F. Maintenance
1. After each use, please rinse suction bottle with clean water rather than any chemical detergents which
may cause cracks or damage to the structure of the suction bottle.
2. Rinse the internal part of silicone tube
、
tube connecting adaptor
、
handle and all fitting adaptors by
sucking 30cc water then sterilize with an autoclave and wipe with a dry cloth after each use.
3. This bottle is made of PC material and available for autoclave at 121°C for 30 min.
4. The rack and liquid collection tray are not available for autoclave; please clean with water and sterilize
with 75% alcohol.
5. If SS tube is blocked, please soak it in water for 30 min. and clean the blocked tube with L type cleaning
rod.
6. It is strongly recommended to replace internal rubber adaptor
、
O-ring and SS spring of the handle with
new parts every year at least to keep good air tightness. Please replace with new parts by referring to
<pic 10>.
Note: O-ring kit (199100-53) available for order
7. Disc filter and silicone tubes are consumables, please replace at least every half year to ensure sufficient
air flow.
G. Troubleshooting
Symptoms
Possible causes and Solution
Device fails to start
1.
Check the power supply conform to power specified on main unit
2.
Blown fuse
replace fuse if it is burnt out
3.
Check if the power switch is in the ON position
4.
Vacuum still exists in system
release vacuum and restart
5.
Defective switch
contact your distributor for repair
Vacuum fails to adjust
1.
Faulty gauge
contact your distributor for repair
2.
Loose regulator
tighten the regulator by yourself or contact your
distributor for repair
3.
Faulty regulator
contact your distributor for repair
Low filtration speed
1.
Regulator improperly set
set regulator vacuum level
following instruction manual
2.
Air leak
Check the tubing and connected accessories for possible
leaks. (Attach, tighten or replace)
3.
Check if the suction bottle is overflowing
4.
Check if the PTFE Disc filter is good part
Weak aspiration
1.
Adjust vacuum to proper level or change vacuum source
2.
Air leakage from tubing
change silicone tube
3.
Air leakage form handle
change rubber adaptor or O-ring
4.
Handle or adaptor is broken
contact your distributor for repair