Version 1.0 May 2016
Pg. 31
User Instruction Manual
NoDec
®
AIRSTREAM 3
High Pressure Alarm
This condition could be caused by an over pressure in the mattress
e.g. by another person sitting on the mattress. This alarm is shown
by the following screen, accompanied by the
two LEDs flashing
RED
.
To resolve this issue, carefully push the CPR button and let out a
small amount of air and then re attach the hose end. Reset the
pump afterwards. Persistent high pressure alarms may indicate an
internal fault. If if this occurs mattress system use should cease.
The pump should immediately be sent for servicing.
Low Pressure Alarm
This condition could be caused by releasing the CPR valve or a
leak in the mattress. This alarm is shown by the following screen,
accompanied by the
two LEDs flashing RED
.