GX-3R Pro Operator’s Manual
Chapter 3: Operation
•
30
4. If
Password Protection
is turned
On
(factory setting is
Off
) using the SDM-3R PC program,
a user-set password is required to turn off the GX-3R Pro. When the password screen appears,
adjust each digit with the AIR button and press and release the POWER MODE button to
move on to the next digit. Once the password is entered, the instrument shuts off.
5. If you plan to leave the instrument off for an extended period of time, see page 134.
Measuring Mode Operation
When the GX-3R Pro completes its startup sequence, it is in Measuring Mode. In Measuring
Mode the GX-3R Pro continuously monitors the sampled atmosphere and displays the target gas
concentrations. The GX-3R Pro is considered to be in Normal Operation if there are no alarm
indications.
Heart Symbol: The heart symbol in the upper left corner of the LCD indicates the operation status
and blinks when normal. A microprocessor error causes the heart symbol to stop flashing or to
disappear.
Check Mark: If
BUMP REMINDER
is set to
ON
and if a bump test is not due, a check mark
appears in the upper left corner of the LCD.
“S”: If the instrument is operating in Stealth Mode, an “S” appears at the top of the LCD.
Buzzer Icon: The buzzer icon indicates the buzzer volume. Lines in front of the icon indicate a
high buzzer volume. No lines in front of the icon indicates a low buzzer volume.
Bluetooth Icon: The Bluetooth icon flashes when the instrument is trying to pair to a phone. The
Bluetooth icon is steadily on when the instrument is paired to a phone. The Bluetooth icon is off
when the instrument is not paired and is not trying to pair to a phone.
Backlight: In a low-light environment, press and release either button to turn on the display
backlight. See page 109 to set the backlight time.
Confirmation/Non-Compliance Indicator: If the
BEEP
menu item in User Mode is set to anything
other than
OFF
, the GX-3R Pro gives periodic indications to confirm that it’s operating or to
indicate a non-compliance (see page 106 for instructions).
7:49
0000
PASSWORD
S
7:49
20.9 0.00
O2 % CO2 vol%
0 0 0.0
CH4 %LEL CO ppm H2S ppm