89
•
Chapter 5: User Mode and Calibration
GX-3R Pro Operator’s Manual
14. Use AIR to place the cursor next to
ESCAPE
.
15. Press and release POWER MODE to return to the
GAS CAL
menu.
16. See “Exiting the GAS CAL Menu” on page 90 to return to User Mode.
Exiting the GAS CAL Menu
1. While in the
GAS CAL
menu, press AIR to scroll to
ESCAPE
.
2. Press and release POWER MODE. The instrument returns to the
GAS CAL
menu item in
User Mode.
3. See “Returning to Measuring Mode (START MEASURE)” on page 117 to enter Measuring
Mode.
Setting Calibration Parameters (CAL SETTING)
The
CAL SETTING
menu item has 5 items:
CAL REMINDER
,
CAL INTERVAL
,
CAL
EXPIRED
,
CAL CHECK GAS
, and
ESCAPE
.
1. While in User Mode, use AIR to place the cursor next to
CAL SETTING
.
2. Press and release POWER MODE. The
CAL SETTING
menu appears.
MAINTE
7:49
GAS CAL
CAL SETTING
BUMP SETTING
USER MODE
MAINTE
7:49
CAL REMINDER
CAL INTERVAL
CAL EXPIRED
CAL SETTING