101 • Downloading Instrument Logger Data
Tab. The test results will display on the Unit Tab until you press the Alarm
Check End button in the Function Tab even if the GX-6000 is turned off.
9 . If a printer is connected, a Print button will appear in the Function Tab’s Alarm
Check screen. Press the Print button to print the alarm check results.
10 . Press the Function Tab to display the Function Tab if it is not already
displayed. Press the Alarm Check End button to return to the top level
Function Tab screen. The test results that were being displayed in the Unit Tab
will disappear once the Alarm Check End button is pressed.
11 . To perform any other operations, press the BUMP, CAL, or Alarm Check
buttons on the Function Tab.
12 . To turn the GX-6000 off:
If the alarm check was successful, the GX-6000 will shut off after 15 seconds.
If the alarm check failed, the GX-6000 will shut off after 10 minutes. If buttons
are pressed before the SDM-6000 turns off the GX-6000, the station will
automatically turn the instrument off 10 minutes after the last button push.
To turn off the GX-6000 before it is automatically turned off, press the Detector
Off button in the lower left corner of the Unit Tab. If you have a multi-station
installation, only the master (leftmost) station will have a Unit Tab. The slave
stations will only have a main screen and the Detector Off button will be
displayed in the lower left corner.
13 . The results of the alarm check will be stored in the SDM-6000’s SD card.
14 . Push the red gas in/gas out cover toward the GX-6000 and then release it. It
should return to its unused state at the top of the SDM-6000.
15 . Remove the GX-6000 from the SDM-6000.
16 . Repeat step 12 - step 15 for any additional GX-6000s.
Downloading Instrument Logger Data
The SDM-6000 allows you to download the data in a connected GX-6000 to the SD
card. This can be done by setting up an SDM-6000 system to download the data
automatically at startup or it can be done manually. See “Updating the Data Logger
Settings” on page 62 for instructions to setup an SDM-6000 system for automatic
instrument data downloading.
To manually download instrument data to the SD card, do the following:
1 . Turn on the SDM-6000 system as described in “Turning on a SDM-6000
2 . Confirm that you have an SD card installed in the system.
NOTE:
If an SD card is not installed, then test data and instrument data logger
data downloaded manually or automatically cannot be saved.