31
•
Chapter 5: Configuration Mode
65-2646XL-05-04 M2A Transmitter Operator’s Manual
3. If you want to change the currently displayed parameter, use the UP/YES and DOWN/NO
buttons to adjust it to the desired setting, then press ENTER to continue to the next parameter.
If the currently displayed parameter is OK, press the ENTER button to proceed to the next
parameter.
Table 5 lists the M2A parameters you can set. Table 5 also lists the factory set value for
each parameter.
Table 5: Configuration Parameters
Parameter
(Factory Set Value)
Description
ALARM-1 (level)
(25 ppm)
The gas reading at which the M2A initiates an alarm 1 condition.
ALARM-1 (activation)
(Increase)
Indicates if the alarm 1 circuit is activated by gas readings
increasing (
Increase
) or decreasing (
Decrease
) to the ALARM-1
Level.
ALARM-1 (relay action)
(N. DE-EN)
If set as
N. DE-EN
, the alarm 1 relay is de-energized in normal
operation and energizes when an alarm 1 condition is initiated.
If set as
N. EN
, the alarm 1 relay is energized in normal operation
and de-energizes when an alarm 1 condition is initiated.
ALARM-1 (relay reset)
(LATCH)
If set as
LATCH
, you must press the ENTER button to reset the
alarm 1 circuit after the alarm 1 condition passes.
If set as
SELF-RST
, the M2A automatically resets the alarm 1
circuit after the alarm 1 condition passes.
A1 OnDy (alarm 1 on
delay)
(1 secs)
The amount of time the M2A delays activation of the alarm 1 cir
-
cuit once an alarm 1 condition is initiated. It can be set in 1 sec
-
ond increments from 0 - 60 seconds, in 1 minute increments from
1 - 15 minutes, and in 15 minute increments from 15 - 60 min
-
utes.
A1 OffDy (alarm 1 off
delay)
(0 sec.)
The amount of time the M2A delays turning off the alarm 1 cir
-
cuit once an alarm 1 condition passes. This parameter appears
only if the alarm 1 relay reset setting is set to
SELF-RST
. It can
be set in 1 second increments from 0 - 60 seconds, in 1 minute
increments from 1 - 15 minutes, and in 15 minute increments
from 15 - 60 minutes.
ALARM-2 (level)
(50 ppm)
The gas reading at which the M2A initiates an alarm 2 condition.
ALARM-2 (activation)
(Increase)
Indicates if the alarm 2 circuit is activated by gas readings
increasing (
Increase
) or decreasing (
Decrease
) to the ALARM-1
Level.