![Roche CoaguCheck Pro II Скачать руководство пользователя страница 153](http://html1.mh-extra.com/html/roche/coagucheck-pro-ii/coagucheck-pro-ii_operators-manual_1472337153.webp)
Maintenance and Care •
8
153
8.5
Cleaning/disinfecting the test strip guide
Use the solutions recommended on page 151 for clean-
ing/disinfecting the test strip guide. Apply the solutions
for a contact time of > 1 minute (refer to the correspond-
ing product labeling) using lint-free cotton swabs/buds.
1
With the meter powered off, remove the test strip
guide cover to clean it. (Use your thumbnail to open
the cover of the test strip guide by pressing its front
edge upward.) Move the cover safely away from the
meter. Then rinse the cover with warm water or
wipe it clean using the solutions recommended
above. Let the test strip guide cover dry
for at least
10 minutes
before re-attaching it.
2
Hold the meter upright with the test strip guide
facing down.
■
Clean the easily accessible white areas with a mois-
tened cotton swab/bud.
■
Ensure the swab/bud is only damp, not wet. Wipe
away residual moisture and fluids.
Содержание CoaguCheck Pro II
Страница 1: ...CoaguChek Pro II Operator s Manual ...
Страница 3: ...0 7154879001 02 2015 11 EN CoaguChek Pro II Operator s Manual Version 2 0 ...
Страница 6: ...6 This page intentionally left blank ...
Страница 28: ...28 1 3 Overview of the meter elements C D F H I G J B A E E ...
Страница 42: ...42 This page intentionally left blank ...
Страница 116: ...116 This page intentionally left blank ...
Страница 138: ...138 This page intentionally left blank ...
Страница 168: ...168 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 PDF417 QR Code DataMatrix ...
Страница 170: ...170 This page intentionally left blank ...
Страница 174: ...174 This page intentionally left blank ...
Страница 182: ...182 This page intentionally left blank ...
Страница 186: ...186 This page intentionally left blank ...
Страница 187: ......