OPERATION
...............................................................................................................................................................................................................
...............................................................................................................................................................................................................
Manual EC 24 · EN05 · 2015-02
27
Error messages
In coordinate G 01 error messages are displayed as error number followed by “Err”.
The messages no. 1 to 8 are fault messages. In this case the main totalizers stop and the disturbing
quantity totalizers start to run. The messages no. 9 to 16 are warnings and have no effect on the
totalizers.
Error no.
Hexadecimal
Description
1
0x0001
Parameters in EEPROM are wrong (may be wrong version)
2
0x0002
EEPROM can not be written (parameters are lost)
3
0x0004
Analog digital converter for temperature measurement
4
0x0008
Analog digital converter for pressure measurement
5 0x0010
Sensor
failure
6 0x0020
Measured
temperature outside limits
7
0x0040
Measured pressure outside limits
8
0x0080
Hardware and software configuration do not fit
9 0x0100
Tamper
contact
10 0x0200
Error
during GERG calculation
11
0x0400
Error in flow rate measurement
12
0x0800
Error in monthly data logger
13
0x1000
Current output
14
0x2000
Too many output pulses in buffer (>500)
15
0x4000
Sensor defect (check hardware)
16
0x8000
Sensor defect (check hardware)
In the column “Hexadecimal” the error bit strings displayed in coordinate Z 09 are listed.