61-1007RK/61-0198RK Combustible Gas Detector • 13

1.

Follow the directions in the controller’s operator’s manual for setting the zero reading 
(sometimes called the fresh air reading).

2.

When the directions call for exposing the detector to zeroing gas, screw the regulator 
into the cylinder, turn the on/off knob counterclockwise to open the regulator, and 
allow the gas to flow to the detector for 1 minute.

3.

Turn the regulator’s on/off knob clockwise to close the regulator.

4.

Set the zero reading according to the controller’s operator’s manual.

5.

Unscrew the 100% nitrogen cylinder from the regulator.

Setting the Response Reading

1.

Follow the directions in the controller’s operator’s manual for setting the response 
reading (span).

2.

When the directions call for exposing the detector to calibration gas, screw the CO

2

 

cylinder onto the regulator, turn the on/off knob counterclockwise to open the 
regulator, and allow the gas to flow to the detector for 1 minute.

3.

Turn the regulator’s on/off knob clockwise to close the regulator.

4.

Set the response reading according to the controller’s operator’s manual.

5.

Unscrew the CO

2

 cylinder from the regulator.

Returning to Normal Operation

1.

Remove the sample tubing from the regulator.

2.

Remove the calibration cup from the detector. Leave the sample tubing connected to 
the calibration cup for convenience.

3.

Allow the reading at the controller to decrease below the alarm points before 
returning the controller to normal operation or enabling external alarms.

NOTE:

If you do not allow the gas reading to decrease below the alarm points, then 
unwanted alarms may occur.

4.

Verify that the controller display reading decreases and stabilizes at a typical 
background CO

2

 level. The 0-5,000 ppm and 0-5 %volume detectors will display a 

reading in a fresh air environment because they have low ranges. The 0-50 %volume 
and 0-100 %volume detectors will not display a reading in fresh air because their 
ranges are too large.

5.

Store the components of the calibration kit in a safe and convenient place.

Содержание 61-0198RK

Страница 1: ...www rkiinstruments com 61 1007RK 61 0198RK CO2 Detector Operator s Manual Part Number 71 0200RK Revision A Released 4 12 11...

Страница 2: ...riodic calibration and maintenance of the detector is essential for proper operation and correct readings Please calibrate and maintain this detector regularly Frequency of calibration depends upon th...

Страница 3: ...eplacement costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIO...

Страница 4: ...tor 4 Wiring the CO2 Detector to a Controller 5 Startup 7 Introducing Incoming Power 7 Setting the Zero Signal 7 Maintenance 8 Preventive Maintenance 8 Troubleshooting 9 Replacing the CO2 Detector 10...

Страница 5: ...x and junction box terminal block Specifications Table 1 lists specifications for the CO2 detector WARNING When using the 61 1007RK you must follow the instructions and warnings in this manual to assu...

Страница 6: ...the 61 1007RK 10 which has a detection range of 0 100 volume The detector is an infrared type of detector This section describes the components of the 61 1007RK and 61 0198RK detectors The 61 1007RK...

Страница 7: ...ignal reaches the controller Housing Cap Cap Gasket The housing cap screws onto the detector housing It retains the plug in sensor and protects it from damage A foam gasket is installed inside the hou...

Страница 8: ...If you purchased a 61 0198RK detector that is factory installed in and factory wired to a controller detector installation is not necessary Mounting the CO2 Detector 1 Select a mounting site that is...

Страница 9: ...e or four wires in conduit through the top conduit hub of the junction box Use appropriate conduit fittings and construction technique for the environmental rating of the junction box The junction box...

Страница 10: ...f using shielded cable connect the cable s drain wire to an available chassis ground at the controller RKI controllers typically have a ground stud that is a convenient grounding location Controller L...

Страница 11: ...kit that includes a calibration cup a calibration cylinder of 100 nitrogen sample tubing and a 0 5 LPM liters per minute fixed flow regulator with an on off knob 1 Screw the calibration cup onto the...

Страница 12: ...ing device during the response test procedure NOTE Performing a response test on the CO2 detector may cause alarms Be sure to put the controller into its calibration program or disable external alarms...

Страница 13: ...ting The troubleshooting guide describes symptoms probable causes and recommended action for problems you may encounter with the CO2 detector NOTE This troubleshooting guide describes detector problem...

Страница 14: ...CAUTION The plug in sensor contains electrolyte which is a dilute acid Do not disassemble the sensor when replacing it with a new one If sensor electrolyte comes in contact with your skin wash affecte...

Страница 15: ...a controller in a junction box disconnect the detector leads from the terminal block in the junction box Note the position of the color coded leads as you remove them 5 Unscrew the detector from the...

Страница 16: ...des procedures to prepare for calibration set the zero reading set the response reading and return to normal operation WARNING The controller is not an active gas monitoring device during the calibrat...

Страница 17: ...he gas to flow to the detector for 1 minute 3 Turn the regulator s on off knob clockwise to close the regulator 4 Set the response reading according to the controller s operator s manual 5 Unscrew the...

Страница 18: ...5040RK 03 CO2 replacement plug in sensor 0 5 Volume 61 5040RK 05 CO2 replacement plug in sensor 0 50 Volume 61 5040RK 10 CO2 replacement plug in sensor 0 100 Volume 71 0200RK 61 1007RK 61 0198RK CO2 D...

Отзывы: