118
Measurement optics error. Clean the measurement
optics (Chapter 7.1). Check that the test strip is
lying flat and straight in the test strip guide. If the
message persists, the meter has a fault. Please call
your customer support and service centre.
Although rarely occuring - you applied too little
blood to the strip. Repeat the test with a new
Accu-Chek Active test strip.
The measurement optics are dirty or you have
inserted a test strip that has already been used
or is expired. Clean the measurement optics
(Chapter 7.1) and repeat the test with a new
Accu-Chek Active test strip.
The meter was exposed to a strong electro-
magnetic field. Move somewhere else.
The coding chip was removed during testing, has
a fault, or is not properly inserted in the meter.
Remove the coding chip and re-insert it.
8.3 Troubleshooting
man_04493168001_02_EnUK.indd 118
1/12/2006, 11:29:25 ðì
Содержание ACCU-CHECK Active
Страница 1: ...ACCU CHEK Active BLOOD GLUCOSE SYSTEM Reference Manual...
Страница 6: ...Parts summary Accu Chek Active 8 7 1 3 2 9...
Страница 8: ...Parts summary Accu Chek Active 10...
Страница 9: ...Coding chip slot Infrared interface 10 11 11...
Страница 24: ...3 2 Basic settings...
Страница 26: ...3 2 Basic settings...
Страница 28: ...3 2 Basic settings...
Страница 30: ...3 2 Basic settings...
Страница 32: ...3 2 Basic settings...
Страница 36: ...3 3 Coding...
Страница 40: ...4 3 Performing a blood glucose test...
Страница 42: ...4 3 Performing a blood glucose test...
Страница 44: ...4 3 Performing a blood glucose test...
Страница 48: ...4 3 Performing a blood glucose test...
Страница 50: ...4 3 Performing a blood glucose test...
Страница 52: ...4 3 Performing a blood glucose test...
Страница 56: ...4 3 Performing a blood glucose test...
Страница 60: ...4 3 Performing a blood glucose test...
Страница 64: ...5 2 Retrieving results...
Страница 66: ...5 2 Retrieving results...
Страница 68: ...5 2 Retrieving results...
Страница 70: ...5 2 Retrieving results...
Страница 72: ...5 2 Retrieving results longer than 3 seconds...
Страница 74: ...5 2 Retrieving results longer than 3 seconds longer than 5 seconds...
Страница 78: ...5 3 Downloading data to a PC longer than 3 seconds...
Страница 80: ...5 3 Downloading data to a PC...
Страница 84: ...6 2 Preparing for the quality control test...
Страница 88: ...6 3 Quality control test procedure...
Страница 90: ...6 3 Quality control test procedure...
Страница 92: ...6 3 Quality control test procedure...
Страница 96: ...6 3 Quality control test procedure...
Страница 98: ...6 4 Performing a full display check longer than 3 seconds...
Страница 102: ...7 1 Cleaning your Accu Chek Active meter...
Страница 104: ...7 1 Cleaning your Accu Chek Active meter...
Страница 108: ...7 2 Battery life and battery replacement...
Страница 127: ...A B D C E F H I G...