: ...use 3 On Off switch 4 Vacuum indicator 5 Vacuum regulator 6 Vacuum pump inlet 7 Suction bottle outlet 8 Suction bottle inlet pic 7 BioVac 225 suction unit and suction bottle part list pic 8 Overflow protection Item Description 1 Overflow protection ...

Страница 6: ...ton 4 Handle end pic 9 Handle part list pic 10 Handle Disassembly pic 11 Operating of 8 channel tip adaptor with ejector pic 12 Change different tip adaptor item Description 1 Rubber adaptor 2 O ring AS004 3 O ring P7 4 Rubber valve 5 S S spring 6 O ring AS012 ...

Страница 7: ...ance with the applicable regulations h Proceed with cleaning and maintenance after every use Note Check the liquid level of the waste bottle before each use and then empty it as per the applicable regulations 2 BioVac 225 Plus pic 2 a Ensure that the overflow protection is installed properly then place the suction bottle on the holder Please refer to pic 9 when installing it Note please always put...

Страница 8: ...mm can be plugged into handle directly for waste suction of microplate or culture dish by pressing the suction button or pressing lock button to keep a continuous operation f The 1 channel tip adaptor is recommended to be plugged into the handle then have it connect with 200 µl disposable tip with which to prevent cross contamination in common microplate and culture dish g For 96 well microplate E...

Страница 9: ...are consumables please replace at least every half year to ensure sufficient air flow G Troubleshooting Symptoms Possible causes and Solution Device fails to start 1 Check the power supply conform to power specified on main unit 2 Blown fuse replace fuse if it is burnt out 3 Check if the power switch is in the ON position 4 Vacuum still exists in system release vacuum and restart 5 Defective switc...

Страница 10: ...lue 199120 00 BioDolphin P Suction kit pink 199130 00 BioDolphin G Suction kit green 199100 30 Handle 199100 46 1 channel tip adaptor 199100 47 1 channel S S adaptor 40 mm 199100 48 1 channel S S adaptor 80 mm 199100 49 8 channel S S adaptor 40 mm 199100 50 8 channel tip adaptor with ejector 199100 75 1 channel tip adaptor with ejector 199100 76 1 channel tip adaptor 150 mm 199100 61 Rack 199100 5...

Отзывы: